Potential for Personal Digital Assistant interference with implantable cardiac devices

被引:19
作者
Tri, JL
Trusty, JM
Hayes, DL
机构
[1] Mayo Clin, Coll Med, Div Fdn Telecommun & Network Serv, Rochester, MN 55905 USA
[2] Mayo Clin, Coll Med, Div Cardiovasc Dis & Internal Med, Rochester, MN 55905 USA
关键词
D O I
10.4065/79.12.1527
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
OBJECTIVE: To determine whether the wireless local area network (WLAN) technology, specifically the Personal Digital Assistant (PDA), Interferes with implantable cardiac pacemakers and defibrillators. MATERIAL AND METHODS: Various pacemakers and defibrillators were tested in vitro at the Mayo Clinic in Rochester, Minn, between March 6 and July 30, 2003. These cardiac devices were exposed to,an HP Compaq iPAQ PDA fitted with a Cisco Aironet WILAN card. Initial testing was designed to show whether the Aironet card radiated energy In a consistent pattern from the antenna of the PDA to ensure that subsequent cardiac device testing would not be affected by the orientation of the PDA to the cardiac device. Testing involved placing Individual cardiac devices in a simulator and uniformly exposing, each device at its most sensitive programmable value to the WILAN card set to maximum power. RESULTS: During testing with the Cisco WILAN Aironet card, all devices programmed to the unipolar or bipolar configuration single- or dual-chamber mode had normal pacing and sensing functions and exhibited no effects of electromagnetic Interference except for 1 Implantable cardioverter-defibrillator (ICD). This aberration was determined to relate to the design of the Investigators' testing apparatus and not to the output of the PDA. The ICD device appropriately identified and labeled the electromagnetic aberration as "noise." CONCLUSIONS: We documented no electromagnetic Interference caused by the WLAN technology by using In vitro testing of pacemakers and ICDs; however, testing Ideally should be completed In vivo to confirm the lack of any clinically Important interactions.
引用
收藏
页码:1527 / 1530
页数:4
相关论文
共 8 条
[1]  
*AM NAT STAND I, 2000, ACT IMPL MED DEV EL
[2]   Digital cellular telephone interaction with implantable cardioverter-defibrillators [J].
Fetter, JG ;
Ivans, V ;
Benditt, DG ;
Collins, J .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1998, 31 (03) :623-628
[3]   Interactions between electronic article surveillance systems and implantable cardioverter-defibrillators [J].
Groh, WJ ;
Boschee, SA ;
Engelstein, ED ;
Miles, WM ;
Burton, ME ;
Foster, PR ;
Crevey, BJ ;
Zipes, DP .
CIRCULATION, 1999, 100 (04) :387-392
[4]   Possible electromagnetic interference with electronic medical equipment by radio waves coming from outside the hospital [J].
Hanada E. ;
Kodama K. ;
Takano K. ;
Watanabe Y. ;
Nose Y. .
Journal of Medical Systems, 2001, 25 (4) :257-267
[5]   Interference with cardiac pacemakers by cellular telephones [J].
Hayes, DL ;
Wang, PJ ;
Reynolds, DW ;
Estes, NAM ;
Griffith, JL ;
Steffens, RA ;
Carlo, GL ;
Findlay, GK ;
Johnson, CM .
NEW ENGLAND JOURNAL OF MEDICINE, 1997, 336 (21) :1473-1479
[6]  
*IEEE, 2003, 80211B9999 IEEE
[7]   Study of pacemaker and implantable cardioverter defibrillator triggering by electronic article surveillance devices (SPICED TEAS) [J].
McIvor, ME ;
Reddinger, J ;
Floden, E ;
Sheppard, RC ;
Johnson, D ;
Becker, GI ;
Mayotte, M .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1998, 21 (10) :1847-1861
[8]   Cellular phone interference with external cardiopulmonary monitoring devices [J].
Tri, JL ;
Hayes, DL ;
Smith, TT ;
Severson, RP .
MAYO CLINIC PROCEEDINGS, 2001, 76 (01) :11-15