Post-reconstruction filtering of positron emission tomography whole-body emission images and attenuation maps using nonlinear diffusion filtering

被引:9
作者
Demirkaya, O [1 ]
机构
[1] King Faisal Specialist Hosp & Res Ctr, Dept Biostat Epidemiol & Sci Comp, Riyadh 11211, Saudi Arabia
关键词
positron emission tomography (PET); whole-body images; anisotropic diffusion filtering;
D O I
10.1016/j.acra.2004.07.012
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Rationale and Objectives. Positron emission tomography has been playing an important role as a quantitative molecular imaging modality to measure and image biochemical processes in vivo. The quality of positron emission tomography images may impact the accuracy of the quantitative or semiquantitative information extracted from them. Materials and Methods. In this study, the anisotropic diffusion filtering technique was used to denoise emission images and attenuation maps of a whole body for improving the quantitative accuracy of emission images. The efficacy of the filtering technique was shown on a hot-lesion phantom as well as on whole-body emission images and attenuation maps. Results. The filtering technique showed a superb performance on images reconstructed using the iterative and the filtered back-projection reconstruction techniques. Conclusion. Filtering may allow the reconstruction of images with optimal parameters with respect to lesion contrast and image resolution by removing the consequential noise. An improvement of as much as six-fold may be attained. It may also improve the accuracy of the quantitative information, such as standard uptake value extracted from emission images. (C) AUR, 2004.
引用
收藏
页码:1105 / 1114
页数:10
相关论文
共 21 条
[11]  
KEYES JW, 1995, J NUCL MED, V36, P1836
[12]  
Lartizien C, 2004, J NUCL MED, V45, P714
[13]   LOCAL GEOMETRY VARIABLE CONDUCTANCE DIFFUSION FOR POSTRECONSTRUCTION FILTERING [J].
LUO, DS ;
KING, MA ;
GLICK, S .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1994, 41 (06) :2800-2806
[14]   THRESHOLD SELECTION METHOD FROM GRAY-LEVEL HISTOGRAMS [J].
OTSU, N .
IEEE TRANSACTIONS ON SYSTEMS MAN AND CYBERNETICS, 1979, 9 (01) :62-66
[15]   SCALE-SPACE AND EDGE-DETECTION USING ANISOTROPIC DIFFUSION [J].
PERONA, P ;
MALIK, J .
IEEE TRANSACTIONS ON PATTERN ANALYSIS AND MACHINE INTELLIGENCE, 1990, 12 (07) :629-639
[16]   Nonlinear multiscale analysis of three-dimensional echocardiographic sequences [J].
Sarti, A ;
Mikula, K ;
Sgallari, F .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 1999, 18 (06) :453-466
[17]   Efficient and reliable schemes for nonlinear diffusion filtering [J].
Weickert, J ;
Romeny, BMT ;
Viergever, MA .
IEEE TRANSACTIONS ON IMAGE PROCESSING, 1998, 7 (03) :398-410
[18]  
WEICKERT J., 1998, Anisotropic Diffusion in Image Processing
[19]   PERFORMANCE EVALUATION OF THE POSITRON SCANNER ECAT EXACT [J].
WIENHARD, K ;
ERIKSSON, L ;
GROOTOONK, S ;
CASEY, M ;
PIETRZYK, U ;
HEISS, WD .
JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 1992, 16 (05) :804-813
[20]  
XU EZ, 1991, J NUCL MED, V32, P161