Chemiluminometric flow-injection method for determination of free L-malate in wine with co-immobilized malate dehydrogenase NADH oxidase

被引:13
作者
Kiba, N
Inagaki, J
Furusawa, M
机构
[1] Department of Applied Chemistry and Biotechnology, Faculty of Engineering, Yamanashi University, Kofu
关键词
chemiluminescence; flow injection; co-immobilized enzyme reactor; malic acid wine;
D O I
10.1016/0039-9140(95)01660-0
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A flow-injection system with a co-immobilized malate dehydrogenase/reduced nicotineamide adenine dinucleotide (NADH) oxidase reactor and a chemiluminometer is described for the determination of free L-malate in wine. Malate dehydrogenase and NADH oxidase were co-immobilized on poly(vinyl alcohol) beads and packed into a stainless-steel column (5 cm x 4 mm i.d.). The hydrogen peroxide produced was detected chemiluminometrically via a luminol-hexacyanoferrate(III) reaction. The calibration graph was linear from 3 x 10(-7) M to 2.5 x 10(-4) M (the linear correlation coefficient was 0.9998); the detection limit (signal-to-noise ratio, 3) was 8 x 10(-8) M. The sample throughput was 30 h(-1) without carryover. The reactor was renewed every 2 weeks.
引用
收藏
页码:1751 / 1755
页数:5
相关论文
共 18 条
[1]   FLOW-INJECTION DETERMINATION OF MALATE WITH IMMOBILIZED MALATE-DEHYDROGENASE [J].
ALMUIABED, AM ;
TOWNSHEND, A .
ANALYTICA CHIMICA ACTA, 1989, 221 (02) :337-340
[2]  
BEUTLER HO, 1990, DEUT LEBENSM-RUNDSCH, V86, P341
[3]  
CHEMNITIUS GC, 1989, ANAL LETT, V22, P2897
[4]   ENZYMATIC DETERMINATION OF FREE L-(-)-MALIC ACID IN MUST AND WINE BY STOPPED-FLOW FLOW-INJECTION ANALYSIS [J].
DEMARIA, CG ;
MUNOZ, TM ;
MATEOS, AA .
ANALYTICA CHIMICA ACTA, 1991, 247 (01) :61-65
[5]   OPTIMIZATION OF FLOW-INJECTION SYSTEMS FOR DETERMINATION OF SUBSTRATES BY MEANS OF ENZYME AMPLIFICATION REACTIONS AND CHEMILUMINESCENCE DETECTION [J].
HANSEN, EH ;
NORGAARD, L ;
PEDERSEN, M .
TALANTA, 1991, 38 (03) :275-282
[6]   DEVELOPMENT OF A HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHIC SYSTEM WITH ENZYME REACTORS FOR THE DETERMINATION OF N-ACETYL BRANCHED-CHAIN AMINO-ACIDS [J].
KIBA, N ;
OYAMA, Y ;
FURUSAWA, M .
TALANTA, 1995, 42 (03) :449-453
[7]   DETERMINATION OF L-MALATE IN WINE BY FLOW-INJECTION WITH CO-IMMOBILIZED MALATE-DEHYDROGENASE OXALOACETATE DECARBOXYLASE [J].
KIBA, N ;
OGUCHI, M ;
FURUSAWA, M .
TALANTA, 1993, 40 (08) :1163-1165
[8]   FLOW-INJECTION DETERMINATION OF GLUCOSE IN SERUM WITH AN IMMOBILIZED PYRANOSE OXIDASE REACTOR [J].
KIBA, N ;
UEDA, F ;
FURUSAWA, M ;
YAMANE, T .
ANALYTICA CHIMICA ACTA, 1992, 269 (02) :187-191
[9]   CHEMILUMINESCENCE FLOW-INJECTION ANALYSIS DETERMINATION OF SUCROSE USING ENZYMATIC CONVERSION AND A MICROPOROUS MEMBRANE FLOW CELL [J].
KOERNER, CA ;
NIEMAN, TA .
ANALYTICAL CHEMISTRY, 1986, 58 (01) :116-119
[10]  
LIMA JLFC, 1992, AM J ENOL VITICULT, V43, P58