Fully automated method for the liquid chromatographic-tandem mass spectrometric determination of cyproterone acetate in human plasma using restricted access material for on-line sample clean-up

被引:30
作者
Christiaens, B
Fillet, A
Chiap, P
Rbeida, O
Ceccato, A
Streel, B
De Graeve, J
Crommen, J
Hubert, P
机构
[1] Univ Liege, CHU, Dept Pharmaceut Analyt Chem, Inst Pharm, B-4000 Liege, Belgium
[2] Galephar MF, B-6900 Marche En Famenne, Belgium
[3] Univ Liege, Dept Environm & Ind Toxicol, B-4000 Liege, Belgium
关键词
validation; mass spectrometry; cyproterone acetate;
D O I
10.1016/j.chroma.2004.06.107
中图分类号
Q5 [生物化学];
学科分类号
071010 [生物化学与分子生物学]; 081704 [应用化学];
摘要
A new automated method for the quantitative analysis of cyproterone acetate (CPA) in human plasma has been developed using on-line solid phase extraction (SPE) prior to the LC-MS/MS determination. The method was based on the use of a pre-column packed with internal-surface reversed-phase material (LiChrospher RP-4 ADS, 25 mm x 2 mm) for sample clean-up coupled to LC separation on an octadecyl silica stationary phase by means of a column switching system. A 30 mul plasma sample volume was injected directly onto the pre-column using a mixture of water, acetonitrile and formic acid (90:10:0.1 (v/v/v)) adjusted to pH 4.0 with diluted ammonia as washing liquid. The analyte was then eluted in the back-flush mode with the LC mobile phase consisting of water, methanol and formic acid (10:90:0.1 (v/v/v)). The dispensing flow rates of the washing liquid and the LC mobile phase were 300 mul min(-1). Medroxyprogesterone acetate (MPA) was used as internal standard. The MS ionization of the analytes was achieved using electrospray (ESI) in the positive ion mode. The pseudomolecular ionic species of CPA and MPA (417.4 and 387.5) were selected to generate daughter ions at 357.4 and 327.5, respectively. Finally, the developed method was validated according to a new approach using accuracy profiles as a decision tool. Very good results with respect to accuracy, detectability, repeatability, intermediate precision and selectivity were obtained. The LOQ of cyproterone acetate was 300 pg ml(-1). (C) 2004 Elsevier B.V. All rights reserved.
引用
收藏
页码:105 / 110
页数:6
相关论文
共 27 条
[1]
DEVELOPMENT AND PERFORMANCE OF AN AUTOMATED HPLC-ANALYZER FOR CATECHOLAMINES [J].
BOOS, KS ;
WILMERS, B ;
SAUERBREY, R ;
SCHLIMME, E .
CHROMATOGRAPHIA, 1987, 24 :363-370
[2]
High-performance liquid chromatography integrated solid-phase extraction in bioanalysis using restricted access precolumn packings [J].
Boos, KS ;
Grimm, CH .
TRAC-TRENDS IN ANALYTICAL CHEMISTRY, 1999, 18 (03) :175-180
[3]
ALKYL-DIOL SILICA (ADS) - RESTRICTED ACCESS PRECOLUMN PACKINGS FOR DIRECT-INJECTION AND COUPLED-COLUMN CHROMATOGRAPHY OF BIOFLUIDS [J].
BOOS, KS ;
RUDOLPHI, A ;
VIELHAUER, S ;
WALFORT, A ;
LUBDA, D ;
EISENBEISS, F .
FRESENIUS JOURNAL OF ANALYTICAL CHEMISTRY, 1995, 352 (7-8) :684-690
[4]
An analysis of the SFSTP guide on validation of chromatographic bioanalytical methods: progresses and limitations [J].
Boulanger, B ;
Chiap, P ;
Dewe, W ;
Crommen, J ;
Hubert, P .
JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 2003, 32 (4-5) :753-765
[5]
Ceccato A, 1999, ENANTIOMER, V4, P305
[6]
Automated determination of pirlindole enantiomers in plasma by on-line coupling of a pre-column packed with restricted access material to a chiral liquid chromatographic column [J].
Chiap, P ;
Ceccato, A ;
Gora, R ;
Hubert, P ;
Géczy, J ;
Crommen, J .
JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 2002, 27 (3-4) :447-455
[7]
Fully automated method for the liquid chromatographic determination of cyproterone acetate in plasma using restricted access material for sample pre-treatment [J].
Christiaens, B ;
Chiap, P ;
Rbeida, O ;
Cello, D ;
Crommen, J ;
Hubert, P .
JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2003, 795 (01) :73-82
[8]
Department of Health and Human Services Food and Drug Administration Center for Drug Evaluation and Research, 2001, GUID IND BIOAN METH
[9]
DETERMINATION OF MEGESTROL-ACETATE AND CYPROTERONE-ACETATE IN SERUM OF PATIENTS WITH ADVANCED BREAST-CANCER BY HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY [J].
DIKKESCHEI, LD ;
WOLTHERS, BG ;
DERUYTERBUITENHUIS, AW ;
NAGEL, GT ;
SLEIJFER, DT ;
WILLEMSE, PHB ;
VANDERSLIK, W .
JOURNAL OF CHROMATOGRAPHY-BIOMEDICAL APPLICATIONS, 1990, 529 (01) :145-154
[10]
15-HYDROXYCYPROTERONE ACETATE AND CYPROTERONE-ACETATE LEVELS IN PLASMA AND URINE [J].
FRITH, RG ;
PHILLIPOU, G .
JOURNAL OF CHROMATOGRAPHY, 1985, 338 (01) :179-186