Quantifying the predictability of diaphragm motion during respiration with a noninvasive external marker

被引:260
作者
Vedam, SS
Kini, VR
Keall, PJ
Ramakrishnan, V
Mostafavi, H
Mohan, R
机构
[1] Virginia Commonwealth Univ, Dept Biomed Engn, Richmond, VA 23298 USA
[2] Virginia Commonwealth Univ, Dept Radiat Oncol, Richmond, VA USA
[3] Virginia Commonwealth Univ, Dept Biostat, Richmond, VA USA
[4] Varian Med Syst Inc, Ginzton Technol Ctr, Palo Alto, CA USA
[5] MD Anderson Canc Ctr, Dept Radiat Phys, Houston, TX USA
关键词
D O I
10.1118/1.1558675
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The aim of this work was to quantify the ability to predict intrafraction diaphragm motion from an external respiration signal during a course of radiotherapy. The data obtained included diaphragm motion traces from 63 fluoroscopic lung procedures for 5 patients, acquired simultaneously with respiratory motion signals (an infrared camera-based system was used to track abdominal wall motion). During these sessions, the patients were asked to breathe either (i) without instruction, (ii) with audio prompting, or (iii) using visual feedback. A statistical general linear model was formulated to describe the relationship between the respiration signal and diaphragm motion over all sessions and for all breathing training types. The model parameters derived from the first session for each patient were then used to predict the diaphragm motion for subsequent sessions based on the respiration signal. Quantification of the difference between the predicted and actual motion during each session determined our ability to predict diaphragm motion during a course of radiotherapy. This measure of diaphragm motion was also used to estimate clinical target volume (CTV) to planning target volume (PTV) margins for conventional, gated, and proposed four-dimensional (4D) radiotherapy. Results from statistical analysis indicated a strong linear relationship between the respiration signal and diaphragm motion (p<0.001) over all sessions, irrespective of session number (p=0.98) and breathing training type (p=0.19). Using model parameters obtained from the first session, diaphragm motion was predicted in subsequent sessions to within 0.1 cm (1 sigma) for gated and 4D radiotherapy. Assuming a 0.4 cm setup error, superior-inferior CTV-PTV margins of 1.1 cm for conventional radiotherapy could be reduced to 0.8 cm for gated and 4D radiotherapy. The diaphragm motion is strongly correlated with the respiration signal obtained from the abdominal wall. This correlation can be used to predict diaphragm motion, based on the respiration signal, to within 0.1 cm (1 a) over a course of radiotherapy. (C) 2003 American Association of Physicists in Medicine.
引用
收藏
页码:505 / 513
页数:9
相关论文
共 56 条
  • [1] [Anonymous], 1993, 50 ICRU
  • [2] Armstrong J., 1995, REED NEWS, V3, P6
  • [3] Improvement of CT-based treatment-planning models of abdominal targets using static exhale imaging
    Balter, JM
    Lam, KL
    McGinn, CJ
    Lawrence, TS
    Ten Haken, RK
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1998, 41 (04): : 939 - 943
  • [4] Daily targeting of intrahepatic tumors for radiotherapy
    Balter, JM
    Brock, KK
    Litzenberg, DW
    McShan, DL
    Lawrence, TS
    Ten Haken, R
    McGinn, CJ
    Lam, KL
    Dawson, LA
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2002, 52 (01): : 266 - 271
  • [5] Determination of ventilatory liver movement via radiographic evaluation of diaphragm position
    Balter, JM
    Dawson, LA
    Kazanjian, S
    McGinn, C
    Brock, KK
    Lawrence, T
    Ten Haken, R
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2001, 51 (01): : 267 - 270
  • [6] Dosimetric evaluation of lung tumor immobilization using breath hold at deep inspiration
    Barnes, EIA
    Murray, BR
    Robinson, DM
    Underwood, LJ
    Hanson, J
    Roa, WHY
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2001, 50 (04): : 1091 - 1098
  • [7] Effects of intra-fraction motion on IMRT dose delivery: statistical analysis and simulation
    Bortfeld, T
    Jokivarsi, K
    Goitein, M
    Kung, J
    Jiang, SB
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2002, 47 (13) : 2203 - 2220
  • [8] Fluoroscopic study of tumor motion due to breathing: Facilitating precise radiation therapy for lung cancer patients
    Chen, QS
    Weinhous, MS
    Deibel, FC
    Ciezki, JP
    Macklis, RM
    [J]. MEDICAL PHYSICS, 2001, 28 (09) : 1850 - 1856
  • [9] ULTRASOUND QUANTITATION OF RESPIRATORY ORGAN MOTION IN THE UPPER ABDOMEN
    DAVIES, SC
    HILL, AL
    HOLMES, RB
    HALLIWELL, M
    JACKSON, PC
    [J]. BRITISH JOURNAL OF RADIOLOGY, 1994, 67 (803) : 1096 - 1102
  • [10] The reproducibility of organ position using active breathing control (ABC) during liver radiotherapy
    Dawson, LA
    Brock, KK
    Kazanjian, S
    Fitch, D
    McGinn, CJ
    Lawrence, TS
    Ten Haken, RK
    Balter, J