PHOTOELECTRON ESCAPE FROM A SILICON X-RAY-DETECTOR - MONTE-CARLO STUDY

被引:12
作者
AOKI, K
KOYAMA, M
机构
[1] Dept. of Radiol., Komazawa Junior Coll., Tokyo
关键词
D O I
10.1088/0031-9155/35/8/002
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Photoelectron escape fractions for small X-ray detectors used in the measurement of diagnostic X-ray spectra are calculated using a Monte Carlo method. The calculation is carried out for detectors 50 to 500 mu m in thickness and at several different diameters with incident photons in the energy range 10-300 keV. When a photoelectron escapes, the partial energy of an incident photon is absorbed in the detector; the spectra of the absorbed energy are shown. The calculated escape fractions are compared with experimental data, and a good agreement is obtained. The diagnostic X-ray spectra measured with a silicon detector 95 mu m in thickness is corrected for detector distortion using the calculated escape fraction; with this correction, a spectrum similar to that determined with a Ge detector is obtained. For an easy calculation of the escape fractions, an empirical equation is presented; this equation gives the escape fractions for any size of detector. The escape fraction depends strongly on detector size and on photon energy; the detector size suitable for measured X-ray energy is also discussed.
引用
收藏
页码:1051 / 1063
页数:13
相关论文
共 30 条
[1]   MEASUREMENT OF DIAGNOSTIC-X-RAY SPECTRA USING A SILICON PHOTODIODE [J].
AOKI, K ;
KOYAMA, M .
MEDICAL PHYSICS, 1989, 16 (04) :529-536
[2]   RESPONSE OF SILICON DETECTORS TO MONOENERGETIC ELECTRONS WITH ENERGIES BETWEEN 0.15 AND 5.0 MEV [J].
BERGER, MJ ;
SELTZER, SM ;
CHAPPELL, SE ;
HUMPHREYS, JC ;
MOTZ, JW .
NUCLEAR INSTRUMENTS & METHODS, 1969, 69 (02) :181-+
[3]  
Berger MJ., 1963, METHODS COMPUT PHYS, P135
[4]   MOLIERE THEORY OF MULTIPLE SCATTERING [J].
BETHE, HA .
PHYSICAL REVIEW, 1953, 89 (06) :1256-1266
[5]   PHOTOELECTRIC CROSS SECTIONS IN KEV RANGE [J].
BRYSK, H ;
ZERBY, CD .
PHYSICAL REVIEW, 1968, 171 (02) :292-&
[6]   MONTE-CARLO SIMULATION STUDIES OF DETECTORS USED IN THE MEASUREMENT OF DIAGNOSTIC-X-RAY SPECTRA [J].
CHEN, CS ;
DOI, K ;
VYBORNY, C ;
CHAN, HP ;
HOLJE, G .
MEDICAL PHYSICS, 1980, 7 (06) :627-635
[7]   THE X-RAY-CALIBRATION OF SILICON P-I-N-DIODES BETWEEN 1.5 AND 17.4 KEV [J].
CORALLO, DM ;
CREEK, DM ;
MURRAY, GM .
JOURNAL OF PHYSICS E-SCIENTIFIC INSTRUMENTS, 1980, 13 (06) :623-626
[8]  
DAVISSON CM, 1968, ALPHA BETA GAMMA RAY, V1
[9]   SPECTROMETRY OF LOW-ENERGY GAMMA- AND X-RAYS WITH GE(LI) DETECTORS [J].
DREXLER, G ;
PERZL, F .
NUCLEAR INSTRUMENTS & METHODS, 1967, 48 (02) :332-+
[10]   EXPERIMENTAL STUDY OF PHOTON ENERGY SPECTRUM OF PRIMARY DIAGNOSTIC X-RAYS [J].
EPP, ER ;
WEISS, H .
PHYSICS IN MEDICINE AND BIOLOGY, 1966, 11 (02) :225-&