Digital chest radiography with a selenium-based flat-panel detector versus a storage phosphor system: Comparison of soft-copy images

被引:31
作者
Goo, JM
Im, JG
Kim, JH
Seo, JB
Kim, TS
Shine, SJ
Lee, W
机构
[1] Seoul Natl Univ, Coll Med, Dept Radiol, Chongro Gu, Seoul 110744, South Korea
[2] SNUMRC, Inst Radiat Med, Chongro Gu, Seoul 110744, South Korea
[3] Gachon Med Sch, Gil Medf Ctr, Dept Radiol, Namdong Gu, Inchon 405220, South Korea
[4] Sungkyunkwan Univ, Coll Med, Samsung Med Ctr, Dept Radiol,Kangnam Ku, Seoul 135710, South Korea
关键词
D O I
10.2214/ajr.175.4.1751013
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
OBJECTIVE, We compared the soft-copy images produced by a digital chest radiograph system that uses a flat-panel X-ray detector based on amorphous selenium with images produced by a storage phosphor radiography system for the visualization of anatomic regions of the chest. MATERIALS AND METHODS. Two chest radiologists and two residents analyzed 46 pairs of posteroanterior chest radiographs on high-resolution video monitors (2560 x 2048 x 8 bits). In each pair, one radiograph was obtained with a storage phosphor radiography system, and the other radiograph was obtained with a selenium-based flat-panel detector radiography system. Each pair of radiographs was obtained at the same exposure settings. The interpreter rated the visibility and radiographic quality of 11 different anatomic regions. Each pair of images was ranked on a five-point scale (1 = prefer image A, 3 = no preference, 5 = prefer image B) for preference of technique. Statistical significance of preference was determined using the Wilcoxon's signed rank test. RESULTS. The interpreters had a statistically significant preference for the selenium-based radiography system in six (unobscured lung, hilum, rib, minor fissure, heart border and overall appearance) of 11 anatomic regions (p < 0.001) and for the storage phosphor system in two regions (proximal airway and thoracic spine) (p < 0.05), Chest radiologists strongly preferred selmium-based images in eight regions, and they did not prefer storage phosphor images in any region. CONCLUSION. The soft-copy images produced by the selenium-based radiography system were perceived as equal or superior to those produced by the storage phosphor system in most but not all anatomic regions.
引用
收藏
页码:1013 / 1018
页数:6
相关论文
共 29 条
[1]   Principles of digital radiography with large-area, electronically readable detectors: A review of the basics [J].
Chotas, HG ;
Dobbins, JT ;
Ravin, CE .
RADIOLOGY, 1999, 210 (03) :595-599
[2]   CHEST RADIOGRAPHY - COMPARISON OF HIGH-RESOLUTION DIGITAL DISPLAYS WITH CONVENTIONAL AND DIGITAL FILM [J].
COX, GG ;
COOK, LT ;
MCMILLAN, JH ;
ROSENTHAL, SJ ;
DWYER, SJ .
RADIOLOGY, 1990, 176 (03) :771-776
[3]   SELENIUM-BASED DIGITAL RADIOGRAPHY OF THE CHEST - RADIOLOGISTS PREFERENCE COMPARED WITH FILM-SCREEN RADIOGRAPHS [J].
FLOYD, CE ;
BAKER, JA ;
CHOTAS, HG ;
DELONG, DM ;
RAVIN, CE .
AMERICAN JOURNAL OF ROENTGENOLOGY, 1995, 165 (06) :1353-1358
[4]   HIGH-RESOLUTION COMPUTER DISPLAY OF PORTABLE, DIGITAL, CHEST RADIOGRAPHS OF ADULTS - SUITABILITY FOR PRIMARY INTERPRETATION [J].
FRANK, MS ;
JOST, RG ;
MOLINA, PL ;
ANDERSON, DJ ;
SOLOMON, SL ;
WHITMAN, RA ;
MOORE, SM .
AMERICAN JOURNAL OF ROENTGENOLOGY, 1993, 160 (03) :473-477
[5]  
GLAZER HS, 1994, RADIOL CLIN N AM, V32, P711
[6]   COMPARISON OF 2048-LINE DIGITAL-DISPLAY FORMATS AND CONVENTIONAL RADIOGRAPHS - AN ROC STUDY [J].
HAYRAPETIAN, A ;
ABERLE, DR ;
HUANG, HK ;
FISKE, R ;
MORIOKA, C ;
VALENTINO, D ;
BOECHAT, MI .
AMERICAN JOURNAL OF ROENTGENOLOGY, 1989, 152 (05) :1113-1118
[7]   DIGITAL CHEST IMAGING WITH A SELENIUM DETECTOR - COMPARISON WITH CONVENTIONAL RADIOGRAPHY FOR VISUALIZATION OF SPECIFIC ANATOMIC REGIONS OF THE CHEST [J].
HEESEWIJK, HPM ;
NEITZEL, U ;
VANDERGRAAF, Y ;
DEVALOIS, JC ;
FELDBERG, MAM .
AMERICAN JOURNAL OF ROENTGENOLOGY, 1995, 165 (03) :535-540
[8]   Subtle pulmonary disease: Detection with computed radiography versus conventional chest radiography [J].
Ishigaki, T ;
Endo, T ;
Ikeda, M ;
Kono, M ;
Yoshida, S ;
Ikezoe, J ;
Murata, K ;
Matsumoto, M .
RADIOLOGY, 1996, 201 (01) :51-60
[9]   INTERPRETATION OF SUBTLE INTERSTITIAL LUNG ABNORMALITIES - CONVENTIONAL VERSUS STORAGE PHOSPHOR RADIOGRAPHY [J].
KIDO, S ;
IKEZOE, J ;
TAKEUCHI, N ;
KONDOH, H ;
TOMIYAMA, N ;
JOKOH, T ;
KOHNO, N ;
TAKASHIMA, S ;
YAMAGAMI, H ;
NAITO, H ;
ARISAWA, J ;
MORIMOTO, S ;
KOZUKA, T .
RADIOLOGY, 1993, 187 (02) :527-533
[10]   Prospective study of PACS: Information flow and clinical action in a medical intensive care unit [J].
Kundel, HL ;
Seshadri, SB ;
Langlotz, CP ;
Lanken, PN ;
Horii, SC ;
Nodine, CF ;
Polansky, M ;
Feingold, E ;
Brikman, I ;
Bozzo, M ;
Redfern, R .