X-RAY AND CHARGED-PARTICLE DETECTION WITH CSI(TL) LAYER COUPLED TO A-SI-H PHOTODIODE LAYERS

被引:47
作者
FUJIEDA, I [1 ]
CHO, G [1 ]
DREWERY, J [1 ]
GEE, T [1 ]
JING, T [1 ]
KAPLAN, SN [1 ]
PEREZMENDEZ, V [1 ]
WILDERMUTH, D [1 ]
STREET, RA [1 ]
机构
[1] UNIV CALIF BERKELEY LAWRENCE BERKELEY LAB,BERKELEY,CA 94720
关键词
D O I
10.1109/23.289306
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
A compact real-time X-ray and charged particle imager with digitized position output can be built either by coupling a fast scintillator to a photodiode array or by forming one on a photodiode array directly. CsI(Tl) layers 100-1000-mu-m thick were evaporated on glass substrates from a crystal CsI(Tl). When coupled to a crystalline Si or amorphous silicon (a-Si:H) photodiode and exposed to calibrated X-ray pulses, their light yields and speed were found to be comparable to those of a crystal CsI(Tl). Single beta particle detection was demonstrated with this combination. The light spread inside evaporated CsI(Tl) was suppressed by its columnar structure. Scintillation detection gives much larger signals than direct X-ray detection due to the increased energy deposition in the detector material. Fabrication of monolithic type X-ray sensors consisting of CsI + a-Si:H photodiodes is discussed.
引用
收藏
页码:255 / 262
页数:8
相关论文
共 22 条
[1]   DEVELOPMENT OF HYDROGENATED AMORPHOUS-SILICON SENSORS FOR HIGH-ENERGY PHOTON RADIOTHERAPY IMAGING [J].
ANTONUK, LE ;
YORKSTON, J ;
BOUDRY, J ;
LONGO, MJ ;
JIMENEZ, J ;
STREET, RA .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1990, 37 (02) :165-170
[2]  
BATES C, 1968, ADV ELECTRONICS ELEC, V28, P451
[3]   RESPONSE OF SILICON DETECTORS TO MONOENERGETIC ELECTRONS WITH ENERGIES BETWEEN 0.15 AND 5.0 MEV [J].
BERGER, MJ ;
SELTZER, SM ;
CHAPPELL, SE ;
HUMPHREYS, JC ;
MOTZ, JW .
NUCLEAR INSTRUMENTS & METHODS, 1969, 69 (02) :181-+
[4]   EFFECT OF PRIMARY IONIZATION IN AMORPHOUS-SILICON DETECTORS [J].
EQUER, B ;
KARAR, A .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1988, 271 (03) :574-584
[5]   APPLICATIONS OF A-SI-H RADIATION DETECTORS [J].
FUJIEDA, I ;
CHO, G ;
KAPLAN, SN ;
PEREZMENDEZ, V ;
QURESHI, S ;
STREET, RA .
JOURNAL OF NON-CRYSTALLINE SOLIDS, 1989, 115 (1-3) :174-176
[6]   TRACE-ELEMENT DETERMINATION WITH SEMICONDUCTOR DETECTOR X-RAY SPECTROMETERS [J].
GIAUQUE, RD ;
GOULDING, FS ;
JAKLEVIC, JM ;
PEHL, RH .
ANALYTICAL CHEMISTRY, 1973, 45 (04) :671-681
[7]   A MEASUREMENT OF THE LIGHT YIELD OF COMMON INORGANIC SCINTILLATORS [J].
HOLL, I ;
LORENZ, E ;
MAGERAS, G .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1988, 35 (01) :105-109
[8]  
ITO H, 1989, JPN J APPL PHYS, V28, pL1476
[9]   RADIATION-DAMAGE OF CSI(TL) CRYSTALS ABOVE 10(3) RAD [J].
KOBAYASHI, M ;
SAKURAGI, S .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1987, 254 (02) :275-280
[10]   A NEW SCINTILLATION MATERIAL - PURE CSI WITH 10 NS DECAY TIME [J].
KUBOTA, S ;
SAKURAGI, S ;
HASHIMOTO, S ;
RUAN, J .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1988, 268 (01) :275-277