Atmospheric pressure MALDI/ion trap mass spectrometry

被引:164
作者
Laiko, VV
Moyer, SC
Cotter, RJ [1 ]
机构
[1] Johns Hopkins Univ, Sch Med, Dept Pharmacol & Mol Sci, Baltimore, MD 21205 USA
[2] Johns Hopkins Univ, Dept Chem, Baltimore, MD 21218 USA
关键词
D O I
10.1021/ac000530d
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A new sample ionization technique, atmospheric pressure matrix-assisted laser desorption/ionization (AP MALDI), was coupled with a commercial ion trap mass spectrometer. This configuration enables the application-specific selection of external atmospheric ionization sources: the electrospray/APCI (commercially available) and AP MALDI (built in-house), which can be readily interchanged within minutes. The detection limit of the novel AP MALDI/ion trap is 10-50 fmol of analyte deposited on the target surface for a four-component mixture of peptides with 800-1700 molecular weight. The possibility of peptide structural analysis by MS/MS and MS3 experiments for AP MALDI-generated ions was demonstrated for the first time.
引用
收藏
页码:5239 / 5243
页数:5
相关论文
共 11 条
[1]   CONTRIBUTIONS OF MASS-SPECTROMETRY TO PEPTIDE AND PROTEIN-STRUCTURE [J].
BIEMANN, K .
BIOMEDICAL AND ENVIRONMENTAL MASS SPECTROMETRY, 1988, 16 (1-12) :99-111
[2]   The quadrupole ion trap mass spectrometer - A small solution to a big challenge [J].
Jonscher, KR ;
Yates, JR .
ANALYTICAL BIOCHEMISTRY, 1997, 244 (01) :1-15
[3]  
LAIKO V, 1999, 47 ASMS C MASS SPECT
[4]   Atmospheric pressure matrix assisted laser desorption/ionization mass spectrometry [J].
Laiko, VV ;
Baldwin, MA ;
Burlingame, AL .
ANALYTICAL CHEMISTRY, 2000, 72 (04) :652-657
[5]  
LAIKO VV, 1999, Patent No. 5965884
[6]  
LAIKO VV, UNPUB
[7]   Characterization of an improved electrodynamic ion funnel interface for electrospray ionization mass spectrometry [J].
Shaffer, SA ;
Tolmachev, A ;
Prior, DC ;
Anderson, GA ;
Udseth, HR ;
Smith, RD .
ANALYTICAL CHEMISTRY, 1999, 71 (15) :2957-2964
[8]  
Shaffer SA, 1997, RAPID COMMUN MASS SP, V11, P1813
[9]   An ion funnel interface for improved ion focusing and sensitivity using electrospray ionization mass spectrometry [J].
Shaffer, SA ;
Prior, DC ;
Anderson, GA ;
Udseth, HR ;
Smith, RD .
ANALYTICAL CHEMISTRY, 1998, 70 (19) :4111-4119
[10]  
Wolfender JL, 1999, J MASS SPECTROM, V34, P447, DOI 10.1002/(SICI)1096-9888(199904)34:4<447::AID-JMS801>3.0.CO