Effect of dose-reduced scan protocols on cardiac coronary image quality with 64-row MDCT: A cardiac phantom study

被引:23
作者
Fei, Xiaolu [1 ]
Du, Xiangying [1 ]
Li, Pengyu [1 ]
Liao, Jingmin [2 ]
Shen, Yun [2 ]
Li, Kuncheng [1 ]
机构
[1] Capital Med Univ, Xuanwu Hosp, Dept Radiol, Beijing 100053, Peoples R China
[2] GE China, CT Imaging Res Ctr, Beijing Econ & Technol Dev Area, Beijing 100176, Peoples R China
关键词
cardiac CT imaging; coronary artery stenosis; stenosis rate; 64-row MDCT; radiation dose;
D O I
10.1016/j.ejrad.2007.07.008
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Objective: To evaluate the feasibility of using relative low-dose scan protocols in coronary imaging with 64-row MDCT. Materials and methods: A pulsating cardiac phantom was used to simulate coronary arteries of two sizes Q and 5 mm in diameter) with three stenosis degrees (25, 50 and 75%) at 55 bpm heart rate. Cardiac scans were performed on a 64-row MDCT scanner (GE LightSpeed VCT) with rotation time of 350 ms and pitch of 0.2 under six different scan protocols: 120 kV/650 mA, 1137.5 mAs (effective) (CTDIvol 121.69 mGy), 120 kV/550 mA, 962.5 mAs (CTDIvol 102.96 mGy), 120 kV/450 mA, 787.5 mAs (CTDIvol 84.24 mGy), 120 kV/350 mA, 612.5 mAs (CTDIvol 65.52 mGy), 100 kV/590 mA, 1032.5 mAs (CTDIvol 65.17 mGy) and 140 kV/390 mA, 682.5 mAs (CTDIvol 102.22 mGy). The simulative coronary arteries were filled with contrast media to reach 300 HU in the lumen. Background noise was measured to describe the basic image quality accordingly. CNR, SNR and contour sharpness represented in slope of CT density curve was calculated as well. Measured stenosis area and rates, described by the percentage area of stenosis on the cross-section images were also calculated. Results: The corresponding image noise levels described in standard deviation of background signals varied with radiation dose, CNR and SNR mainly varied with tube current. The contour sharpness, which can reflect actual spatial resolution, is affected mainly by tube voltage. The first five protocols depicted obviously steeper curves than the sixth one (P < 0.05). As for 25% stenosis, there was no significant difference among the stenosis rates of the six protocols (P > 0.05). As for evaluation on 50 and 75% stenosis, there was no significant difference between the first two protocols, and between the second two protocols as well. However, significant difference presented between these two groups (P > 0.05). When comparing the groups with similar radiation dose, protocols with lower tube voltage gain more accuracy in representing stenosis area and rate. Conclusion: Dose level and corresponding image quality is relevant to the accuracy of stenosis evaluation on simulated coronary arteries with 64-row MDCT. In this study, we find relative low-dose protocols with acceptable image quality showed a tendency of overestimating stenosis. Furthermore, using a lower tube voltage and higher tube current to gain accurate imaging result is more applicable than other protocols with the same radiation dose level. (c) 2007 Elsevier Ireland Ltd. All rights reserved.
引用
收藏
页码:85 / 91
页数:7
相关论文
共 19 条
[1]  
Achenbach S, 2000, CIRCULATION, V102, P2823
[2]   Comparison of image quality in contrast-enhanced coronary artery visualization by electron beam tomography and retrospectively electrocardiogram-gated multislice spiral computed tomography [J].
Achenbach, S ;
Giesler, T ;
Ropers, D ;
Ulzheimer, S ;
Anders, K ;
Wenkel, E ;
Pohle, K ;
Kachelriess, M ;
Derlien, H ;
Kalender, WA ;
Daniel, WG ;
Bautz, W ;
Baum, U .
INVESTIGATIVE RADIOLOGY, 2003, 38 (02) :119-128
[3]  
Bash S, 2005, AM J NEURORADIOL, V26, P1012
[4]   Cardiac applications of multislice computed tomography [J].
de Roos, A ;
Kroft, LJM ;
Bax, JJ ;
Lamb, HJ ;
Geleijns, J .
BRITISH JOURNAL OF RADIOLOGY, 2006, 79 (937) :9-16
[5]  
FEI X, 2007, INF MED EQUIP, V22, P3
[6]   Estimation of radiation exposure in low-dose multislice computed tomography of the heart and comparison with a calculation program [J].
Hohl, C. ;
Muehlenbruch, G. ;
Wildberger, J. E. ;
Leidecker, C. ;
Suss, C. ;
Schmidt, T. ;
Guenther, R. W. ;
Mahnken, A. H. .
EUROPEAN RADIOLOGY, 2006, 16 (08) :1841-1846
[7]   Electron beam CT versus 16-MDCT on the variability of repeated coronary artery calcium measurements in a variable heart rate phantom [J].
Horiguchi, J ;
Shen, Y ;
Akiyama, Y ;
Hirai, N ;
Sasaki, K ;
Ishifuro, M ;
Nakanishi, T ;
Ito, K .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2005, 185 (04) :995-1000
[8]   Technical innovation of cardiac multirow detector CT using multisector reconstruction [J].
Horiguchi, J ;
Nakanishi, T ;
Tamura, A ;
Ito, K ;
Sasaki, K ;
Shen, Y .
COMPUTERIZED MEDICAL IMAGING AND GRAPHICS, 2002, 26 (04) :217-226
[9]   Ultra-low-dose coronary artery calcium screening using multislice CT with retrospective ECG gating [J].
Jakobs, TF ;
Wintersperger, BJ ;
Herzog, P ;
Flohr, T ;
Suess, C ;
Knez, A ;
Reiser, MF ;
Becker, CR .
EUROPEAN RADIOLOGY, 2003, 13 (08) :1923-1930
[10]   Quantification of obstructive and nonobstructive coronary lesions by 64-slice computed tomography - A comparative study with quantitative coronary angiography and intravascular ultrasound [J].
Leber, AW ;
Knez, A ;
von Ziegler, F ;
Becker, A ;
Nikolaou, K ;
Paul, S ;
Wintersperger, B ;
Reiser, M ;
Becker, CR ;
Steinbeck, G ;
Boekstegers, P .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2005, 46 (01) :147-154