Validation of automated contour analysis of intravascular ultrasound images after vascular intervention

被引:3
作者
Hartlooper, A
van Essen, JA
van der Lugt, A
van Sambeek, MRHM
van Overhagen, H
Gussenhoven, EJ
机构
[1] Erasmus Univ Ee 2312, NL-3000 DR Rotterdam, Netherlands
[2] Univ Hosp Dijkzigt, Dept Cardiol, NL-3015 GD Rotterdam, Netherlands
[3] Univ Hosp Dijkzigt, Dept Radiol, NL-3015 GD Rotterdam, Netherlands
[4] Univ Hosp Dijkzigt, Dept Vasc Surg, NL-3015 GD Rotterdam, Netherlands
关键词
D O I
10.1016/S0741-5214(98)99998-0
中图分类号
R61 [外科手术学];
学科分类号
摘要
Purpose: The purpose of this study was to determine the feasibility of automated contour analysis of intravascular ultrasound images obtained after vascular intervention. Study design: This was a descriptive study. Methods: Intravascular ultrasound images obtained from patients after balloon angioplasty (n = 10), stent (n = 10), or stent graft placement (n = 10) were analyzed. A comparison was made between lumen area measured with an automated and a manual system. The location showing the smallest lumen area derived from the automated system was compared with the smallest lumen area selected by visual estimation. Results: Images containing a dissection as a result of balloon angioplasty could not be analyzed by the automated system. The coefficient of variation between the lumen area measurements obtained with the automated system and the manual tracing system of images with a stent (n = 76) or stent graft (n = 79) was 2.7% and 2.1%, respectively. Correlation between the two systems was high (r = 1.00, p < 0.01) both for images containing stents or stent grafts. Minimum lumen area measured with the automated analysis system was smaller than minimum lumen area selected by visual estimation (mean difference 0.8 mm(2) (4.9%) for stents and 2.4 mm(2) (10.9%) for stent grafts). The location of the smallest lumen area determined with both systems was the same (<1 cm) in 16 cases and differed more than 1 cm in 4 other cases. Conclusions: The automated analysis system shows good agreement with manual contour analysis of lumen area in images with a stent or stent graft and is a reliable tool for determination of the smallest lumen area. The system is not able to analyze an irregular shaped lumen area caused by a dissection.
引用
收藏
页码:486 / 491
页数:6
相关论文
共 12 条
[1]  
[Anonymous], 1994, P COMP CARD
[2]   MECHANISMS AND OUTCOMES OF ANGIOPLASTY AND ATHERECTOMY ASSESSED BY INTRAVASCULAR ULTRASOUND IMAGING [J].
FITZGERALD, PJ ;
YOCK, PG .
JOURNAL OF CLINICAL ULTRASOUND, 1993, 21 (09) :579-588
[3]   ARTERIAL-WALL CHARACTERISTICS DETERMINED BY INTRAVASCULAR ULTRASOUND IMAGING - AN INVITRO STUDY [J].
GUSSENHOVEN, EJ ;
ESSED, CE ;
LANCEE, CT ;
MASTIK, F ;
FRIETMAN, P ;
VANEGMOND, FC ;
REIBER, J ;
BOSCH, H ;
VANURK, H ;
ROELANDT, J ;
BOM, N .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1989, 14 (04) :947-952
[4]   Temporal correlation of blood scattering signals in vivo from radiofrequency intravascular ultrasound [J].
Li, W ;
vanderSteen, AFW ;
Lancee, CT ;
Honkoop, J ;
Gussenhoven, EJ ;
Bom, N .
ULTRASOUND IN MEDICINE AND BIOLOGY, 1996, 22 (05) :583-590
[5]   Intravascular ultrasound predictors of restenosis after percutaneous transcatheter coronary revascularization [J].
Mintz, GS ;
Popma, JJ ;
Pichard, AD ;
Kent, KM ;
Satler, LF ;
Chuang, YC ;
Griffin, J ;
Leon, MB .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1996, 27 (07) :1678-1687
[6]   Reliability and reproducibility of automated contour analysis in intravascular ultrasound images of femoropopliteal arteries [J].
van der Lugt, A ;
Hartlooper, A ;
van Essen, JA ;
Li, WG ;
von Birgelen, C ;
Reiber, JHC ;
Gussenhoven, EJ .
ULTRASOUND IN MEDICINE AND BIOLOGY, 1998, 24 (01) :43-50
[7]   COMPARISON OF INTRAVASCULAR ULTRASONIC FINDINGS AFTER CORONARY BALLOON ANGIOPLASTY EVALUATED IN-VITRO WITH HISTOLOGY [J].
VANDERLUGT, A ;
GUSSENHOVEN, EJ ;
STIJNEN, T ;
VANSTRIJEN, M ;
VANDRIEL, E ;
VANEGMOND, FC ;
SUYLEN, RJ ;
VANURK, H .
AMERICAN JOURNAL OF CARDIOLOGY, 1995, 76 (10) :661-666
[8]   A NEW APPROACH FOR THE QUANTIFICATION OF COMPLEX LESION MORPHOLOGY - THE GRADIENT FIELD TRANSFORM - BASIC PRINCIPLES AND VALIDATION RESULTS [J].
VANDERZWET, PMJ ;
REIBER, JHC .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1994, 24 (01) :216-224
[9]   Computerized assessment of coronary lumen and atherosclerotic plaque dimensions in three-dimensional intravascular ultrasound correlated with histomorphometry [J].
vonBirgelen, C ;
vanderLugt, A ;
Nicosia, A ;
Mintz, GS ;
Gussenhoven, EJ ;
deVrey, E ;
Mallus, MT ;
Roelandt, JRTC ;
Serruys, PW ;
deFeyter, PJ .
AMERICAN JOURNAL OF CARDIOLOGY, 1996, 78 (11) :1202-1209
[10]   Morphometric analysis in three-dimensional intracoronary ultrasound: An in vitro and in vivo study performed with a novel system for the contour detection of lumen and plaque [J].
vonBirgelen, C ;
DiMario, C ;
Li, WG ;
Schuurbiers, JCH ;
Slager, CJ ;
deFeyter, PJ ;
Roelandt, JRTC ;
Serruys, PW .
AMERICAN HEART JOURNAL, 1996, 132 (03) :516-527