Detection of simulated chest lesions by using soft-copy reading: Comparison of an amorphous silicon flat-panel-detector system and a storage-phosphor system

被引:20
作者
Goo, JM
Im, JG
Lee, HJ
Chung, MJ
Seo, JB
Kim, HY
Lee, YJ
Kang, JW
Kim, JH
机构
[1] Seoul Natl Univ, Coll Med, Dept Radiol, Chongno Gu, Seoul 110744, South Korea
[2] SNUMRC, Inst Radiat Med, Clin Res Inst, Chongno Gu, Seoul 110744, South Korea
[3] Eulgi Univ, Sch Med, Dept Radiol, Seoul, South Korea
[4] Univ Ulsan, Asan Med Ctr, Dept Radiol, Seoul, South Korea
[5] Natl Canc Ctr, Dept Radiol, Goyang, South Korea
关键词
phantoms; radiography; comparative studies; digital; flat panel; thorax;
D O I
10.1148/radiol.2241011441
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
PURPOSE: To compare observer performance by using soft-copy images produced by an amorphous silicon flat-panel- detector system and a storage-phosphor system for the detection of simulated chest lesions. MATERIALS AND METHODS: To test the diagnostic performance of these two systems, four types of simulated lesions (nodules, micronodules, lines, and reticular opacities) were superimposed over an anthropomorphic chest phantom. Digital chest radiographs were acquired with amorphous silicon flat-panel-detector (3-K [K = 1,000] matrix, 12 bits) and storage-phosphor radiography (4-K matrix, 10 bits). Six board-certified radiologists evaluated soft-copy images on a high-resolution video monitor (2,048 X 2,560 X 8 bits). A total of 14,400 observations were analyzed in terms of receiver operating characteristics. RESULTS: Average performance in terms of nodule detection was significantly better (P <.05) with the flat-panel- detector system than with the storage-phosphor system. For micronodules, lines, and reticular opacities, no significant detection differences in averaged performance were found between the two detector systems. CONCLUSION: In the evaluation of soft-copy images, the amorphous silicon detector system appears to be superior to the storage-phosphor system for the detection of pulmonary nodules. (C) RSNA 2002.
引用
收藏
页码:242 / 246
页数:5
相关论文
共 19 条
[1]   Detection of simulated interstitial lung disease and catheters with selenium, storage phosphor, and film-based radiography [J].
Bernhardt, TM ;
Otto, D ;
Reichel, G ;
Ludwig, K ;
Seifert, S ;
Kropf, S ;
Rapp-Bernhardt, U .
RADIOLOGY, 1999, 213 (02) :445-454
[2]  
BULPITT CJ, 1987, LANCET, V1, P494
[3]   Principles of digital radiography with large-area, electronically readable detectors: A review of the basics [J].
Chotas, HG ;
Dobbins, JT ;
Ravin, CE .
RADIOLOGY, 1999, 210 (03) :595-599
[4]   Digital chest radiography with a solid-state flat-panel X-ray detector Contrast-detail evaluation with processed images printed on film hard copy [J].
Chotas, HG ;
Ravin, CE .
RADIOLOGY, 2001, 218 (03) :679-682
[5]   RECEIVER OPERATING CHARACTERISTIC RATING ANALYSIS - GENERALIZATION TO THE POPULATION OF READERS AND PATIENTS WITH THE JACKKNIFE METHOD [J].
DORFMAN, DD ;
BERBAUM, KS ;
METZ, CE .
INVESTIGATIVE RADIOLOGY, 1992, 27 (09) :723-731
[6]   Imaging characteristics of an amorphous silicon flat-panel detector for digital chest radiography [J].
Floyd, CE ;
Warp, RJ ;
Dobbins, JT ;
Chotas, HG ;
Baydush, AH ;
Vargas-Voracek, R ;
Ravin, CE .
RADIOLOGY, 2001, 218 (03) :683-688
[7]   HIGH-RESOLUTION COMPUTER DISPLAY OF PORTABLE, DIGITAL, CHEST RADIOGRAPHS OF ADULTS - SUITABILITY FOR PRIMARY INTERPRETATION [J].
FRANK, MS ;
JOST, RG ;
MOLINA, PL ;
ANDERSON, DJ ;
SOLOMON, SL ;
WHITMAN, RA ;
MOORE, SM .
AMERICAN JOURNAL OF ROENTGENOLOGY, 1993, 160 (03) :473-477
[8]   EVALUATION OF A DIGITAL WORKSTATION FOR INTERPRETING NEONATAL EXAMINATIONS - A RECEIVER OPERATING CHARACTERISTIC STUDY [J].
FRANKEN, EA ;
BERBAUM, KS ;
MARLEY, SM ;
SMITH, WL ;
SATO, Y ;
KAO, SCS ;
MILAM, SG .
INVESTIGATIVE RADIOLOGY, 1992, 27 (09) :732-737
[9]  
GLAZER HS, 1994, RADIOL CLIN N AM, V32, P711
[10]   Digital chest radiography with a selenium-based flat-panel detector versus a storage phosphor system: Comparison of soft-copy images [J].
Goo, JM ;
Im, JG ;
Kim, JH ;
Seo, JB ;
Kim, TS ;
Shine, SJ ;
Lee, W .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2000, 175 (04) :1013-1018