Orientational analysis of the Achilles tendon and enthesis using an ultrashort echo time spectroscopic imaging sequence

被引:55
作者
Du, Jiang [1 ]
Chiang, Alan Jing-Tzyh [1 ]
Chung, Christine B. [1 ]
Statum, Sheronda [1 ]
Znamirowski, Richard [1 ]
Takahashi, Atsushi [2 ]
Bydder, Graeme M. [1 ]
机构
[1] Univ Calif San Diego, Dept Radiol, San Diego, CA 92103 USA
[2] GE Healthcare, Global Appl Sci Lab, Menlo Pk, CA 94025 USA
关键词
Ultrashort TE; Spectroscopic imaging; Short-T2; Tendon; Enthesis; Magic angle; Bulk susceptibility; UTE PULSE SEQUENCES; MAGNETIC-RESONANCE; PROJECTION RECONSTRUCTION; CORTICAL BONE; ANGIOGRAPHY; EXCITATION; ANISOTROPY; COLLAGEN; TISSUES; FIELD;
D O I
10.1016/j.mri.2009.06.002
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
100231 [临床病理学]; 100902 [航空航天医学];
摘要
Tendons and entheses are magnetic resonance (MR) "invisible" when imaged with conventional clinical pulse sequences. When the highly ordered, collagen-rich fibers in tendons and entheses are placed at the magic angle, dipolar interactions are decreased and their T2s are often considerably increased. The bulk magnetic susceptibility of tendons and entheses also varies with orientation to B-0, leading to a direction-dependent resonance frequency shift. Ultrashort echo time (UTE) sequences with a minimum TE of 8 mu s provide high signal from both tendons and entheses. The combination of a UTE sequence with an interleaved undersampled variable TE acquisition scheme provides a new approach for fast spectroscopic imaging of short T2 tissues. This UTE spectroscopic imaging (UTESI) technique provides quantitative information including T2*, chemical shift and resonance frequency shift due to bulk susceptibility effect. In this article, the orientational effects on tendons and entheses were investigated using a UTESI sequence on a clinical 3-T scanner. T2* was found to increase fivefold for tendons and twofold for entheses due to the magic angle effect. A resonance frequency shift up to 1.2 ppm was observed for both tendons and entheses due to the bulk susceptibility effect when their orientation was changed from 0 degrees to 90 degrees relative to B-0. (C) 2010 Elsevier Inc. All rights reserved.
引用
收藏
页码:178 / 184
页数:7
相关论文
共 32 条
[1]
Effects of gradient anisotropy in MRI [J].
Aldefeld, B ;
Börnert, P .
MAGNETIC RESONANCE IN MEDICINE, 1998, 39 (04) :606-614
[2]
Magnetic resonance imaging of entheses using ultrashort TE (UTE) pulse sequences [J].
Benjamin, Michael ;
Bydder, Graeme M. .
JOURNAL OF MAGNETIC RESONANCE IMAGING, 2007, 25 (02) :381-389
[3]
LUNG PARENCHYMA - PROJECTION RECONSTRUCTION MR-IMAGING [J].
BERGIN, CJ ;
PAULY, JM ;
MACOVSKI, A .
RADIOLOGY, 1991, 179 (03) :777-781
[4]
NMR CHEMICAL-SHIFT IMAGING IN 3 DIMENSIONS [J].
BROWN, TR ;
KINCAID, BM ;
UGURBIL, K .
PROCEEDINGS OF THE NATIONAL ACADEMY OF SCIENCES OF THE UNITED STATES OF AMERICA-BIOLOGICAL SCIENCES, 1982, 79 (11) :3523-3526
[5]
Carl M, 2009, P 17 ANN M ISMRM HON, P1991
[6]
VARIABLE-RATE SELECTIVE EXCITATION [J].
CONOLLY, S ;
NISHIMURA, D ;
MACOVSKI, A ;
GLOVER, G .
JOURNAL OF MAGNETIC RESONANCE, 1988, 78 (03) :440-458
[7]
Contrast-enhanced peripheral magnetic resonance angiography using time-resolved vastly undersampled isotropic projection reconstruction [J].
Du, J ;
Carroll, TJ ;
Brodsky, E ;
Lu, AM ;
Grist, TM ;
Mistretta, CA ;
Block, WF .
JOURNAL OF MAGNETIC RESONANCE IMAGING, 2004, 20 (05) :894-900
[8]
Time-resolved, undersampled projection reconstruction imaging for high-resolution CE-MRA of the distal runoff vessels [J].
Du, J ;
Carroll, TJ ;
Wagner, HJ ;
Vigen, K ;
Fain, SB ;
Block, WF ;
Korosec, FR ;
Grist, TM ;
Mistretta, CA .
MAGNETIC RESONANCE IN MEDICINE, 2002, 48 (03) :516-522
[9]
Two-dimensional ultrashort echo time imaging using a spiral trajectory [J].
Du, Jiang ;
Bydder, Mark ;
Takahashi, Atsushi M. ;
Chung, Christine B. .
MAGNETIC RESONANCE IMAGING, 2008, 26 (03) :304-312
[10]
Ultrashort echo time spectroscopic imaging (UTESI) of cortical bone [J].
Du, Jiang ;
Hamilton, Gavin ;
Takahashi, Atsushi ;
Bydder, Mark ;
Chung, Christine B. .
MAGNETIC RESONANCE IN MEDICINE, 2007, 58 (05) :1001-1009