AN ANALYTIC CALCULATION OF THE ENERGY FLUENCE SPECTRUM OF A LINEAR-ACCELERATOR

被引:21
作者
DESOBRY, GE [1 ]
BOYER, AL [1 ]
机构
[1] UNIV TEXAS,MD ANDERSON CANC CTR,HOUSTON,TX 77030
关键词
BREMSSTRAHLUNG; SCHIFF SPECTRUM; ENERGY FLUENCE SPECTRUM;
D O I
10.1118/1.597179
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Calculation of photon dose by convolution methods requires a knowledge of the fluence spectrum of photons produced by the linear accelerator treatment head. But this spectrum is very difficult to measure accurately, and is often derived by Monte Carlo calculations modeling the elements of the treatment head. In this paper an analytic calculation technique and the corresponding computer tool for calculating the photon energy fluence spectral distribution at any point in a bremsstrahlung field are presented. Primary bremsstrahlung photon distributions are computed by modeling electron dispersion in layers in a thick target and using the thin target bremsstrahlung cross section formulas of Schiff. The first Compton scatter from all materials in a linear accelerator treatment head is computed analytically. Higher-order Compton scatter events and pair production annihilation photons are ignored, but the attenuation of both primary and first scattered photon fluence is computed. Predictions of the computer implementation of the model are compared to measurements of bremsstrahlung production in a thick target and to Monte Carlo calculations of the energy fluence emerging from a linear accelerator. Finally, the computer tool is used to investigate the source of collimator- dependent fluence fluctuations in air. In agreement with other measurements, the principal contribution to fluence outside the geometric field is found to be from scatter in the flattening filter.
引用
收藏
页码:1943 / 1952
页数:10
相关论文
共 15 条
[1]   DETERMINATION OF EFFECTIVE BREMSSTRAHLUNG SPECTRA AND ELECTRON CONTAMINATION FOR PHOTON DOSE CALCULATIONS [J].
AHNESJO, A ;
ANDREO, P .
PHYSICS IN MEDICINE AND BIOLOGY, 1989, 34 (10) :1451-1464
[2]  
BIELAJEW AF, 1990, MED PHYS, V17, P522
[3]   DETERMINATION OF THE PHOTON SPECTRUM OF A CLINICAL ACCELERATOR [J].
BROWNRIDGE, J ;
SAMNICK, S ;
STILES, P ;
TIPTON, P ;
VESELKA, J ;
YEH, N .
MEDICAL PHYSICS, 1984, 11 (06) :794-796
[4]   BREMSSTRAHLUNG REVIEW - AN ANALYSIS OF THE SCHIFF SPECTRUM [J].
DESOBRY, GE ;
BOYER, AL .
MEDICAL PHYSICS, 1991, 18 (03) :497-505
[5]   FORWARD-DIRECTED BREMSSTRAHLUNG OF 10-MEV TO 30-MEV ELECTRONS INCIDENT ON THICK TARGETS OF AL AND PB [J].
FADDEGON, BA ;
ROSS, CK ;
ROGERS, DWO .
MEDICAL PHYSICS, 1990, 17 (05) :773-785
[6]   ANGULAR-DISTRIBUTION OF BREMSSTRAHLUNG FROM 15-MEV ELECTRONS INCIDENT ON THICK TARGETS OF BE, AL, AND PB [J].
FADDEGON, BA ;
ROSS, CK ;
ROGERS, DWO .
MEDICAL PHYSICS, 1991, 18 (04) :727-739
[8]   HEAD SCATTER DATA FOR SEVERAL LINEAR ACCELERATORS (4-18MV) [J].
KASE, KR ;
SVENSSON, GK .
MEDICAL PHYSICS, 1986, 13 (04) :530-532
[9]   MEASUREMENT OF ACCELERATOR BREMSSTRAHLUNG SPECTRA WITH A HIGH-EFFICIENCY GE DETECTOR [J].
LANDRY, DJ ;
ANDERSON, DW .
MEDICAL PHYSICS, 1991, 18 (03) :527-532
[10]   ENERGY AND ANGULAR-DISTRIBUTIONS OF PHOTONS FROM MEDICAL LINEAR ACCELERATORS [J].
MOHAN, R ;
CHUI, C ;
LIDOFSKY, L .