ION-SELECTIVE MICROCHEMICAL SENSORS WITH REDUCED PRECONDITIONING TIME - MEMBRANE BIOSTABILITY STUDIES AND APPLICATIONS IN BLOOD ANALYSIS

被引:38
作者
COSOFRET, VV
ERDOSY, M
LINDNER, E
JOHNSON, TA
BUCK, RP
KAO, WJ
NEUMAN, MR
ANDERSON, JM
机构
[1] UNIV N CAROLINA,DEPT MED,CHAPEL HILL,NC 27599
[2] CASE WESTERN RESERVE UNIV,DEPT BIOMED ENGN,CLEVELAND,OH 44106
关键词
PLANAR MICROCHEMICAL SENSORS; POLYMERIC MEMBRANES; BLOOD ANALYSIS; MEMBRANE BIOSTABILITY;
D O I
10.1080/00032719408000310
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Progress on solution of two general problems regarding the use of in vivo planar microchemical sensors is reported. These are issues of short term and long term response stability. Reduction of preconditioning time (hydration period), i.e., the time needed by the planar microchemical sensors based on Kapton(R) substrate to achieve the optimal analytical performances, has been achieved. By storing the electrodes in containers with humid atmospheres (100% humidity) their short time responses, e.g. measured potential, when placed in samples to be analyzed, are practically constant after one minute of immersion. The electrode sensitivity, potential reproducibility and membrane resistance of both pH and K+ sensors were evaluated and compared before and after placing them in whole blood samples for specified periods of time. Blood serum samples were successfully assayed and the results compared with those obtained with a pH glass electrode and a blood gas analyzer, respectively. The long term stability of the membranes for in vivo use was investigated by determination of cell adhesion and membrane biostability (at 14 days of subcutaneous implantation in rats) using scanning electron microscopy.
引用
收藏
页码:3039 / 3063
页数:25
相关论文
共 32 条
  • [11] CUTLER SG, 1978, POLYM PREPRS AM CHEM, V19, P330
  • [12] LIFETIME OF NEUTRAL-CARRIER-BASED LIQUID MEMBRANES IN AQUEOUS SAMPLES AND BLOOD AND THE LIPOPHILICITY OF MEMBRANE-COMPONENTS
    DINTEN, O
    SPICHIGER, UE
    CHANIOTAKIS, N
    GEHRIG, P
    RUSTERHOLZ, B
    MORF, WE
    SIMON, W
    [J]. ANALYTICAL CHEMISTRY, 1991, 63 (06) : 596 - 603
  • [13] GUNARATNA PC, 1987, METHODOLOGY CLIN APP, V9, P52
  • [14] ENHANCED LIFETIME AND ADHESION OF K+-SENSITIVE, NH4+-SENSITIVE, AND CA-2+-SENSITIVE MEMBRANES ON SOLID-SURFACES USING HYDROXYL-MODIFIED POLYVINYLCHLORIDE MATRICES
    HARRISON, DJ
    CUNNINGHAM, LL
    LI, XZ
    TECLEMARIAM, A
    PERMANN, D
    [J]. JOURNAL OF THE ELECTROCHEMICAL SOCIETY, 1988, 135 (10) : 2473 - 2478
  • [15] PLASTICIZED POLYVINYL-CHLORIDE) PROPERTIES AND CHARACTERISTICS OF VALINOMYCIN ELECTRODES .1. HIGH-FREQUENCY RESISTANCES AND DIELECTRIC-PROPERTIES
    HORVAI, G
    GRAF, E
    TOTH, K
    PUNGOR, E
    BUCK, RP
    [J]. ANALYTICAL CHEMISTRY, 1986, 58 (13) : 2735 - 2740
  • [16] HORVAI G, 1989, ION SELECTIVE ELECTR, V5, P397
  • [17] A NOVEL SPECTROSCOPIC METHOD TO IMAGE H2O DISTRIBUTION IN ION-SELECTIVE MEMBRANES
    LI, XZ
    PETROVIC, S
    HARRISON, DJ
    [J]. SENSORS AND ACTUATORS B-CHEMICAL, 1990, 1 (1-6) : 275 - 280
  • [18] LI Z, 1993, ANAL METHOD INSTRUM, V1, P30
  • [19] FLEXIBLE (KAPTON-BASED) MICROSENSOR ARRAYS OF HIGH-STABILITY FOR CARDIOVASCULAR APPLICATIONS
    LINDNER, E
    COSOFRET, VV
    UFER, S
    BUCK, RP
    KUSY, RP
    ASH, RB
    NAGLE, HT
    [J]. JOURNAL OF THE CHEMICAL SOCIETY-FARADAY TRANSACTIONS, 1993, 89 (02): : 361 - 367
  • [20] ION-SELECTIVE MEMBRANES WITH LOW PLASTICIZER CONTENT - ELECTROANALYTICAL CHARACTERIZATION AND BIOCOMPATIBILITY STUDIES
    LINDNER, E
    COSOFRET, VV
    UFER, S
    BUCK, RP
    KAO, WJ
    NEUMAN, MR
    ANDERSON, JM
    [J]. JOURNAL OF BIOMEDICAL MATERIALS RESEARCH, 1994, 28 (05): : 591 - 601