Total body and regional bone mineral densitometry (BMD) and soft tissue measurements: Correlations of BMD parameter to lumbar spine and hip

被引:41
作者
Franck, H [1 ]
Munz, M [1 ]
机构
[1] Ctr Rheumatol, D-82487 Oberammergau, Germany
关键词
bone densitometry; total body; subregional BMD;
D O I
10.1007/s00223001124
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Bone loss in men and women seems to differ according to the skeletal regions or particular areas being evaluated. Dual energy X-ray absorptiometry (DXA) is the method of choice for measuring total body and regional bone mineral area density (BMD). The aim of the study was to evaluate the importance of DXA measurements of total body in relation to lumbar spine and hip in different scan beam designs. In 300 patients, ages 43-80 years, lumbar spine, hip, total body and regional bone mineral area density, and soft tissue measurements were performed on all subjects in the supine position on a QDR 2000 using single beam (SB) and fan beam (FB). Short-term precision errors were 0.7% (SB) and 1.2% (FB) for BMD total of the total body and between 1.2% and 8.0% for soft tissue measurements. All mid-term precision errors of BMD total, right and left leg, and pelvis were below 2.0% with SE and FB, whereas precision errors of thoracic and lumbar spine varied depending on the scan mode being applied. In contrast, all mid-term precision errors of soft tissue measurements were greater (2.6-11.0%). All SE values of BMD and soft tissue measurement were significantly higher than FB values, except for BMD values of the head, thoracic spine, and pelvis. Furthermore, BMD total of the total body scan correlated significantly (P < 0.001) with all subregional parameters with best "r"-values (0.86-0.92) for the right and left leg in SE and FB design. In addition, there were excellent correlations (r > 0.94, P < 0.001) between the right and left legs (SB and FB) or arms (SB). There were also highly significant correlations between the lumbar spine (or hip) and total body, being best for the subregional thorax. Our data demonstrate short-and mid-term precision errors of BMD with reproducible results for most areas in SE and FB design, whereas soft tissue measurements vary depending on the area being measured. Furthermore, there is a close relationship between BMD values of total body total and subregional parameters and lumbar spine and hip scans, respectively.
引用
收藏
页码:111 / 115
页数:5
相关论文
共 14 条
[1]   DUAL X-RAY ABSORPTIOMETRY - A COMPARISON BETWEEN FAN-BEAM AND PENCIL BEAM SCANS [J].
BLAKE, GM ;
PARKER, JC ;
BUXTON, FMA ;
FOGELMAN, I .
BRITISH JOURNAL OF RADIOLOGY, 1993, 66 (790) :902-906
[2]  
DOOK JE, 1997, OFFICIAL AM COLL SPO, P291
[3]   A multi-center comparison of dual energy X-ray absorptiometers: In vivo and in vitro soft tissue measurement [J].
Economos, CD ;
Nelson, ME ;
Fiatarone, MA ;
Dallal, GE ;
Heymsfield, SB ;
Wang, J ;
Yasumara, S ;
Ma, R ;
Vaswani, AN ;
RussellAulet, M ;
Pierson, RN .
EUROPEAN JOURNAL OF CLINICAL NUTRITION, 1997, 51 (05) :312-317
[4]   BILATERAL COMPARISON OF FEMORAL BONE-DENSITY AND HIP AXIS LENGTH FROM SINGLE AND FAN-BEAM DXA SCANS [J].
FAULKNER, KG ;
GENANT, HK ;
MCCLUNG, M .
CALCIFIED TISSUE INTERNATIONAL, 1995, 56 (01) :26-31
[5]   Bone mineral density of opposing hips using dual energy X-ray absorptiometry in single-beam and fan-beam design [J].
Franck, H ;
Munz, M ;
Scherrer, M .
CALCIFIED TISSUE INTERNATIONAL, 1997, 61 (06) :445-447
[6]   EVALUATION OF DUAL-ENERGY X-RAY ABSORPTIOMETRY BONE-MINERAL MEASUREMENT COMPARISON OF A SINGLE-BEAM AND FAN-BEAM DESIGN - THE EFFECT OF OSTEOPHYTIC CALCIFICATION ON SPINE BONE-MINERAL DENSITY [J].
FRANCK, H ;
MUNZ, M ;
SCHERRER, M .
CALCIFIED TISSUE INTERNATIONAL, 1995, 56 (03) :192-195
[7]   ASSESSMENT OF THE COMPOSITION OF MAJOR BODY REGIONS BY DUAL-ENERGY X-RAY ABSORPTIOMETRY (DEXA), WITH SPECIAL REFERENCE TO LIMB MUSCLE MASS [J].
FULLER, NJ ;
LASKEY, MA ;
ELIA, M .
CLINICAL PHYSIOLOGY, 1992, 12 (03) :253-266
[8]   PRECISION AND STABILITY OF DUAL-ENERGY X-RAY ABSORPTIOMETRY MEASUREMENTS [J].
JOHNSON, J ;
DAWSONHUGHES, B .
CALCIFIED TISSUE INTERNATIONAL, 1991, 49 (03) :174-178
[9]   DUAL-ENERGY X-RAY ABSORPTIOMETRY FOR TOTAL-BODY AND REGIONAL BONE-MINERAL AND SOFT-TISSUE COMPOSITION [J].
MAZESS, RB ;
BARDEN, HS ;
BISEK, JP ;
HANSON, J .
AMERICAN JOURNAL OF CLINICAL NUTRITION, 1990, 51 (06) :1106-1112
[10]   In vivo validation of whole body composition estimates from dual-energy X-ray absorptiometry [J].
Prior, BM ;
Cureton, KJ ;
Modlesky, CM ;
Evans, EM ;
Sloniger, MA ;
Saunders, M ;
Lewis, RD .
JOURNAL OF APPLIED PHYSIOLOGY, 1997, 83 (02) :623-630