Influence of Na, K, Ca and Mg on lead atomization by tungsten coil atomic absorption spectrometry

被引:16
作者
de Oliveira, PV
Krug, FJ
Silva, MM
Nóbrega, JA
Queiroz, ZF
Rocha, FRP
机构
[1] Univ Sao Paulo, Inst Quim, BR-05513970 Sao Paulo, Brazil
[2] Univ Sao Paulo, Ctr Energia Nucl Na Agr, BR-13400970 Piracicaba, SP, Brazil
[3] Univ Fed Rio Grande Sul, Dept Quim, BR-91501970 Porto Alegre, RS, Brazil
[4] Univ Fed Sao Carlos, Dept Quim, BR-13565905 Sao Carlos, SP, Brazil
[5] Univ Sao Paulo, Inst Quim Sao Carlos, BR-13560970 Sao Carlos, SP, Brazil
关键词
lead atomization; atomic absorption spectrometry; tungsten coil atomizer; interference;
D O I
10.1590/S0103-50532000000200006
中图分类号
O6 [化学];
学科分类号
0703 ;
摘要
The atomization of lead in an electrothermal tungsten coil atomizer in the presence and absence of Na+, K+, Ca2+ and Mg2+ was investigated with the objective of understanding the interference processes. The lead atomization was less affected by Ca2+ and Mg2+ than by Na+ and K+. In the absence of concomitants, lead atomization efficiency was improved by the presence of H-2 (10% v/v) in the purge gas composition, during pyrolysis and atomization steps. The interference caused by Na+ and Ca2+ was negligible when the pyrolysis step was accomplished without H-2 in the purge gas composition The results showed that Na+, K+, Ca2+ and Mg2+ are directly involved in competition reactions for H-2 in condensed phase.
引用
收藏
页码:136 / 142
页数:7
相关论文
共 30 条
[1]  
[Anonymous], 1982, METAL VAPOURS FLAMES
[2]   Determination of cadmium with a portable, battery-powered tungsten coil atomic absorption spectrometer [J].
Batchelor, JD ;
Thomas, SE ;
Jones, BT .
APPLIED SPECTROSCOPY, 1998, 52 (08) :1086-1091
[3]   SIMPLE LOW-COST TUNGSTEN-COIL ATOMIZER FOR ELECTROTHERMAL ATOMIC-ABSORPTION SPECTROMETRY [J].
BERNDT, H ;
SCHALDACH, G .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1988, 3 (05) :709-712
[4]   ANALYTICAL EVALUATION OF A TUNGSTEN COIL ATOMIZER FOR CADMIUM, LEAD, CHROMIUM, MANGANESE, NICKEL AND COBALT DETERMINATION BY ELECTROTHERMAL ATOMIC-ABSORPTION SPECTROMETRY [J].
BRUHN, CG ;
AMBIADO, FE ;
CID, HJ ;
WOERNER, R ;
TAPIA, J ;
GARCIA, R .
ANALYTICA CHIMICA ACTA, 1995, 306 (2-3) :183-192
[5]  
Bruhn CG, 1996, QUIM ANAL, V15, P191
[6]   Chemical modifiers in a tungsten coil electrothermal atomizer. Part 1. Determination of lead in hair and blood [J].
Bruhn, CG ;
Neira, JY ;
Valenzuela, GD ;
Nobrega, JA .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1998, 13 (01) :29-35
[7]   Determination of cadmium in hair and blood by tungsten coil electrothermal atomic absorption spectrometry with chemical modifiers [J].
Bruhn, CG ;
Neira, JY ;
Valenzuela, GD ;
Nóbrega, JA .
TALANTA, 1999, 48 (03) :537-549
[8]  
BRUHN CG, 1998, QUIMICA IND, V37
[9]  
BRUHN CG, IN PRESS FRESENIUS J
[10]   CHEMICAL MODIFICATION BY ASCORBIC-ACID AND OXALIC-ACID IN GRAPHITE-FURNACE ATOMIC-ABSORPTION SPECTROMETRY [J].
BYRNE, JP ;
CHAKRABARTI, CL ;
GILCHRIST, GFR ;
LAMOUREUX, MM ;
BERTELS, P .
ANALYTICAL CHEMISTRY, 1993, 65 (09) :1267-1272