Influence of monitor luminance change on observer performance for detection of abnormalities depicted on chest radiographs

被引:11
作者
Ikeda, M
Ishigaki, T
Shimamoto, K
Kodera, Y
Koyama, S
Usami, H
Kato, K
机构
[1] Nagoya Univ Hosp, Dept Med Informat & Med Records, Showa Ku, Nagoya, Aichi 4668560, Japan
[2] Nagoya Univ, Sch Med, Dept Radiol, Nagoya, Aichi 466, Japan
[3] Nagoya Univ, Sch Hlth Sci, Dept Radiol Technol, Div Med Radiol Technol, Nagoya, Aichi, Japan
关键词
picture archiving and communication system (PACS); observer performance; radiography; comparative study; digital; storage phosphor;
D O I
10.1097/00004424-200301000-00008
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 [临床医学]; 100207 [影像医学与核医学]; 1009 [特种医学];
摘要
RATIONALE AND OBJECTIVES. To investigate how changes in luminance affect the detection accuracy of radiologists viewing chest radiograph images on high-resolution CRT monitors. MATERIALS AND METHODS. Thirteen radiologists performed a detection task for 11 chest radiograph images With simulated nodules on a monitor with 11 luminance conditions (the maximum luminance ranges from 157.4-369.0 candela/m(2)) simulating CRT degraded by long-term usage, under the ambient illumination of 200 lux; the observation order was always from the darkest to the brightest. RESULTS. There was a statistically reliable effect of the 11 monitor display conditions on the detection of nodules (P < 0.001). In the conditions in Which the maximum luminance of the CRT was 60.7% or below that of the standard display luminance, the correctly detected nodule number reliably deteriorated. CONCLUSIONS. The luminance change in CRT monitor display under long-term usage will have a detrimental effect on nodule detection performance in chest radiograph images.
引用
收藏
页码:57 / 63
页数:7
相关论文
共 10 条
[1]
[Anonymous], 2001, Computer-Assisted Research Design and Analysis
[2]
Comparison of a cathode-ray-tube and film for display of computed radiographic images [J].
Cook, LT ;
Cox, GG ;
Insana, MF ;
McFadden, MA ;
Hall, TJ ;
Gaborski, RS ;
Lure, FYM .
MEDICAL PHYSICS, 1998, 25 (07) :1132-1138
[3]
High-fidelity electronic display of digital radiographs [J].
Flynn, MJ ;
Kanicki, J ;
Badano, A ;
Eyler, WR .
RADIOGRAPHICS, 1999, 19 (06) :1653-1669
[4]
Cathode ray tube quality control and acceptance testing program: Initial results for clinical PACS displays [J].
Groth, DS ;
Bernatz, SN ;
Fetterly, KA ;
Hangiandreou, NJ .
RADIOGRAPHICS, 2001, 21 (03) :719-732
[5]
Effects of luminance and resolution on observer performance with chest radiographs [J].
Herron, JM ;
Bender, TM ;
Campbell, WL ;
Sumkin, JH ;
Rockette, HE ;
Gur, D .
RADIOLOGY, 2000, 215 (01) :169-174
[6]
ISHIGAKI T, 2001, REPORT SCI RES MINIS, P4
[7]
INFLUENCE OF CRT WORKSTATION ON OBSERVERS PERFORMANCE [J].
ITOH, Y ;
ISHIGAKI, T ;
SAKUMA, S ;
HIROSE, M ;
FUKATSU, H ;
ITOH, S ;
HORIKAWA, Y ;
SHIMAMOTO, K ;
TADOKORO, T ;
IKEDA, M ;
ITOH, K .
COMPUTER METHODS AND PROGRAMS IN BIOMEDICINE, 1992, 37 (04) :253-258
[8]
Influence of film and monitor display luminance on observer performance and visual search [J].
Krupinski, E ;
Roehrig, H ;
Furukawa, T .
ACADEMIC RADIOLOGY, 1999, 6 (07) :411-418
[9]
The influence of a perceptually linearized display on observer performance and visual search [J].
Krupinski, EA ;
Roehrig, H .
ACADEMIC RADIOLOGY, 2000, 7 (01) :8-13
[10]
Detection of subtle lung nodules: Relative influence of quantum and anatomic noise on chest radiographs [J].
Samei, E ;
Flynn, MJ ;
Eyler, WR .
RADIOLOGY, 1999, 213 (03) :727-734