Comparison of 5 and 10 mm thick HgI2 pixelated γ-ray spectrometers
被引:33
作者:
Baciak, JE
[1
]
He, Z
[1
]
机构:
[1] Univ Michigan, Ann Arbor, MI 48109 USA
来源:
关键词:
mercuric iodide;
room temperature;
compound semiconductor;
gamma-ray detectors;
D O I:
10.1016/S0168-9002(03)01049-0
中图分类号:
TH7 [仪器、仪表];
学科分类号:
0804 ;
080401 ;
081102 ;
摘要:
HgI2 detectors fabricated with pixelated anodes were tested as potential room temperature spectrometers. Spectra were obtained with 5 and 10 mm thick devices, and their results were compared. The 5 mm thick devices showed some effect of the anode weighting potential, while the 10 mm thick devices showed the effects of charge diffusion/sharing and electron trapping. A resolution near 1.4% FWHM using a Cs-137 source was achieved after correcting for the interaction depth within the detector. The resolution from individual depths was as low as approximately 1.1% FWHM. (C) 2003 Elsevier Science B.V. All rights reserved.