Verification and correction of setup deviations in tangential breast irradiation using EPID: Gain versus workload

被引:31
作者
Pouliot, J
Lirette, A
机构
[1] Department of Radiation Oncology, Ctr. Hosp. Univ. de Québec, Pavillon l'Hotel-Dieu de Quebec, Que. G1R 2J6
关键词
treatment verification; correction; breast cancer; electronic portal imaging;
D O I
10.1118/1.597722
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
When a deviation in the treatment setup is identified, when or how should it be corrected? With the aim of improving the precision and reducing the systematic errors while maintaining the workload to a minimum, a study was performed to define the proper set of actions for the correction of tangential breast setup deviations during the course of the treatment. Patients and Methods: Clinical data were taken from a prospective study of more than 2200 on-line Electronic Portal Images (EPI) from 20 patients treated with tangential irradiation following a partial mastectomy for an early stage (I and II) breast cancer. The values of the central lung distance were then entered in the verification/correction procedure and modified to correct only the portion of the deviation most likely attributed to systematic errors according to the maximum likelihood. The verification/correction procedure uses an action level equal to FML . alpha/root N, where alpha is proportional to the standard deviation (alpha=n sigma) and N is the number of consecutive fractions delivered after the start of the treatment. FML is the fraction of the deviation due to the systematic errors estimated from the maximum likelihood of the two distributions. In addition to n, one needs to specify N-max, the maximum number of consecutive measurements without correction, to apply the procedure. The combination of these two parameters C(n,N-max) will determine the reduction of systematic errors (gain) and the number of measurements and corrections (workload) associated to the procedure. Results: The effects of three combinations, C(1,1), C(2,2), and C(3,4) were studied. Also, the analysis of the results after application of the procedure with and without the factor of maximum likelihood made individually for each patient demonstrates the importance of the FML. Conclusions: The verification/correction procedure with the inclusion of the FML can effectively improve the accuracy when applied to clinical data. With the specific workload related to measurements and corrections performed at our institution, we have found that in the case of the tangential breast treatment, the optimum combination uses an action level equal to 2 sigma and a maximum of two consecutive measurements without correction. (C) 1996 American Association of Physicists in Medicine.
引用
收藏
页码:1393 / 1398
页数:6
相关论文
共 29 条
[1]  
[Anonymous], 1978, 29 ICRU
[2]   A VERIFICATION PROCEDURE TO IMPROVE PATIENT SET-UP ACCURACY USING PORTAL IMAGES [J].
BEL, A ;
VANHERK, M ;
BARTELINK, H ;
LEBESQUE, JV .
RADIOTHERAPY AND ONCOLOGY, 1993, 29 (02) :253-260
[3]   MAXIMIZING SETUP ACCURACY USING PORTAL IMAGES AS APPLIED TO A CONFORMAL BOOST TECHNIQUE FOR PROSTATIC-CANCER [J].
BIJHOLD, J ;
LEBESQUE, JV ;
HART, AAM ;
VIJLBRIEF, RE .
RADIOTHERAPY AND ONCOLOGY, 1992, 24 (04) :261-271
[4]   ROLE OF PORTAL IMAGING IN CLINICAL RADIOTHERAPY - FLORENCE EXPERIENCE [J].
CIONINI, L ;
BUCCIOLINI, M .
RADIOTHERAPY AND ONCOLOGY, 1993, 29 (02) :230-236
[5]   ROUTINE CLINICAL ONLINE PORTAL IMAGING FOLLOWED BY IMMEDIATE FIELD ADJUSTMENT USING A TELE-CONTROLLED PATIENT COUCH [J].
DENEVE, W ;
VANDENHEUVEL, F ;
DEBEUKELEER, M ;
COGHE, M ;
THON, L ;
DEROOVER, P ;
VANLANCKER, M ;
STORME, G .
RADIOTHERAPY AND ONCOLOGY, 1992, 24 (01) :45-54
[6]   WHEN AND HOW CAN WE IMPROVE PRECISION IN RADIOTHERAPY [J].
DUTREIX, A .
RADIOTHERAPY AND ONCOLOGY, 1984, 2 (04) :275-292
[7]   TIME TREND OF PATIENT SETUP DEVIATIONS DURING PELVIC IRRADIATION USING ELECTRONIC PORTAL IMAGING [J].
ELGAYED, AAH ;
BEL, A ;
VIJLBRIEF, R ;
BARTELINK, H ;
LEBESQUE, JV .
RADIOTHERAPY AND ONCOLOGY, 1993, 26 (02) :162-171
[8]   DAILY MONITORING AND CORRECTION OF RADIATION-FIELD PLACEMENT USING A VIDEO-BASED PORTAL IMAGING-SYSTEM - A PILOT-STUDY [J].
EZZ, A ;
MUNRO, P ;
PORTER, AT ;
BATTISTA, J ;
JAFFRAY, DA ;
FENSTER, A ;
OSBORNE, S .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1992, 22 (01) :159-165
[9]   5-YEAR RESULTS OF A RANDOMIZED CLINICAL-TRIAL COMPARING TOTAL MASTECTOMY AND SEGMENTAL MASTECTOMY WITH OR WITHOUT RADIATION IN THE TREATMENT OF BREAST-CANCER [J].
FISHER, B ;
BAUER, M ;
MARGOLESE, R ;
POISSON, R ;
PILCH, Y ;
REDMOND, C ;
FISHER, E ;
WOLMARK, N ;
DEUTSCH, M ;
MONTAGUE, E ;
SAFFER, E ;
WICKERHAM, L ;
LERNER, H ;
GLASS, A ;
SHIBATA, H ;
DECKERS, P ;
KETCHAM, A ;
OISHI, R ;
RUSSELL, I .
NEW ENGLAND JOURNAL OF MEDICINE, 1985, 312 (11) :665-673
[10]   IMMOBILIZATION ERROR - SOME THEORETICAL CONSIDERATIONS [J].
GOITEIN, M ;
BUSSE, J .
RADIOLOGY, 1975, 117 (02) :407-412