PET DETECTOR MODULES BASED ON NOVEL DETECTOR TECHNOLOGIES

被引:42
作者
MOSES, WW
DERENZO, SE
BUDINGER, TF
机构
[1] Lawrence Berkeley Laboratory, University of California, Berkeley
关键词
D O I
10.1016/0168-9002(94)91635-7
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
A successful PET detector module must identify 511 keV photons with: high efficiency (> 85%), high spatial resolution (<5 mm fwhm), low cost (<$600/in.(2)), low dead time (<4 mu s in.(2)), good timing resolution (<5 ns fwhm for conventional PET, < 200 ps fwhm for time of flight), and good energy resolution (< 100 keV fwhm), where these requirements are listed in order of decreasing importance. The ''high efficiency'' requirement also implies that the detector modules must pack together without inactive gaps. Several novel and emerging radiation detector technologies could improve the performance of PET detectors. Avalanche photodiodes, PIN photodiodes, metal channel dynode photomultiplier tubes, and new scintillators all have the potential to improve PET detectors significantly.
引用
收藏
页码:189 / 194
页数:6
相关论文
共 21 条
[1]   SILICON DRIFT PHOTODIODES [J].
AVSET, BS ;
ELLISON, J ;
EVENSEN, L ;
HALL, G ;
HANSEN, TE ;
ROE, S ;
WHEADON, R .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1990, 288 (01) :131-136
[2]   CD1-XZNXTE4 GAMMA-RAY DETECTORS [J].
BUTLER, JF ;
LINGREN, CL ;
DOTY, FP .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1992, 39 (04) :605-609
[3]   A NEW LIQUID XENON SCINTILLATION DETECTOR FOR POSITRON EMISSION TOMOGRAPHY [J].
CHEPEL, VY .
NUCLEAR TRACKS AND RADIATION MEASUREMENTS, 1993, 21 (01) :47-51
[4]   A PHOTOMULTIPLIER TUBE INCORPORATING AN AVALANCHE PHOTODIODE [J].
CUSHMAN, P ;
RUSACK, R .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1993, 333 (2-3) :381-390
[5]  
DERENZO SE, 1993, INT CONGR SER, V1030, P25
[6]   APPLICATION OF CADMIUM TELLURIDE DETECTORS TO HIGH-ENERGY COMPUTED-TOMOGRAPHY [J].
GLASSER, F ;
THOMAS, G ;
CUZIN, M ;
VERGER, L .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1992, 322 (03) :619-622
[7]   FAST, HIGH-DENSITY AVALANCHE PHOTODIODE-ARRAY [J].
GRAMSCH, E ;
SZAWLOWSKI, M ;
ZHANG, S ;
MADDEN, M .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1994, 41 (04) :762-766
[8]   20-ELEMENT HGI2 ENERGY DISPERSIVE-X-RAY ARRAY DETECTOR SYSTEM [J].
IWANCZYK, JS ;
DORRI, N ;
WANG, M ;
SZCZEBIOT, RW ;
DABROWSKI, AJ ;
HEDMAN, B ;
HODGSON, KO ;
PATT, BE .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1992, 39 (05) :1275-1280
[9]  
KWAIT PG, 1993, PHYS REV A, V48, pR867
[10]   PHOTOMULTIPLIER TUBE OF NEW DYNODE CONFIGURATION [J].
KYUSHIMA, H ;
HASEGAWA, Y ;
ATSUMI, A ;
NAGURA, K ;
YOKOTA, H ;
ITO, M ;
TAKEUCHI, J ;
OBA, K ;
MATSUURA, H ;
SUZUKI, S .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1994, 41 (04) :725-729