Core-based portal image registration for automatic radiotherapy treatment verification

被引:37
作者
Fritsch, DS
Chaney, EL
Boxwala, A
McAuliffe, MJ
Raghavan, S
Thall, A
Earnhart, JRD
机构
[1] UNIV N CAROLINA, DEPT BIOMED ENGN, CHAPEL HILL, NC 27599 USA
[2] UNIV N CAROLINA, DEPT COMP SCI, CHAPEL HILL, NC 27599 USA
来源
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS | 1995年 / 33卷 / 05期
关键词
radiotherapy; portal images; patient setup analysis; treatment verification; image registration;
D O I
10.1016/0360-3016(95)02092-6
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Purpose: Portal imaging is the most important quality assurance procedure for monitoring the reproducibility of setup geometry in radiation therapy. The role of portal imaging has become even more critical in recent years due to the migration of three-dimensional (3D) treatment planning technology, including high precision conformal therapy, from the research setting to routine clinical practice. Unfortunately, traditional methods for acquiring and interpreting portal images suffer from a number of deficiencies that contribute to the well-documented observation that many setup errors go undetected, and some persist for a clinically significant portion of the prescribed dose. Significant improvements in both accuracy and efficiency of detecting setup errors can, in principle, be achieved by using automatic image registration for on-line screening of images obtained from electronic portal imaging devices (EPIDs). Methods and Materials: This article presents recent developments in a method called core-based image analysis that shows great promise for achieving the desired improvements in error detection. Core-based image analysis is a fundamental computer vision method that is capable of exploiting the full power of EPIDs by providing for on-line detection of setup errors via automatic registration of user-selected anatomical structures. We describe a robust method for automatic portal image registration based on core analysis and demonstrate an approach for assessing both accuracy and precision of registration methods using realistic, digitally reconstructed portal radiographs (DRPRs) where truth is known. Results: Automatic core-based analysis of a set of 20 DRPRs containing known, random field positioning errors was performed for a patient undergoing treatment for prostate cancer. ln all cases,the translation was within 1 mm of the actual translation with mean absolute errors of 0.3 mm and standard deviations of 0.3 mm. In all cases, the reported rotation was within 0.6 degrees of the actual rotation with a mean absolute error of 0.18 degrees and a standard deviation of 0.23 degrees. Conclusion: Our results, using digitally reconstructed portal radiographs that closely resemble clinical portal images, suggest that automatic core-based registration is suitable as an on-line screening tool for detecting and quantifying patient setup errors.
引用
收藏
页码:1287 / 1300
页数:14
相关论文
共 56 条
[1]  
BAILY NA, 1980, INT J RADIAT ONCOL, V6, P935, DOI 10.1016/0360-3016(80)90341-7
[2]   CORRELATION OF PROJECTION RADIOGRAPHS IN RADIATION-THERAPY USING OPEN CURVE SEGMENTS AND POINTS [J].
BALTER, JM ;
PELIZZARI, CA ;
CHEN, GTY .
MEDICAL PHYSICS, 1992, 19 (02) :329-334
[3]  
Barrow H. G., 1977, TECHNICAL REPORT, P659
[4]   A VERIFICATION PROCEDURE TO IMPROVE PATIENT SET-UP ACCURACY USING PORTAL IMAGES [J].
BEL, A ;
VANHERK, M ;
BARTELINK, H ;
LEBESQUE, JV .
RADIOTHERAPY AND ONCOLOGY, 1993, 29 (02) :253-260
[5]   RADIATION-FIELD EDGE-DETECTION IN PORTAL IMAGES [J].
BIJHOLD, J ;
GILHUIJS, KGA ;
VANHERK, M ;
MEERTENS, H .
PHYSICS IN MEDICINE AND BIOLOGY, 1991, 36 (12) :1705-1710
[6]   FAST EVALUATION OF PATIENT SET-UP DURING RADIOTHERAPY BY ALIGNING FEATURES IN PORTAL AND SIMULATOR IMAGES [J].
BIJHOLD, J ;
VANHERK, M ;
VIJLBRIEF, R ;
LEBESQUE, JV .
PHYSICS IN MEDICINE AND BIOLOGY, 1991, 36 (12) :1665-1679
[7]  
BIJHOLD J, 1992, MED PHYS, V19, P408
[8]  
BOYER AL, 1992, MED PHYS, V19, P1455
[9]   OBJECT REPRESENTATION BY CORES - IDENTIFYING AND REPRESENTING PRIMITIVE SPATIAL REGIONS [J].
BURBECK, CA ;
PIZER, SM .
VISION RESEARCH, 1995, 35 (13) :1917-1930
[10]   A P-THETA TECHNIQUE FOR TREATMENT VERIFICATION IN RADIOTHERAPY AND ITS CLINICAL-APPLICATIONS [J].
DING, GX ;
SHALEV, S ;
GLUCHEV, G .
MEDICAL PHYSICS, 1993, 20 (04) :1135-1143