Three dimensional echo-planar imaging at 7 Tesla

被引:209
作者
Poser, B. A. [1 ,2 ]
Koopmans, P. J. [2 ]
Witzel, T. [3 ,4 ]
Wald, L. L. [3 ,4 ]
Barth, M. [1 ,2 ]
机构
[1] Univ Duisburg Essen, Erwin L Hahn Inst Magnet Resonance Imaging, D-45141 Essen, Germany
[2] Radboud Univ Nijmegen, Donders Inst Brain Cognit & Behav, Ctr Cognit Neuroimaging, NL-6525 ED Nijmegen, Netherlands
[3] Massachusetts Gen Hosp, Dept Radiol, AA Martinos Ctr, Charlestown, MA USA
[4] MIT, Harvard Mit Div Hlth Sci & Technol, Cambridge, MA 02139 USA
基金
美国国家卫生研究院;
关键词
FUNCTIONAL MRI; PHYSIOLOGICAL NOISE; FMRI TECHNIQUE; PRESTO-SENSE; BRAIN; SENSITIVITY; EPI; OPTIMIZATION; ACQUISITION; PRINCIPLES;
D O I
10.1016/j.neuroimage.2010.01.108
中图分类号
Q189 [神经科学];
学科分类号
071006 ;
摘要
Functional MRI (fMRI) most commonly employs 2D echo-planar imaging (EPI). The advantages for fMRI brought about by the increasingly popular ultra-high field strengths are best exploited in high-resolution acquisitions, but here 2D EPI becomes unpractical for several reasons, including the very long volume acquisitions times. In this study at 7 T, a 3D EPI sequence with full parallel and partial Fourier imaging capability along both phase encoding axes was implemented and used to evaluate the sensitivity of 3D and corresponding 2D EPI acquisitions at four different spatial resolutions ranging from small to typical voxel sizes (1.5-3.0 mm isotropic). Whole-brain resting state measurements (N = 4) revealed a better, or at least comparable sensitivity of the 3D method for gray and white matter. The larger vulnerability of 3D to physiological effects was outweighed by the much shorter volume TR, which moreover allows whole-brain coverage at high resolution within fully acceptable limits for event-related fMRI: TR was only 3.07 s for 1.5 mm, 1.88 s for 2.0 mm, 1.38 s for 2.5 mm and 1.07 s for 3.0 mm isotropic resolution. In order to investigate the ability to detect and spatially resolve BOLD activation in the visual cortex, functional 3D EPI experiments (N = 8) were performed at 1 mm isotropic resolution with parallel imaging acceleration of 3 x 3, resulting in a TR of only 3.2 s for whole-brain coverage. From our results, and several other practical advantages of 3D over 2D EPI found in the present study, we conclude that 3D EPI provides a useful alternative for whole-brain fMRI at 7 T, not only when high-resolution data are required. (C) 2010 Elsevier Inc. All rights reserved.
引用
收藏
页码:261 / 266
页数:6
相关论文
共 30 条
[1]   Accelerated parallel imaging for functional imaging of the human brain [J].
de Zwart, JA ;
van Gelderen, P ;
Golay, X ;
Ikonomidou, VN ;
Duyn, JH .
NMR IN BIOMEDICINE, 2006, 19 (03) :342-351
[2]   Application of sensitivity-encoded echo-planar imaging for blood oxygen level-dependent functional brain imaging [J].
de Zwart, JA ;
van Gelderen, P ;
Kellman, P ;
Duyn, JH .
MAGNETIC RESONANCE IN MEDICINE, 2002, 48 (06) :1011-1020
[3]  
Golay X, 2000, MAGNET RESON MED, V43, P779, DOI 10.1002/1522-2594(200006)43:6<779::AID-MRM1>3.0.CO
[4]  
2-4
[5]   Generalized Autocalibrating Partially Parallel Acquisitions (GRAPPA) [J].
Griswold, MA ;
Jakob, PM ;
Heidemann, RM ;
Nittka, M ;
Jellus, V ;
Wang, JM ;
Kiefer, B ;
Haase, A .
MAGNETIC RESONANCE IN MEDICINE, 2002, 47 (06) :1202-1210
[6]  
HU Y, 2007, P 15 ANN M ISMRM BER, P1945
[7]   Three-dimensional spiral technique for high-resolution functional MRI [J].
Hu, Yanle ;
Glover, Gary H. .
MAGNETIC RESONANCE IN MEDICINE, 2007, 58 (05) :947-951
[8]   Improved optimization for the robust and accurate linear registration and motion correction of brain images [J].
Jenkinson, M ;
Bannister, P ;
Brady, M ;
Smith, S .
NEUROIMAGE, 2002, 17 (02) :825-841
[9]   Single-shot half k-space high-resolution gradient-recalled EPI for fMRI at 3 tesla [J].
Jesmanowicz, A ;
Bandettini, PA ;
Hyde, JS .
MAGNETIC RESONANCE IN MEDICINE, 1998, 40 (05) :754-762
[10]   A PRESTO-SENSE sequence with alternating partial-Fourier encoding for rapid susceptibility-weighted 3D MRI time series [J].
Klarhöfer, M ;
Dilharreguy, B ;
van Gelderen, P ;
Moonen, CTW .
MAGNETIC RESONANCE IN MEDICINE, 2003, 50 (04) :830-838