The double K+/Ca2+ sensor based on laser scanned silicon transducer (LSST) for multi-component analysis

被引:20
作者
Ermolenko, Y
Yoshinobu, T
Mourzina, Y
Furuichi, K
Levichev, S
Schöning, MJ
Vlasov, Y
Iwasaki, H
机构
[1] St Petersburg State Univ, Fac Chem, St Petersburg 199034, Russia
[2] Osaka Univ, Inst Sci & Ind Res, Ibaraki, Osaka 5670047, Japan
[3] Res Ctr Julich, Inst Thin Films & Interfaces, D-52425 Julich, Germany
[4] Univ Appl Sci Aachen, Div Julich, D-52428 Julich, Germany
关键词
laser scanned semiconductor transducer (LSST); double sensor; potassium and calcium selectivity; solvent-polymeric membrane;
D O I
10.1016/S0039-9140(02)00620-3
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
In the present work a double ion sensor based on a laser scanned semiconductor transducer (LSST) for the simultaneous determination of K+- and Ca2+-ions in solutions has been developed. Specially elaborated ion-sensitive membrane compositions based on valinomycin and calcium ionophore calcium bis[4-(1,1,3,3-tetramethylbutyl)phenyl] phosphate (t-HDOPP-Ca) were deposited as separate layers on a silanized surface of the Si/SiO2/Si3N4-transducer. The proposed multi-sensor exhibits theoretical sensitivities and the detection limits of the sensor were found to be 2 x 10(-6) mol 1(-1) for K+ and 5 x 10(-6) Mol 1(-1) for Ca2+. The elaborated double sensor is proposed for the first time as a prototype of a new type of multi-sensor systems for chemical analysis. (C) 2002 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:785 / 795
页数:11
相关论文
共 37 条
[1]  
AKADOV JY, 1972, DIELECTRIC PROPERTIE, P64
[2]  
AKKER E, 1997, CHEM REV, V97, P3080
[3]   DETERMINATION OF COMPLEX-FORMATION CONSTANTS OF NEUTRAL CATION-SELECTIVE IONOPHORES IN SOLVENT POLYMERIC MEMBRANES [J].
BAKKER, E ;
WILLER, M ;
LERCHI, M ;
SEILER, K ;
PRETSCH, E .
ANALYTICAL CHEMISTRY, 1994, 66 (04) :516-521
[4]   Carrier-based ion-selective electrodes and bulk optodes. 1. General characteristics [J].
Bakker, E ;
Buhlmann, P ;
Pretsch, E .
CHEMICAL REVIEWS, 1997, 97 (08) :3083-3132
[6]   EMF RESPONSE OF NEUTRAL-CARRIER BASED ION-SENSITIVE FIELD-EFFECT TRANSISTORS WITH MEMBRANES FREE OF IONIC SITES [J].
BUHLMANN, P ;
YAJIMA, S ;
TOHDA, K ;
UMEZAWA, Y .
ELECTROCHIMICA ACTA, 1995, 40 (18) :3021-3027
[7]   ELECTROCHEMICAL PERFORMANCE, BIOCOMPATIBILITY, AND ADHESION OF NEW POLYMER MATRICES FOR SOLID-STATE ION SENSORS [J].
CHA, GS ;
LIU, D ;
MEYERHOFF, ME ;
CANTOR, HC ;
MIDGLEY, AR ;
GOLDBERG, HD ;
BROWN, RB .
ANALYTICAL CHEMISTRY, 1991, 63 (17) :1666-1672
[8]   Photocurable membranes for ion-selective light-addressable potentiometric sensor [J].
Ermolenko, Y ;
Yoshinobu, T ;
Mourzina, Y ;
Levichev, S ;
Furuichi, K ;
Vlasov, Y ;
Schöning, MJ ;
Iwasaki, H .
SENSORS AND ACTUATORS B-CHEMICAL, 2002, 85 (1-2) :79-85
[9]   Lithium sensor based on the laser scanning semiconductor transducer [J].
Ermolenko, Y ;
Yoshinobu, T ;
Mourzina, Y ;
Furuichi, K ;
Levichev, S ;
Vlasov, Y ;
Schöning, MJ ;
Iwasaki, H .
ANALYTICA CHIMICA ACTA, 2002, 459 (01) :1-9
[10]  
ERMOLENKO Y, 2001, P 5 E AS C CHEM SENS, P66