Declotting of embolized temporary vena cava filter by ultrasound and the Angiojet -: Comparative experimental in vitro studies

被引:11
作者
Dick, A
Neuerburg, J
Schmitz-Rode, T
Alliger, H
Günther, R
机构
[1] Univ Technol Aachen, Dept Diagnost Radiol, D-52057 Aachen, Germany
[2] SUNY Farmingdale, Farmingdale, NY USA
关键词
ultrasound; ultrasound thrombolysis; Angiojet; vena cava filter; in vitro study;
D O I
10.1097/00004424-199802000-00006
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
RATIONALE AND OBJECTIVES. The authors perform an in vitro evaluation of the thrombolytic efficacy and the amount of "downstream" embolization induced by two new mechanical thrombectomy devices when applied to clots trapped in a temporary vena cava filter. METHODS. The first device used was a 22.5-kHz prototype intravascular ultrasound device with a flexible 0.8-mm (.032-inch) titanium 2-mm ball-tipped wire probe ensheathed in a 7-French teflon guide catheter. The device was inserted through a 10-French steering catheter. Under fluoroscopic control, ultrasound energy (26 +/- 4 watts/cm(2), maximal longitudinal catheter tip amplitude 54 mu m) was applied to 10 Ultravist-filled porcine thrombi (mean, 3500 mg). The second device, the Angiojet catheter, was applied to five Ultravist-filled porcine thrombi (mean, 3640 mg). The thrombi were treated while trapped in a temporary Gunther vena cava filter (Cook Europe, Bjaverskov, Demnark) mounted in a vena cava flow model. The resultant "downstream" emboli were trapped in two tandem filters of decreasing pore size and weighed. RESULTS. Mean thrombus dissolution rate was 53% +/- 22% standard deviation (SD) for the ultrasound device (n = 10) and 63% +/- 8% SD for the Angiojet (n = 5) (difference statistically significant at P = 0.03). For the ultrasound device, the mean embolic particle weight caught by the filters with mesh widths of 1 mm and 0.1 mm was 42% +/- 14% SD and 4% +/- 2% SD, respectively, of the initial thrombus weight. For the Angiojet, the respective numbers were 35% +/- 16% SD and 3% +/- 1% SD. Mean treatment time was 216 +/- 45 seconds SD for the ultrasound device and 153 +/- 21 seconds SD for the Angiojet. CONCLUSIONS. The thrombolytic efficacy of the Angiojet was significantly greater and the treatment time was significantly shorter than that of the ultrasound device. Both systems had a high embolization rate.
引用
收藏
页码:91 / 97
页数:7
相关论文
共 28 条
[1]   DISSOLUTION OF PERIPHERAL ARTERIAL THROMBI BY ULTRASOUND [J].
ARIANI, M ;
FISHBEIN, MC ;
CHAE, JS ;
SADEGHI, H ;
MICHAEL, TAD ;
DUBIN, SB ;
SIEGEL, RJ .
CIRCULATION, 1991, 84 (04) :1680-1688
[2]   Comparative in vitro study of two percutaneous hydrodynamic thrombectomy systems [J].
Bucker, A ;
SchmitzRode, T ;
Vorwerk, D ;
Gunther, RW .
JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 1996, 7 (03) :445-449
[3]   VENA TECH-LGM FILTER - LONG-TERM RESULTS OF A PROSPECTIVE-STUDY [J].
CROCHET, DP ;
STORA, O ;
FERRY, D ;
GROSSETETE, R ;
LEURENT, B ;
BRUNEL, P ;
NGUYEN, JM .
RADIOLOGY, 1993, 188 (03) :857-860
[4]   SHORT-TERM PROPHYLAXIS OF PULMONARY-EMBOLISM BY USING A RETRIEVABLE VENA-CAVA FILTER [J].
DARCY, MD ;
SMITH, TP ;
HUNTER, DW ;
CASTANEDAZUNIGA, W ;
LUND, G ;
AMPLATZ, K .
AMERICAN JOURNAL OF ROENTGENOLOGY, 1986, 147 (04) :836-838
[5]   Thrombolysis of mural thrombus by ultrasound -: An experimental in vitro study [J].
Dick, A ;
Neuerburg, J ;
Schmitz-Rode, T ;
Alliger, H ;
Schmid-Schönbein, H ;
Günther, RW .
INVESTIGATIVE RADIOLOGY, 1998, 33 (02) :85-90
[6]   ABILITY OF HIGH-INTENSITY ULTRASOUND TO ABLATE HUMAN ATHEROSCLEROTIC PLAQUES AND MINIMIZE DEBRIS SIZE [J].
ERNST, A ;
SCHENK, EA ;
GRACEWSKI, SM ;
WOODLOCK, TJ ;
MURANT, FG ;
ALLIGER, H ;
MELTZER, RS .
AMERICAN JOURNAL OF CARDIOLOGY, 1991, 68 (02) :242-246
[7]   FEASIBILITY OF RECANALIZATION OF HUMAN CORONARY-ARTERIES USING HIGH-INTENSITY ULTRASOUND [J].
ERNST, A ;
SCHENK, EA ;
WOODLOCK, TJ ;
ALLIGER, H ;
GOTTLIEB, S ;
CHILD, SZ ;
MELTZER, RS .
AMERICAN JOURNAL OF CARDIOLOGY, 1994, 73 (02) :126-132
[8]  
ERNST A, 1989, CIRCULATION S2, V80, P1220
[9]   PERCUTANEOUS INFERIOR VENA-CAVAL FILTERS - FOLLOW-UP OF 7 DESIGNS IN 320 PATIENTS [J].
FERRIS, EJ ;
MCCOWAN, TC ;
CARVER, DK ;
MCFARLAND, DR .
RADIOLOGY, 1993, 188 (03) :851-856
[10]  
GREENFIELD LJ, 1992, ARCH SURG-CHICAGO, V127, P969