Storage phosphor and film-screen mammography:: performance with different mammographic techniques

被引:14
作者
Kheddache, S [1 ]
Thilander-Klang, A
Lanhede, B
Månsson, LG
Bjurstam, N
Ackerholm, P
Björneld, L
机构
[1] Univ Gothenburg, Sahlgrens Univ Hosp, Dept Diagnost Radiol, S-41345 Gothenburg, Sweden
[2] Univ Gothenburg, Sahlgrens Univ Hosp, Dept Radiat Phys, S-41345 Gothenburg, Sweden
关键词
digital radiography; storage phosphors; mammography; anthropomorphic phantom;
D O I
10.1007/s003300050716
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The aim of this study was to compare the image quality of storage phosphor plates with that in screen-film radiograms in mammography. Two anode/filter combinations were also compared - Mo/Mo and W/Rh. S Storage phosphor plates, generation IIIN (Fuji, Tokyo, Japan) and a conventional screen-film system (Kodak, Rochester, N. Y.) were evaluated using two mammographic units. One unit had a 0.6-mm focal spot, an anode/filter combination of Mo/Mo: and no grid (A(Mo)); the other had a 0.3-mm focal spot, a grid, and two possible combinations of anode/filter Mo/Mo (B-Mo) and W/Rh (B-W). Simulated tumours and microcalcifications were randomly positioned;in an anthropomorphic breast phantom (RMI model 165, no. 210-009, Radiation Measurements Inc., Middleton, Wisconsin). The image quality was evaluated using a modified version of receiver operating characteristics analysis. Five observers evaluated 300 films and 300 hard copy images each. Radiation doses were also determined. The image quality of the conventional screen-film images was significantly better than that for the storage phosphor plate mammograms. The B-Mo system rated best, for the detection of both tumours and microcalcifications,: all;:hough it was not significantly different from the Bw system. Systems B-Mo and B-W rated significantly better than the A(Mo) system for both image receptors studied. The mean absorbed dose was twice as high for the B-Mo system as for the A(Mo) and B-W systems for both conventional and digital technique. The mammograms produced with the screen-film combination gave a significantly better detectability than the storage phosphor plates used in this study. Substantial dose reduction could be achieved using anode/filter combination of W/Rh instead of Mo/Mo with no significant loss of information in the images.
引用
收藏
页码:591 / 597
页数:7
相关论文
共 21 条
  • [1] DEVELOPMENT OF AN ANTHROPOMORPHIC BREAST PHANTOM
    CALDWELL, CB
    YAFFE, MJ
    [J]. MEDICAL PHYSICS, 1990, 17 (02) : 273 - 280
  • [2] FREE-RESPONSE METHODOLOGY - ALTERNATE ANALYSIS AND A NEW OBSERVER-PERFORMANCE EXPERIMENT
    CHAKRABORTY, DP
    WINTER, LHL
    [J]. RADIOLOGY, 1990, 174 (03) : 873 - 881
  • [3] DIGITAL MAMMOGRAPHY - ROC STUDIES OF THE EFFECTS OF PIXEL SIZE AND UNSHARP-MASK FILTERING ON THE DETECTION OF SUBTLE MICROCALCIFICATIONS
    CHAN, HP
    VYBORNY, CJ
    MACMAHON, H
    METZ, CE
    DOI, K
    SICKLES, EA
    [J]. INVESTIGATIVE RADIOLOGY, 1987, 22 (07) : 581 - 589
  • [4] Visibility of microcalcifications in computed and screen-film mammography
    Cowen, AR
    Launders, JH
    Jadav, M
    Brettle, DS
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 1997, 42 (08) : 1533 - 1548
  • [5] DETECTION OF SUBTLE MICROCALCIFICATIONS - COMPARISON OF COMPUTED RADIOGRAPHY AND SCREEN-FILM MAMMOGRAPHY
    HIGASHIDA, Y
    MORIBE, N
    MORITA, K
    KATSUDA, N
    HATEMURA, M
    TAKADA, T
    TAKAHASHI, M
    YAMASHITA, J
    [J]. RADIOLOGY, 1992, 183 (02) : 483 - 486
  • [6] JARLMAN O, 1991, ACTA RADIOL, V32, P110
  • [7] JARLMAN O, 1994, CANCER DETECT PREV, V18, P375
  • [8] Digital mammography using storage phosphor plate technique - Optimizing image processing parameters for the visibility of lesions and anatomy
    Kheddache, S
    Kvist, H
    [J]. EUROPEAN JOURNAL OF RADIOLOGY, 1997, 24 (03) : 237 - 244
  • [9] MAMMOGRAMS OBTAINED WITH RHODIUM VS MOLYBDENUM ANODES - CONTRAST AND DOSE DIFFERENCES
    KIMMESMITH, C
    WANG, J
    DEBRUHL, N
    BASIC, M
    BASSETT, LW
    [J]. AMERICAN JOURNAL OF ROENTGENOLOGY, 1994, 162 (06) : 1313 - 1317
  • [10] THE INFLUENCE OF FOCAL SPOT SIZE ON IMAGE-RESOLUTION AND TEST PHANTOM SCORES IN MAMMOGRAPHY
    LAW, J
    [J]. BRITISH JOURNAL OF RADIOLOGY, 1993, 66 (785) : 441 - 446