Effects of the 14F Hemopump on coronary hemodynamics in patients undergoing high-risk coronary angioplasty

被引:4
作者
Dubois-Randé, JL [1 ]
Teiger, E [1 ]
Garot, J [1 ]
Aptecar, E [1 ]
Pernès, JM [1 ]
Tixier, D [1 ]
Gueret, P [1 ]
Loisance, D [1 ]
Dupouy, P [1 ]
机构
[1] Hop Henri Mondor, Federat Cardiol, Serv Explorat Fonct, F-94010 Creteil, France
关键词
D O I
10.1016/S0002-8703(98)70044-4
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background The influence of the 14F Hemopump on coronary hemodynamics in patients with coronary artery disease remains unknown. Methods Systemic and coronary hemodynamic measurements were obtained in eight patients among 13 who underwent high-risk coronary angioplasty in our institution with the support of the Hemopump. Coronary blood flow velocity was measured with a 0.014-inch Doppler-tipped guide wire both proximal and distal to the target lesion. Results Angioplasty decreases the diameter coronary stenosis from 76% +/- 21% to 22% +/- 11%. Hemopump support did not change systemic hemodynamics either before or after angioplasty. During angioplasty Hemopump support decreased the pulmonary capillary wedge pressure from 23.5 +/- 8.5 mm Hg to 18.6 +/- 7 mm Hg (p = 0.013). No changes in either heart rate, mean and systolic aortic pressures, and cardiac index were observed throughout the procedure. After successful angioplasty was performed, the ratio of proximal to distal flow velocity decreased from 2.11 +/- 1 to 1.65 +/- 0.2 (p = 0.05). However, Hemopump did not affect absolute coronary blood flow velocities or the phasic pattern of flow velocities (diastolic systolic velocity ratio, diastolic and systolic velocity integrals) either in proximal or distal locations either before or after angioplasty. Conclusions This study shows that although the 14F Hemopump produces unloading of the left ventricle, it does not importantly alter coronary hemodynamics when systemic hemodynamics are stable. Whether the Hemopump would maintain or improve coronary blood flow in compromised patients remains to be determined.
引用
收藏
页码:844 / 849
页数:6
相关论文
共 9 条
[1]   VALIDATION OF A DOPPLER GUIDE WIRE FOR INTRAVASCULAR MEASUREMENT OF CORONARY-ARTERY FLOW VELOCITY [J].
DOUCETTE, JW ;
CORL, PD ;
PAYNE, HM ;
FLYNN, AE ;
GOTO, M ;
NASSI, M ;
SEGAL, J .
CIRCULATION, 1992, 85 (05) :1899-1911
[2]  
DUBOISRANDE JL, 1994, ASAIO T, V40, P486
[3]   AUGMENTATION OF CORONARY BLOOD-FLOW BY INTRAAORTIC BALLOON PUMPING IN PATIENTS AFTER CORONARY ANGIOPLASTY [J].
KERN, MJ ;
AGUIRRE, F ;
BACH, R ;
DONOHUE, T ;
SIEGEL, R ;
SEGAL, J .
CIRCULATION, 1993, 87 (02) :500-511
[4]   PROPHYLACTIC INTRAVENTRICULAR PUMPING IN HIGH-RISK CORONARY ANGIOPLASTY [J].
LOISANCE, D ;
DUBOISRANDE, JL ;
DELEUZE, P ;
OKUDE, J ;
ROSENVAL, O ;
GESCHWIND, H .
LANCET, 1990, 335 (8687) :438-440
[5]   METHODS OF MEASUREMENT OF MYOCARDIAL BLOOD-FLOW IN PATIENTS - A CRITICAL-REVIEW [J].
MARCUS, ML ;
WILSON, RF ;
WHITE, CW .
CIRCULATION, 1987, 76 (02) :245-253
[6]   MECHANICAL LEFT-VENTRICULAR UNLOADING DURING HIGH-RISK CORONARY ANGIOPLASTY - FIRST USE OF A NEW PERCUTANEOUS TRANSVALVULAR LEFT-VENTRICULAR ASSIST DEVICE [J].
SCHOLZ, KH ;
FIGULLA, HR ;
SCHWEDA, F ;
SMALLING, RW ;
HELLIGE, G ;
KREUZER, H ;
ABOULHOSN, W ;
WAMPLER, RK .
CATHETERIZATION AND CARDIOVASCULAR DIAGNOSIS, 1994, 31 (01) :61-69
[7]   PROTECTIVE EFFECTS OF THE HEMOPUMP LEFT-VENTRICULAR ASSIST DEVICE IN EXPERIMENTAL CARDIOGENIC-SHOCK [J].
SCHOLZ, KH ;
HERING, JP ;
SCHRODER, T ;
UHLIG, P ;
KREUZER, H ;
TEBBE, U ;
FERRARI, M ;
HELLIGE, G .
EUROPEAN JOURNAL OF CARDIO-THORACIC SURGERY, 1992, 6 (04) :209-214
[8]   IMPROVED REGIONAL MYOCARDIAL BLOOD-FLOW, LEFT-VENTRICULAR UNLOADING, AND INFARCT SALVAGE USING AN AXIAL-FLOW, TRANSVALVULAR LEFT-VENTRICULAR ASSIST DEVICE - A COMPARISON WITH INTRAAORTIC BALLOON COUNTERPULSATION AND REPERFUSION ALONE IN A CANINE INFARCTION MODEL [J].
SMALLING, RW ;
CASSIDY, DB ;
BARRETT, R ;
LACHTERMAN, B ;
FELLI, P ;
AMIRIAN, J .
CIRCULATION, 1992, 85 (03) :1152-1159
[9]  
Wampler R K, 1988, ASAIO Trans, V34, P450