picture archiving and communication system (PACS);
observer performance;
radiography;
comparative study;
digital;
storage phosphor;
D O I:
10.1097/00004424-200301000-00008
中图分类号:
R8 [特种医学];
R445 [影像诊断学];
学科分类号:
1002 [临床医学];
100207 [影像医学与核医学];
1009 [特种医学];
摘要:
RATIONALE AND OBJECTIVES. To investigate how changes in luminance affect the detection accuracy of radiologists viewing chest radiograph images on high-resolution CRT monitors. MATERIALS AND METHODS. Thirteen radiologists performed a detection task for 11 chest radiograph images With simulated nodules on a monitor with 11 luminance conditions (the maximum luminance ranges from 157.4-369.0 candela/m(2)) simulating CRT degraded by long-term usage, under the ambient illumination of 200 lux; the observation order was always from the darkest to the brightest. RESULTS. There was a statistically reliable effect of the 11 monitor display conditions on the detection of nodules (P < 0.001). In the conditions in Which the maximum luminance of the CRT was 60.7% or below that of the standard display luminance, the correctly detected nodule number reliably deteriorated. CONCLUSIONS. The luminance change in CRT monitor display under long-term usage will have a detrimental effect on nodule detection performance in chest radiograph images.