Photon beam relative dose validation of the DPM Monte Carlo code in lung-equivalent media

被引:60
作者
Chetty, IJ [1 ]
Charland, PM
Tyagi, N
McShan, DL
Fraass, BA
Bielajew, AF
机构
[1] Univ Michigan, Dept Radiat Oncol, Ann Arbor, MI 48109 USA
[2] Univ Michigan, Dept Nucl Engn, Ann Arbor, MI 48109 USA
关键词
DPM Monte Carlo code; phase space; BEAMnrc; lateral electronic disequilibrium; ion chamber measurements;
D O I
10.1118/1.1555671
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Validation experiments have been conducted using 6 and 15 MV photons in inhomogeneous (water/ lung/water) media to benchmark the accuracy of the DPM Monte Carlo code for photon beam dose calculations. Small field sizes (down to 2 X 2 cm(2)) and low-density media were chosen for this investigation because the intent was to test the DPM code under conditions where lateral electronic disequilibrium effects are emphasized. The treatment head components of a Varian 21EX linear accelerator, including the jaws (defining field sizes of 2X2, 3X3 and 10X10 cm(2)), were simulated using the BEAMnrc code. The phase space files were integrated within the DPM code system, and central axis depth dose and profile calculations were compared against diode measurements in a homogeneous water phantom in order to validate the phase space. Results of the homogeneous phantom study indicated that the relative differences between DPM calculations and measurements were within +/-1% (based on the rms deviation) for the depth dose curves; relative profile dose differences were on average within +/-1%/1 mm. Depth dose and profile measurements were carried out using an ion-chamber and film, within an inhomogeneous phantom consisting of a 6 cm slab of lung-equivalent material embedded within solid water. For the inhomogeneous phantom experiment, DPM depth dose calculations were within +/-1% (based on the rms deviation) of measurements; relative profile differences at depths within and beyond the lung were, on average, within +/-2% in the inner and outer beam regions, and within 1-2 mm distance-to-agreement within the penumbral region. Relative point differences on the order of 2-3% were within the estimated experimental uncertainties. This work demonstrates that the DPM Monte Carlo code is capable of accurate photon beam dose calculations in situations where lateral electron disequilibrium effects are pronounced. (C) 2003 American Association of Physicists in Medicine.
引用
收藏
页码:563 / 573
页数:11
相关论文
共 36 条
[11]   Fast Monte Carlo dose calculation for photon beams based on the VMC electron algorithm [J].
Fippel, M .
MEDICAL PHYSICS, 1999, 26 (08) :1466-1475
[12]   A multiple source model for 6 MV photon beam dose calculations using Monte Carlo [J].
Fix, MK ;
Stampanoni, M ;
Manser, P ;
Born, EJ ;
Mini, R ;
Rüegsegger, P .
PHYSICS IN MEDICINE AND BIOLOGY, 2001, 46 (05) :1407-1427
[13]   American Association of Physicists in Medicine radiation therapy committee task group 53: Quality assurance for clinical radiotherapy treatment planning [J].
Fraass, B ;
Doppke, K ;
Hunt, M ;
Kutcher, G ;
Starkschall, G ;
Stern, R ;
Van Dyke, J .
MEDICAL PHYSICS, 1998, 25 (10) :1773-1829
[14]   Accurate condensed history Monte Carlo simulation of electron transport. I. EGSnrc, the new EGS4 version [J].
Kawrakow, I .
MEDICAL PHYSICS, 2000, 27 (03) :485-498
[15]   3D electron dose calculation using a Voxel based Monte Carlo algorithm (VMC) [J].
Kawrakow, I ;
Fippel, M ;
Friedrich, K .
MEDICAL PHYSICS, 1996, 23 (04) :445-457
[16]   Super-Monte Carlo: A 3-D electron beam dose calculation algorithm [J].
Keall, PJ ;
Hoban, PW .
MEDICAL PHYSICS, 1996, 23 (12) :2023-2034
[17]   MONTE-CARLO MODEL OF PHOTON BEAMS USED IN RADIATION-THERAPY [J].
LOVELOCK, DMJ ;
CHUI, CS ;
MOHAN, R .
MEDICAL PHYSICS, 1995, 22 (09) :1387-1394
[18]   Clinical implementation of a Monte Carlo treatment planning system [J].
Ma, CM ;
Mok, E ;
Kapur, A ;
Pawlicki, T ;
Findley, D ;
Brain, S ;
Forster, K ;
Boyer, AL .
MEDICAL PHYSICS, 1999, 26 (10) :2133-2143
[19]  
MA CM, 1998, 509B NRCC PIRS
[20]   A CONVOLUTION METHOD OF CALCULATING DOSE FOR 15-MV X-RAYS [J].
MACKIE, TR ;
SCRIMGER, JW ;
BATTISTA, JJ .
MEDICAL PHYSICS, 1985, 12 (02) :188-196