Rapid, accurate and precise quantitative drug analysis:: comparing liquid chromatography tandem mass spectrometry and chip-based nanoelectrospray ionisation mass spectrometry

被引:8
作者
Alfazema, LN
Richards, DS
Gélébart, S
Mitchell, JC
Snowden, MJ
机构
[1] Univ Greenwich, Chatham ME4 4TB, Kent, England
[2] Pfizer Global Res & Dev, Analyt R&D, Sandwich CT13 9NJ, Kent, England
关键词
accuracy precision; 5-methoxytryptamine; amitriptyline; quantitative MS/MS; chip-based nanoelectrospray;
D O I
10.1255/ejms.772
中图分类号
O64 [物理化学(理论化学)、化学物理学]; O56 [分子物理学、原子物理学];
学科分类号
070203 ; 070304 ; 081704 ; 1406 ;
摘要
We have developed a liquid chromatography tandem mass spectrometry (LC-MS/MS) system capable of achieving better than 2% accuracy, routinely over a wide concentration range of 1-800 ngmL(-1). We demonstrate that the necessary high precision, high accuracy and rapid analysis can be achieved using LC-MS/MS technology. Automated nanoelectrospray ionisation tandem mass spectrometry (nanoESI-MS/MS) technology can be employed to eliminate the chromatographic step completely. In this paper, nanoESI-MS/MS is evaluated and compared directly with LC-MS/MS for the quantitative analysis of two-test analytes, amitriptyline (ATT) and 5-methoxytryptamine (5-MTT), in aqueous/organic mixture. Calibration curves were found to be linear over a wide concentration range of 1-800 ngmL(-1) for both analytes using LC-MS/MS. Using nanoESI-MS/MS ATT gave a linear response while 5-MTT gave a non-linear response using nanoESI-MS/MS over the same concentration range as in LC-MS/MS. Accuracy and precision values of quality control samples (QCs) at four concentration levels were analysed in replicates of six at each level using 5-MTT and ATT as test analytes for both techniques. The LC-MS/MS system was capable of achieving accuracy levels of 99.50-101.96% for ATT and 100.17-100.40% for 5-MTT. Accuracy levels using nanoESI-MS/MS were not comparable to LC-MS/MS, they ranged from 90.09-100.18% for ATT and 95.95-113.55% for 5-MTT. The precision values obtained for nanoESI-MS/MS were in good agreement with those obtained by LC-MS/MS.
引用
收藏
页码:393 / 402
页数:10
相关论文
共 17 条
[1]   Efficacy of amitriptyline for relief of pain in spinal cord injury: results of a randomized controlled trial [J].
Cardenas, DD ;
Warms, CA ;
Turner, JA ;
Marshall, H ;
Brooke, MM ;
Loeser, JD .
PAIN, 2002, 96 (03) :365-373
[2]   Determination of SCH 211803 by nanoelectrospray infusion mass spectrometry: evaluation of matrix effect and comparison with liquid chromatography-tandem mass spectrometry [J].
Chen, JW ;
Yang, LY ;
Kapron, JT ;
Ma, L ;
Pace, E ;
Van Pelt, CK ;
Rudewicz, PJ .
JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2004, 809 (02) :205-210
[3]   Demonstration of direct bioanalysis of drugs in plasma using nanoelectrospray infusion from a silicon chip coupled with tandem mass spectrometry [J].
Dethy, JM ;
Ackermann, BL ;
Delatour, C ;
Henion, JD ;
Schultz, GA .
ANALYTICAL CHEMISTRY, 2003, 75 (04) :805-811
[4]  
*FDA, 2001, FDA GUID IND
[5]  
*FDA, 2000, FDA GUID IND
[6]   Quantitation of midazolam in human plasma by automated chip-based infusion nanoelectrospray tandem mass spectrometry [J].
Kapron, JT ;
Pace, E ;
Van Pelt, CK ;
Henion, J .
RAPID COMMUNICATIONS IN MASS SPECTROMETRY, 2003, 17 (18) :2019-2026
[7]   Simultaneous selected reaction monitoring, MS/MS and MS3 quantitation for the analysis of pharmaceutical compounds in human plasma using chip-based infusion [J].
Leuthold, LA ;
Grivet, C ;
Allen, M ;
Baumert, M ;
Hopfgartner, G .
RAPID COMMUNICATIONS IN MASS SPECTROMETRY, 2004, 18 (17) :1995-2000
[8]   Improving quantitative measurements for the evaporative light scattering detector [J].
Mathews, BT ;
Higginson, PD ;
Lyons, R ;
Mitchell, JC ;
Sach, NW ;
Snowden, MJ ;
Taylor, MR ;
Wright, AG .
CHROMATOGRAPHIA, 2004, 60 (11-12) :625-633
[9]   Wide concentration range investigation of recovery, precision and error structure in liquid chromatography [J].
Schepers, U ;
Ermer, J ;
Preu, L ;
Wätzig, H .
JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2004, 810 (01) :111-118
[10]   Application of co-eluting structural analog internal standards for expanded linear dynamic range in liquid chromatography/electrospray mass spectrometry [J].
Shi, GE .
RAPID COMMUNICATIONS IN MASS SPECTROMETRY, 2003, 17 (03) :202-206