X-ray imaging spectrometers for Astro-E: Ground calibration in soft X-ray range

被引:8
作者
Hayashida, K [1 ]
Kitamoto, S [1 ]
Miyata, E [1 ]
Tsunemi, H [1 ]
Yoshita, E [1 ]
Kohmura, T [1 ]
Mori, K [1 ]
Katayama, K [1 ]
Katayama, H [1 ]
Shouho, M [1 ]
Ohta, M [1 ]
Dotani, T [1 ]
Ozaki, M [1 ]
Koyama, K [1 ]
Awaki, H [1 ]
Tsuru, TG [1 ]
Ricker, G [1 ]
Bautz, MW [1 ]
Foster, R [1 ]
Kissel, S [1 ]
机构
[1] Osaka Univ, Dept Earth & Space Sci, Toyonaka, Osaka 5600043, Japan
来源
X-RAY OPTICS, INSTRUMENTS, AND MISSIONS III | 2000年 / 4012卷
关键词
Astro-E; X-ray CCD; calibration;
D O I
10.1117/12.391548
中图分类号
V [航空、航天];
学科分类号
08 ; 0825 ;
摘要
Soft X-ray response of X-ray Imaging Spectrometers (XIS) for the Astro-E satellite is measured with a grating spectrometer system at Osaka. First, relation between incident X-ray energy and output pulse height peak (E-PH relation) is examined with an SX grating. It is found that jump in the E-PH relation around Si-K edge is at most 2.7 eV. Second, quantum efficiency (QE) of the XIS in 0.4-2.2 keV range is measured relatively to the reference CCD of which absolute QE was calibrated with a gas proportional counter. The QE is-fitted with a model in which CCD gate structures are considered. Systematic error on the QE results is estimated by referring an independent measurement. Third, tuning and improvement of the response function is performed. We employ six components to reproduce the response profile of the XIS. In this paper, improvement of one component which is originated in the events absorbed in the channel-stop is presented. Neverthless, Astro-E was lost due to the launch failure. We overview the XIS project in its flight model phase, modified points of the design, problems and solutions etc., in order to be utilized in a possible recovery of the satellite.
引用
收藏
页码:123 / 136
页数:14
相关论文
共 20 条
[1]   X-ray CCD calibration for the AXAF CCD imaging spectrometer [J].
Bautz, M ;
Pivovaroff, M ;
Baganoff, F ;
Isobe, T ;
Jones, S ;
Kissel, S ;
LaMarr, B ;
Manning, H ;
Prigozhin, G ;
Ricker, G ;
Nousek, J ;
Grant, C ;
Nishikida, K ;
Scholze, F ;
Thornagel, R ;
Ulm, G .
X-RAY OPTICS, INSTRUMENTS, AND MISSIONS, 1998, 3444 :210-224
[2]  
BAUTZ MW, 1999, SCI INSTRUMENT SI C
[3]   THE X-RAY-ENERGY RESPONSE OF SILICON .A. THEORY [J].
FRASER, GW ;
ABBEY, AF ;
HOLLAND, A ;
MCCARTHY, K ;
OWENS, A ;
WELLS, A .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1994, 350 (1-2) :368-378
[4]   Continuum x-ray source as a calibration system for charge coupled devices [J].
Hashimotodani, K ;
Toneri, T ;
Kitamoto, S ;
Tsunemi, H ;
Hayashida, K ;
Miyata, E ;
Katayama, K ;
Kohomura, T ;
Asakura, R ;
Koyama, K ;
Yamamoto, K ;
Miyaguchi, K ;
Suzuki, H .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1998, 69 (02) :392-395
[5]   Soft X-ray response of the proto-type CCD camera (XIS) for Astro-E [J].
Hayashida, K ;
Kitamoto, S ;
Miyata, E ;
Tsunemi, H ;
Hashimotodani, K ;
Katayama, K ;
Kohmura, T ;
Asakura, R ;
Yoshita, K ;
Katayama, H ;
Shouho, M ;
Koyoma, K ;
Tsuru, TG ;
Awaki, H ;
Dotani, T ;
Ozaki, M ;
Ricker, G ;
Doty, J ;
Bautz, M ;
Kissel, S ;
Foster, R .
EUV, X-RAY, AND GAMMA-RAY INSTRUMENTATION FOR ASTRONOMY IX, 1998, 3445 :278-290
[6]  
IMANISHI K, 2000, P SPIE, V4012
[7]   Quantum efficiency of the CCD camera (XIS) for the ASTRO-E mission [J].
Katayama, H ;
Shouho, M ;
Kohmura, T ;
Katayama, K ;
Yoshita, K ;
Tsunemi, H ;
Kitamoto, S ;
Hayashida, K ;
Miyata, E ;
Hashimotodani, K ;
Koyama, K ;
Ricker, G ;
Bautz, MW ;
Foster, R ;
Kissel, S .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1999, 436 (1-2) :74-78
[8]   Calibration of the x-ray CCD camera (XIS) for the ASTRO-E in the soft x-ray band [J].
Katayama, H ;
Shouho, M ;
Kohmura, T ;
Katayama, K ;
Ohta, M ;
Tsunemi, H ;
Kitamoto, S ;
Hayashida, K ;
Miyata, E ;
Yoshita, K ;
Koyama, K ;
Ricker, G ;
Bautz, MW ;
Foster, R ;
Kissel, S ;
Doty, J .
EUV, X-RAY, AND GAMMA-RAY INSTRUMENTATION FOR ASTRONOMY X, 1999, 3765 :721-728
[9]   Reports of soft X-ray response of the engineering model of the CCD camera (XIS) for the ASTRO-E mission [J].
Katayama, K ;
Kohmura, T ;
Katayama, H ;
Shouho, M ;
Tsunemi, H ;
Kitamoto, S ;
Hayashida, K ;
Miyata, E ;
Hashimotodani, K ;
Koyama, K ;
Ricker, G ;
Bautz, M ;
Foster, R .
BROAD BAND X-RAY SPECTRA OF COSMIC SOURCES, 2000, 25 (3-4) :881-884
[10]   Quantum efficiency measurement of the X-ray CCD camera (XIS) for the ASTRO-E mission in the soft X-ray band [J].
Kohmura, T ;
Katayama, K ;
Katayama, H ;
Shouho, M ;
Tsunemi, H ;
Kitamoto, S ;
Hayashida, K ;
Miyata, E ;
Ohtani, M ;
Ohta, M ;
Yoshita, K ;
Koyama, K ;
Ricker, G ;
Bautz, MW ;
Foster, R ;
Kissel, S .
EUV, X-RAY, AND GAMMA-RAY INSTRUMENTATION FOR ASTRONOMY X, 1999, 3765 :588-596