Energy- and intensity-modulated electron beams for radiotherapy

被引:91
作者
Ma, CM [1 ]
Pawlicki, T [1 ]
Lee, MC [1 ]
Jiang, SB [1 ]
Li, JS [1 ]
Deng, J [1 ]
Yi, B [1 ]
Mok, E [1 ]
Boyer, AL [1 ]
机构
[1] Stanford Univ, Sch Med, Dept Radiat Oncol, Stanford, CA 94305 USA
关键词
D O I
10.1088/0031-9155/45/8/316
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
This work investigates the feasibility of optimizing energy- and intensity-modulated electron beams for radiation therapy. A multileaf collimator (MLC) specially designed for modulated electron radiotherapy (MERT) was investigated both experimentally and by Monte Carlo simulations: An inverse-planning system based on Monte Carlo dose calculations was developed to optimize electron beam energy and intensity to achieve dose conformity for target volumes near the surface. The results showed that an MLC with 5 mm leaf widths could produce complex held shapes for MERT. Electron intra- and inter-leaf leakage had negligible effects on the dose distributions delivered with the MLC, even at shallow depths. Focused leaf ends reduced the electron scattering contributions to the dose compared with straight leaf ends. As anticipated, moving the MLC position toward the patient surface reduced the penumbra significantly. There were significant differences in the beamlet distributions calculated by an analytic 3-D pencil beam algorithm and the Monte Carlo method. The Monte Carlo calculated beamlet distributions were essential to the accuracy of the MERT dose distribution in cases involving large air gaps, oblique incidence and heterogeneous treatment targets (at the tissue-bone and bone-lung interfaces). To demonstrate the potential of MERT for target dose coverage and normal tissue sparing for treatment of superficial targets, treatment plans For a hypothetical treatment were compared using photon beams and MERT.
引用
收藏
页码:2293 / 2311
页数:19
相关论文
共 58 条
[51]   ELECTRON-BEAM DOSE DISTRIBUTIONS NEAR STANDARD INHOMOGENEITIES [J].
SHORTT, KR ;
ROSS, CK ;
BIELAJEW, AF ;
ROGERS, DWO .
PHYSICS IN MEDICINE AND BIOLOGY, 1986, 31 (03) :235-249
[52]  
Takahashi S., 1965, Acta Radiol Diagn Suppl, V242, P1
[53]   A patient-specific Monte Carlo dose-calculation method for photon beams [J].
Wang, L ;
Chui, CS ;
Lovelock, M .
MEDICAL PHYSICS, 1998, 25 (06) :867-878
[55]  
WEBB S, 1997, PHYSICS CONFORMAL RA
[56]   A METHOD FOR IMPLEMENTING DYNAMIC PHOTON-BEAM INTENSITY MODULATION USING INDEPENDENT JAWS AND A MULTILEAF COLLIMATOR [J].
YU, CX ;
SYMONS, MJ ;
DU, MN ;
MARTINEZ, AA ;
WONG, JW .
PHYSICS IN MEDICINE AND BIOLOGY, 1995, 40 (05) :769-787
[57]   Matching of electron beams for conformal therapy of target volumes at moderate depths [J].
Zackrisson, B ;
Karlsson, M .
RADIOTHERAPY AND ONCOLOGY, 1996, 39 (03) :261-270
[58]   Monte Carlo investigation of electron beam output factors versus size of square cutout [J].
Zhang, GG ;
Rogers, DWO ;
Cygler, JE ;
Mackie, TR .
MEDICAL PHYSICS, 1999, 26 (05) :743-750