EX-VIVO EVALUATION OF FERROMAGNETISM, HEATING, AND ARTIFACTS PRODUCED BY HEART-VALVE PROSTHESES EXPOSED TO A 1.5-T MR SYSTEM

被引:43
作者
SHELLOCK, FG
MORISOLI, SM
机构
[1] ST JOHNS HOSP, SANTA MONICA, CA USA
[2] UNIV CALIF LOS ANGELES, SCH MED, DEPT RADIOL SCI, LOS ANGELES, CA USA
关键词
HEART; MR; PROSTHESES; SAFETY;
D O I
10.1002/jmri.1880040521
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Ex vivo testing techniques were used to determine the ferromagnetic qualities of, presence of heating in, and artifacts produced by 13 different heart valve prostheses exposed to a 1.5-T (64-MHz) magnetic resonance (MR) system. None of the heart valve prostheses showed a measurable deflection in the 1.5-T static magnetic field. Only minimal artifacts were produced during MR imaging with a fast spoiled GRASS (gradient-recalled acquisition in the steady state) pulse sequence. The largest temperature changes measured during a ''worst case'' MR imaging sequence (estimated average specific absorption rate, 2.5 W/kg; estimated spatial peak specific absorption rate, 7.6 W/kg) were +0.2-degrees-C with the implant imaged ''in air'' and +0.3-degrees-C with the implant imaged in normal saline. Therefore, MR procedures performed with a 1.5-T (64-MHz) MR system may be performed safely in patients with any of the 13 different heart valve prostheses evaluated in this study.
引用
收藏
页码:756 / 758
页数:3
相关论文
共 21 条
  • [1] BELLON EM, 1984, MAGN RESON IMAGING, V2, P41
  • [2] POTENTIAL HAZARDS IN NMR IMAGING - HEATING EFFECTS OF CHANGING MAGNETIC-FIELDS AND RF FIELDS ON SMALL METALLIC IMPLANTS
    DAVIS, PL
    CROOKS, L
    ARAKAWA, M
    MCREE, R
    KAUFMAN, L
    MARGULIS, AR
    [J]. AMERICAN JOURNAL OF ROENTGENOLOGY, 1981, 137 (04) : 857 - 860
  • [3] Frank H, 1992, EUR J RADIOL, V2, P555, DOI DOI 10.1007/BF00187551
  • [4] HASSLER M, 1986, Journal de Radiologie (Paris), V67, P661
  • [5] IMPORTANT CONSIDERATIONS IN MEASUREMENT OF ATTRACTIVE FORCE ON METALLIC IMPLANTS IN MR IMAGERS
    KAGETSU, NJ
    LITT, AW
    [J]. RADIOLOGY, 1991, 179 (02) : 505 - 508
  • [6] SAFETY CONSIDERATIONS IN MR IMAGING
    KANAL, E
    SHELLOCK, FG
    TALAGALA, L
    [J]. RADIOLOGY, 1990, 176 (03) : 593 - 606
  • [7] Pusey E, 1986, Radiographics, V6, P891
  • [8] MAGNETIC-RESONANCE IMAGING OF PROSTHETIC CARDIAC VALVES INVITRO AND INVIVO
    RANDALL, PA
    KOHMAN, LJ
    SCALZETTI, EM
    SZEVERENYI, NM
    PANICEK, DM
    [J]. AMERICAN JOURNAL OF CARDIOLOGY, 1988, 62 (13) : 973 - 976
  • [9] MR IMAGING AND METALLIC IMPLANTS FOR ANTERIOR CRUCIATE LIGAMENT RECONSTRUCTION - ASSESSMENT OF FERROMAGNETISM AND ARTIFACT
    SHELLOCK, FG
    MINK, JH
    CURTIN, S
    FRIEDMAN, MJ
    [J]. JMRI-JOURNAL OF MAGNETIC RESONANCE IMAGING, 1992, 2 (02): : 225 - 228
  • [10] MR IMAGING AND BIOMEDICAL IMPLANTS, MATERIALS, AND DEVICES - AN UPDATED REVIEW
    SHELLOCK, FG
    CURTIS, JS
    [J]. RADIOLOGY, 1991, 180 (02) : 541 - 550