Blood pressure in children aged 4-8 years: comparison of Omron HEM 711 and sphygmomanometer blood pressure measurements

被引:11
作者
Midgley, P. C. [1 ]
Wardhaugh, B. [2 ]
Macfarlane, C. [2 ]
Magowan, R. [2 ]
Kelnar, C. J. H. [2 ]
机构
[1] Univ Edinburgh, Dept Reprod & Dev Med, Edinburgh, Midlothian, Scotland
[2] Royal Hosp Sick Children, Lothian Univ Hosp NHS Trust, Edinburgh EH9 1LF, Midlothian, Scotland
关键词
MEASURING DEVICES; VALIDATION; PROTOCOL;
D O I
10.1136/adc.2008.137059
中图分类号
R72 [儿科学];
学科分类号
100202 ;
摘要
Objective: To collect normal data on blood pressure (BP) in healthy children aged 4-8 and to compare measurements of BP made in the same subjects with a sphygmomanometer and a portable automated oscillometric BP monitor (Omron HEM 711 with child cuff). Methods: Cross-sectional observational study of 764 children. BP measurements were made at school, using both a sphygmomanometer and an Omron HEM 711. Immediately after the BP measurement children were asked to state which device they preferred (if any). Results: Children had no preference for whether the sphygmomanometer or the Omron was used. Bland-Altman plots showed a lack of consistency between the two methods of BP measurement. With systolic BP there was a trend for the Omron to underestimate when low and overestimate when high. Conclusions: Children were equally distributed in their preference for BP device. There was a wide variation between the two methods of BP measurement, which suggests that comparison of automated BP measurements with normative data obtained by sphygmomanometer is not valid.
引用
收藏
页码:955 / 958
页数:4
相关论文
共 10 条
  • [1] [Anonymous], 1987, Pediatrics, V79, P1, DOI 10.1542/peds.79.1.1
  • [2] Artigao L M, 2000, Aten Primaria, V25, P96
  • [3] Validation of the Omron 705IT (HEM-759-E) oscillometric blood pressure monitoring device according to the British Hypertension Society protocol
    Coleman, A
    Freeman, P
    Steel, S
    Shennan, A
    [J]. BLOOD PRESSURE MONITORING, 2006, 11 (01) : 27 - 32
  • [4] BLOOD-PRESSURE IN 1ST 10 YEARS OF LIFE - THE BROMPTON STUDY
    DESWIET, M
    FAYERS, P
    SHINEBOURNE, EA
    [J]. BRITISH MEDICAL JOURNAL, 1992, 304 (6818) : 23 - 26
  • [5] Falkner B, 1996, PEDIATRICS, V98, P649
  • [6] Working Group on Blood Pressure Monitoring of the European Society of Hypertension International Protocol for validation of blood pressure measuring devices in adults
    O'Brien, E
    Pickering, T
    Asmar, R
    Myers, M
    Parati, G
    Staessen, J
    Mengden, T
    Imai, Y
    Waeber, B
    Palatini, P
    Atkins, N
    Gerin, W
    [J]. BLOOD PRESSURE MONITORING, 2002, 7 (01) : 3 - 17
  • [7] Ambulatory blood pressure in schoolchildren
    O'Sullivan, JJ
    Derrick, G
    Griggs, P
    Foxall, R
    Aitkin, M
    Wren, C
    [J]. ARCHIVES OF DISEASE IN CHILDHOOD, 1999, 80 (06) : 529 - 532
  • [8] THE BRITISH-HYPERTENSION-SOCIETY PROTOCOL FOR THE EVALUATION OF AUTOMATED AND SEMIAUTOMATED BLOOD-PRESSURE MEASURING DEVICES WITH SPECIAL REFERENCE TO AMBULATORY SYSTEMS
    OBRIEN, E
    PETRIE, J
    LITTLER, W
    DESWIET, M
    PADFIELD, PL
    OMALLEY, K
    JAMIESON, M
    ALTMAN, D
    BLAND, M
    ATKINS, N
    [J]. JOURNAL OF HYPERTENSION, 1990, 8 (07) : 607 - 619
  • [9] COMPARISON OF OSCILLOMETRIC AND INTRAARTERIAL BLOOD-PRESSURE AND PULSE MEASUREMENT
    RITHALIA, SVS
    EDWARDS, D
    [J]. JOURNAL OF MEDICAL ENGINEERING & TECHNOLOGY, 1994, 18 (05) : 179 - 181
  • [10] Validation of the Omron 705 IT oscillometric device for home blood pressure measurement in children and adolescents: The Arsakion School Study
    Stergiou, George S.
    Yiannes, Nikolaos G.
    Rarra, Vayia C.