Improvement in the quality of the cardiac vein images by optimizing the scan protocol of multidetector-row computed tomography

被引:5
作者
Hara, Tetsuya [2 ,3 ]
Yamashiro, Kohei [1 ,2 ]
Okajima, Katsunori [2 ]
Hayashi, Takatoshi [2 ]
Kajiya, Teishi [2 ]
机构
[1] Toyohashi Heart Ctr, Dept Cardiol, Aichi 4418530, Japan
[2] Himeji Cardiovasc Ctr, Dept Cardiol, Himeji, Hyogo 6700981, Japan
[3] Kobe Univ, Grad Sch, Div Cardiol, Dept Internal Med,Chuo Ku, Kobe, Hyogo 6500017, Japan
关键词
Multidetector-row computed tomography; Coronary veins; Cardiac resynchronization therapy; Scan protocol; CORONARY-SINUS; HEART-FAILURE; VENOUS SYSTEM; RESYNCHRONIZATION THERAPY; ANGIOGRAPHY; VISUALIZATION; ABLATION; ARTERIES;
D O I
10.1007/s00380-008-1142-x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
The present study aimed at optimizing the scan protocol for multidetector-row computed tomography (MDCT) to adequately visualize coronary veins. Circulation time (Cir.T) was defined as the time period from the injection of contrast media into the coronary artery to the pervasion of the contrast media into the coronary sinus as observed by coronary angiography. We investigated the relation between the Cir.T and echocardiographic parameters in 64 patients. The left ventricular end-diastolic diameter (LVDd) and left ventricular end-systolic diameter (LVDs) were correlated with the Cir.T (r = 0.58, P < 0.0001, and r = 0.60, P < 0.0001 respectively). In addition, the left ventricular ejection fraction (LVEF) was negatively correlated with the Cir.T (r = 0.48, P < 0.0001). The average Cir. T was longer in patients with LVEF < 35% (8.0 s vs 6.7 s; P < 0.05) or LVDd > 55 mm (7.9 s vs 6.2 s; P < 0.05) than in the other patients. The quality of the MDCT images of the coronary veins obtained at different scan timings (coronary artery phase and 10 s or 15 s after the coronary artery phase) were graded and classified into four categories (0 = worst, 3 = best) in 25 patients with LVEF < 35%. The delays of 10 and 15 s after the coronary artery phase significantly improved the mean image quality (P < 0.05). The Cir.T was prolonged in patients with low LVEF and LV dilation. An appropriate delay improved the quality of the MDCT images of the coronary veins in patients with LV dysfunction.
引用
收藏
页码:434 / 439
页数:6
相关论文
共 22 条
[11]   Intrapericardial administration of basic fibroblast growth factor: Myocardial and tissue distribution and comparison with intracoronary and intravenous administration [J].
Laham, RJ ;
Rezaee, M ;
Post, M ;
Xu, XY ;
Sellke, FW .
CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2003, 58 (03) :375-381
[12]   Coronary venous imaging with electron beam computed tomographic angiography: Three-dimensional mapping and relationship with coronary arteries [J].
Mao, SS ;
Shinbane, JS ;
Girsky, MJ ;
Child, J ;
Carson, S ;
Oudiz, RJ ;
Budoff, MJ .
AMERICAN HEART JOURNAL, 2005, 150 (02) :315-322
[13]   Imaging of the cardiac venous system:: Comparison of MDCT and conventional angiography [J].
Mühlenbruch, G ;
Koos, R ;
Wildberger, JE ;
Günther, RW ;
Mahnken, AH .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2005, 185 (05) :1252-1257
[14]   Reliable noninvasive coronary angiography with fast submillimeter multislice spiral computed tomography [J].
Nieman, K ;
Cademartiri, F ;
Lemos, PA ;
Raaijmakers, R ;
Pattynama, PMT ;
de Feyter, PJ .
CIRCULATION, 2002, 106 (16) :2051-2054
[15]  
OZAWA M, 2007, CIRC J S, V71, P404
[16]   SIZE OF CORONARY SINUS AT NECROPSY IN SUBJECTS WITHOUT CARDIAC DISEASE AND IN PATIENTS WITH VARIOUS CARDIAC CONDITIONS [J].
POTKIN, BN ;
ROBERTS, WC .
AMERICAN JOURNAL OF CARDIOLOGY, 1987, 60 (16) :1418-1421
[17]   Initial evaluation of coronary images from 320-detector row computed tomography [J].
Rybicki, Frank J. ;
Otero, Hansel J. ;
Steigner, Michael L. ;
Vorobiof, Gabriel ;
Nallamshetty, Leelakrishna ;
Mitsouras, Dimitrios ;
Ersoy, Hale ;
Mather, Richard T. ;
Judy, Philip F. ;
Cai, Tianxi ;
Coyner, Karl ;
Schultz, Kurt ;
Whitmore, Amanda G. ;
Di Carli, Marcelo F. .
INTERNATIONAL JOURNAL OF CARDIOVASCULAR IMAGING, 2008, 24 (05) :535-546
[18]  
Schiller N B, 1989, J Am Soc Echocardiogr, V2, P358
[19]   The coronary venous anatomy - A segmental approach to aid cardiac resynchronization therapy [J].
Singh, JP ;
Houser, S ;
Heist, EK ;
Ruskin, JN .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2005, 46 (01) :68-74
[20]   Three-dimensional visualization of the coronary venous system using multidetector row computed tomography [J].
Tada, H ;
Kurosaki, K ;
Naito, S ;
Koyama, K ;
Itoi, K ;
Ito, S ;
Ueda, M ;
Shinbo, G ;
Hoshizaki, H ;
Nogami, A ;
Oshima, S ;
Taniguchi, K .
CIRCULATION JOURNAL, 2005, 69 (02) :165-170