Weighing of biomolecules, single cells and single nanoparticles in fluid

被引:940
作者
Burg, Thomas P.
Godin, Michel
Knudsen, Scott M.
Shen, Wenjiang
Carlson, Greg
Foster, John S.
Babcock, Ken
Manalis, Scott R.
机构
[1] MIT, Dept Biol Engn, Cambridge, MA 02139 USA
[2] MIT, Dept Mech Engn, Cambridge, MA 02139 USA
[3] Innovat Micro Technol, Santa Barbara, CA 93117 USA
[4] Affin Biosensors, Santa Barbara, CA 93117 USA
关键词
D O I
10.1038/nature05741
中图分类号
O [数理科学和化学]; P [天文学、地球科学]; Q [生物科学]; N [自然科学总论];
学科分类号
07 ; 0710 ; 09 ;
摘要
Nanomechanical resonators enable the measurement of mass with extraordinary sensitivity(1-7). Previously, samples as light as 7 zeptograms ( 1 zg = 10(-21) g) have been weighed in vacuum, and proton-level resolution seems to be within reach(8). Resolving small mass changes requires the resonator to be light and to ring at a very pure tone - that is, with a high quality factor(9). In solution, viscosity severely degrades both of these characteristics, thus preventing many applications in nanotechnology and the life sciences where fluid is required(10). Although the resonant structure can be designed to minimize viscous loss, resolution is still substantially degraded when compared to measurements made in air or vacuum(11-14). An entirely different approach eliminates viscous damping by placing the solution inside a hollow resonator that is surrounded by vacuum(15,16). Here we demonstrate that suspended microchannel resonators can weigh single nanoparticles, single bacterial cells and sub-monolayers of adsorbed proteins in water with sub-femtogram resolution ( 1 Hz bandwidth). Central to these results is our observation that viscous loss due to the fluid is negligible compared to the intrinsic damping of our silicon crystal resonator. The combination of the low resonator mass ( 100 ng) and high quality factor ( 15,000) enables an improvement in mass resolution of six orders of magnitude over a high-end commercial quartz crystal microbalance(17). This gives access to intriguing applications, such as mass-based flow cytometry, the direct detection of pathogens, or the non-optical sizing and mass density measurement of colloidal particles.
引用
收藏
页码:1066 / 1069
页数:4
相关论文
共 30 条
[1]   Counting single native biomolecules and intact viruses with color-coded nanoparticles [J].
Agrawal, A ;
Zhang, CY ;
Byassee, T ;
Tripp, RA ;
Nie, SM .
ANALYTICAL CHEMISTRY, 2006, 78 (04) :1061-1070
[2]   ANALYSIS OF CELL-SIZE AND DNA CONTENT IN EXPONENTIALLY GROWING AND STATIONARY-PHASE BATCH CULTURES OF ESCHERICHIA-COLI [J].
AKERLUND, T ;
NORDSTROM, K ;
BERNANDER, R .
JOURNAL OF BACTERIOLOGY, 1995, 177 (23) :6791-6797
[3]   Next generation of protein microarray support materials:: Evaluation for protein and antibody microarray applications [J].
Angenendt, P ;
Glökler, J ;
Sobek, J ;
Lehrach, H ;
Cahill, DJ .
JOURNAL OF CHROMATOGRAPHY A, 2003, 1009 (1-2) :97-104
[4]   A label-free immunosensor array using single-chain antibody fragments [J].
Backmann, N ;
Zahnd, C ;
Huber, F ;
Bietsch, A ;
Plückthun, A ;
Lang, HP ;
Güntherodt, HJ ;
Hegner, M ;
Gerber, C .
PROCEEDINGS OF THE NATIONAL ACADEMY OF SCIENCES OF THE UNITED STATES OF AMERICA, 2005, 102 (41) :14587-14592
[5]   Micromechanical mass sensors for biomolecular detection in a physiological environment [J].
Braun, T ;
Barwich, V ;
Ghatkesar, MK ;
Bredekamp, AH ;
Gerber, C ;
Hegner, M ;
Lang, HP .
PHYSICAL REVIEW E, 2005, 72 (03)
[6]   Vacuum-packaged suspended microchannel resonant mass sensor for biomolecular detection [J].
Burg, Thomas P. ;
Mirza, Amir R. ;
Milovic, Nebojsa ;
Tsau, Christine H. ;
Popescu, George A. ;
Foster, John S. ;
Manalis, Scott R. .
JOURNAL OF MICROELECTROMECHANICAL SYSTEMS, 2006, 15 (06) :1466-1476
[7]   Suspended microchannel resonators for biomolecular detection [J].
Burg, TP ;
Manalis, SR .
APPLIED PHYSICS LETTERS, 2003, 83 (13) :2698-2700
[8]   Ultimate limits to inertial mass sensing based upon nanoelectromechanical systems [J].
Ekinci, KL ;
Yang, YT ;
Roukes, ML .
JOURNAL OF APPLIED PHYSICS, 2004, 95 (05) :2682-2689
[9]   Silicon tube structures for a fluid-density sensor [J].
Enoksson, P ;
Stemme, G ;
Stemme, E .
SENSORS AND ACTUATORS A-PHYSICAL, 1996, 54 (1-3) :558-562
[10]   Ultrasensitive mass sensor fully integrated with complementary metal-oxide-semiconductor circuitry [J].
Forsen, E ;
Abadal, G ;
Ghatnekar-Nilsson, S ;
Teva, J ;
Verd, J ;
Sandberg, R ;
Svendsen, W ;
Perez-Murano, F ;
Esteve, J ;
Figueras, E ;
Campabadal, F ;
Montelius, L ;
Barniol, N ;
Boisen, A .
APPLIED PHYSICS LETTERS, 2005, 87 (04)