Multislice CT angiography: Optimization of scan parameters in a vascular phantom.

被引:9
作者
Funke, C [1 ]
Alamo, L [1 ]
Castillo, E [1 ]
Kopka, L [1 ]
Grabbe, E [1 ]
机构
[1] Univ Gottingen, Abt Rontgendiagnost 1, Klinikum Georg August, D-37075 Gottingen, Germany
来源
ROFO-FORTSCHRITTE AUF DEM GEBIET DER RONTGENSTRAHLEN UND DER BILDGEBENDEN VERFAHREN | 2000年 / 172卷 / 11期
关键词
multislice-helical CT; CT angiography; vascular stenoses; vascular phantom; scan parameters;
D O I
10.1055/s-2000-8373
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: The aim of this study is to determine the optimal scan parameters for the evaluation of experimental vascular stenoses with a multislice-helical CT. Material and Methods: A vascular phantom consisting of four tubes with an inner diameter of 8 mm and with experimental stenoses of 50%, 75% and 90% was scanned in different tube orientations using a multislice-CT scanner (LightSpeed QX/i, GE, Milwaukee, USA). Examinations were performed with increasing collimations (1.25-5mm), tube currents (100-300mA) and two different table speeds (0.75 HQ mode and 1.5 HS mode). Results: The most exact measurements were obtained in tubes angulated parallel to the scan direction with a collimation of 2.5 mm in the HQ mode (7.5 mm/rot.). An almost equivalent accuracy was obtained in the HS mode (15 mm/rot.) with a collimation of 2.5 mm when higher tube currents (300 mA) were employed. The degree of stenoses was overestimated when the tube was angulated perpendicular to the z-axis. Conclusion: Multislice-CT provides a good detection rate of vascular stenoses especially at 0 degrees and also at 45 degrees angulation in the HQ mode. The use of the HS mode with higher tube currents allows scanning of longer distances with almost identical accuracy.
引用
收藏
页码:927 / 933
页数:7
相关论文
共 16 条
[1]   Vitro examinations for detection of vascular stenoses using spiral CT angiography. [J].
Beilicke, M ;
Kloppel, R ;
Lieberenz, S .
ROFO-FORTSCHRITTE AUF DEM GEBIET DER RONTGENSTRAHLEN UND DER BILDGEBENDEN VERFAHREN, 1998, 168 (03) :217-221
[2]   TECHNICAL OPTIMIZATION OF SPIRAL CT FOR DEPICTION OF RENAL-ARTERY STENOSIS - IN-VITRO ANALYSIS [J].
BRINK, JA ;
LIM, JT ;
WANG, G ;
HEIKEN, JP ;
DEYOE, LA ;
VANNIER, MW .
RADIOLOGY, 1995, 194 (01) :157-163
[3]   RENAL ARTERIAL STENOSES - SPIRAL CT ANGIOGRAPHY [J].
GALANSKI, M ;
PROKOP, M ;
CHAVAN, A ;
SCHAEFER, CM ;
JANDELEIT, K ;
NISCHELSKY, JE .
RADIOLOGY, 1993, 189 (01) :185-192
[4]   Multi-slice helical CT: Scan and reconstruction [J].
Hu, H .
MEDICAL PHYSICS, 1999, 26 (01) :5-18
[5]  
Kalender W A, 1994, Aktuelle Radiol, V4, P287
[6]   Subsecond multi-slice computed tomography: basics and applications [J].
Klingenbeck-Regn, K ;
Schaller, S ;
Flohr, T ;
Ohnesorge, B ;
Kopp, AF ;
Baum, U .
EUROPEAN JOURNAL OF RADIOLOGY, 1999, 31 (02) :110-124
[7]  
Kopka L, 1999, RADIOLOGE, V39, P971, DOI 10.1007/s001170050590
[8]   Influence of helical CT parameters on spatial resolution in CT angiography performed with a subsecond scanner [J].
Luboldt, W ;
Weber, R ;
Seemann, M ;
Desantis, M ;
Reiser, M .
INVESTIGATIVE RADIOLOGY, 1999, 34 (06) :421-426
[9]  
Ohnesorge B, 1999, RADIOLOGE, V39, P923, DOI 10.1007/s001170050583
[10]  
PROKOP M, 1993, RADIOLOGE, V33, P694