INSTRUMENTATION AND STRATEGIES FOR HIGH-SPEED GAS-CHROMATOGRAPHY

被引:29
作者
PETERS, A [1 ]
KLEMP, M [1 ]
PUIG, L [1 ]
RANKIN, C [1 ]
SACKS, R [1 ]
机构
[1] UNIV MICHIGAN,DEPT CHEM,ANN ARBOR,MI 48109
关键词
HIGH-SPEED GAS CHROMATOGRAPHY; INSTRUMENTATION; VACUUM BACKFLUSH; CRYOFOCUSING; POROUS-LAYER OPEN-TUBULAR COLUMNS;
D O I
10.1039/an9911601313
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Short lengths of capillary column operated at unusually high carrier-gas velocities are used to separate relatively simple mixtures on a timescale of 20 s or less. Two inlet systems are described which can generate the narrow injection bands needed for these elution times. A cryofocusing inlet using a vacuum-pump-operated gas flow direction control is used for compounds generally less volatile than n-pentane. A low-friction gas valve inlet combined with porous-layer open-tubular columns is used for volatile compounds of low molecular mass. Strategies for high-speed separations including high-speed backflush, backflush with recycle, vacuum-outlet operation and selective detection are described. These techniques are used to increase the operational flexibility and the amount of information that can be obtained using high-speed gas chromatography.
引用
收藏
页码:1313 / 1320
页数:8
相关论文
共 34 条
[11]  
INGRAHAM D, 1981, HRC-J HIGH RES CHROM, V5, P227
[12]   IMPROVEMENT OF SPEED OF SEPARATION IN PACKED-COLUMN GAS-CHROMATOGRAPHY [J].
JONKER, RJ ;
POPPE, H ;
HUBER, JFK .
ANALYTICAL CHEMISTRY, 1982, 54 (14) :2447-2456
[13]   VACUUM BACKFLUSH RECYCLE SYSTEM FOR GC ANALYSIS OF HIGH-PURITY SOLVENTS [J].
KLEMP, M ;
SACKS, RD .
JOURNAL OF CHROMATOGRAPHIC SCIENCE, 1991, 29 (03) :114-121
[14]   STRATEGIES FOR RAPID GC IMPURITY ANALYSIS OF HIGH-PURITY SOLVENTS .1. INSTRUMENTATION AND PROCEDURES [J].
KLEMP, M ;
SACKS, R .
JOURNAL OF CHROMATOGRAPHIC SCIENCE, 1991, 29 (06) :243-247
[15]  
KLEMP M, 1991, HRC-J HIGH RES CHROM, V14, P235
[16]   STRATEGIES FOR RAPID GC IMPURITY ANALYSIS OF HIGH-PURITY SOLVENTS .2. COLUMN SELECTION AND SOLVENT EFFECTS [J].
KLEMP, M ;
SACKS, R .
JOURNAL OF CHROMATOGRAPHIC SCIENCE, 1991, 29 (06) :248-252
[17]   ELECTRICALLY HEATED COLD TRAP INLET SYSTEM FOR COMPUTER-CONTROLLED HIGH-SPEED GAS-CHROMATOGRAPHY [J].
LANNING, LA ;
SACKS, RD ;
MOURADIAN, RF ;
LEVINE, SP ;
FOULKE, JA .
ANALYTICAL CHEMISTRY, 1988, 60 (18) :1994-1996
[18]   MINIMUM ANALYSIS TIME IN CAPILLARY GAS-CHROMATOGRAPHY - VACUUM-OUTLET VERSUS ATMOSPHERIC-OUTLET COLUMN OPERATION [J].
LECLERCQ, PA ;
CRAMERS, CA .
JOURNAL OF HIGH RESOLUTION CHROMATOGRAPHY & CHROMATOGRAPHY COMMUNICATIONS, 1987, 10 (05) :269-272
[19]  
Liu Z, 1989, J MICROCOLUMN SEP, V1, P249
[20]   MEASUREMENT OF ORGANIC VAPORS AT SUB-TLV CONCENTRATIONS USING FAST GAS-CHROMATOGRAPHY [J].
MOURADIAN, RF ;
LEVINE, SP ;
SACKS, RD ;
SPENCE, MW .
AMERICAN INDUSTRIAL HYGIENE ASSOCIATION JOURNAL, 1990, 51 (02) :90-95