Megavoltage CT imaging as a by-product of multileaf collimator leakage

被引:21
作者
Ruchala, KJ
Olivera, GH
Kapatoes, JM
Schloesser, EA
Reckwerdt, PJ
Mackie, TR
机构
[1] Univ Wisconsin, Dept Med Phys, Madison, WI 53706 USA
[2] Tomotherapy Inc, Middleton, WI 53562 USA
[3] Univ Wisconsin, Dept Human Oncol, Madison, WI 53706 USA
关键词
D O I
10.1088/0031-9155/45/7/401
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
In addition to their potential for the delivery of highly conformal radiation therapy treatments, tomotherapeutic treatments also Feature increased potential for verification. For example, megavoltage CT allows one to use the megavoltage linac to generate tomographic images of the patient in the treatment position. This is typically done before or after radiation therapy treatments. However, it is also possible to collect MVCT images entirely during the treatment itself. This process utilizes the leakage radiation through the closed leaves of the Nomos MIMiC MLC, along with slight inefficiencies in treatment delivery, to generate MVCT images during treatment that require neither additional time nor dose. The image quality is limited, yet sufficient to see a patient's external boundary, density differences over 8% for 25.0 mm objects and resolutions of 3.0 mm for high-contrast objects. Such images can potentially be viewed during treatment, used to flag additional CT immediately after the treatment and provide a representation of the patient's exact position during treatment for use with dose reconstruction.
引用
收藏
页码:N61 / N70
页数:10
相关论文
共 32 条
  • [1] Aoki Y, 1990, Radiat Med, V8, P107
  • [2] BALOG J, 1998, THESIS U WISCONSIN M
  • [3] RADIOTHERAPEUTIC COMPUTED-TOMOGRAPHY WITH SCANNED PHOTON BEAMS
    BRAHME, A
    LIND, B
    NAFSTADIUS, P
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1987, 13 (01): : 95 - 101
  • [4] THE IMPLEMENTATION OF PATIENT POSITION CORRECTION USING A MEGAVOLTAGE IMAGING DEVICE ON A LINEAR-ACCELERATOR
    EVANS, PM
    GILDERSLEVE, JQ
    RAWLINGS, C
    SWINDELL, W
    [J]. BRITISH JOURNAL OF RADIOLOGY, 1993, 66 (789) : 833 - 838
  • [5] IMAGE COMPARISON TECHNIQUES FOR USE WITH MEGAVOLTAGE IMAGING-SYSTEMS
    EVANS, PM
    GILDERSLEVE, JQ
    MORTON, EJ
    SWINDELL, W
    COLES, R
    FERRARO, M
    RAWLINGS, C
    XIAO, ZR
    DYER, J
    [J]. BRITISH JOURNAL OF RADIOLOGY, 1992, 65 (776) : 701 - 709
  • [6] Registration of synthetic tomographic projection data sets using cross-correlation
    Fitchard, EE
    Aldridge, JS
    Reckwerdt, PJ
    Mackie, TR
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 1998, 43 (06) : 1645 - 1657
  • [7] Registration using tomographic projection files
    Fitchard, EE
    Aldridge, JS
    Ruchala, K
    Fang, G
    Balog, J
    Pearson, DW
    Olivera, GH
    Schloesser, EA
    Wenman, D
    Reckwerdt, PJ
    Mackie, TR
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 1999, 44 (02) : 495 - 507
  • [8] Six parameter patient registration directly from projection data
    Fitchard, EE
    Aldridge, JS
    Reckwerdt, PJ
    Olivera, GH
    Mackie, TR
    Iosevich, A
    [J]. NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1999, 421 (1-2) : 342 - 351
  • [9] A RANDOMIZED TRIAL OF PATIENT REPOSITIONING DURING RADIOTHERAPY USING A MEGAVOLTAGE IMAGING-SYSTEM
    GILDERSLEVE, J
    DEARNALEY, DP
    EVANS, PM
    LAW, M
    RAWLINGS, C
    SWINDELL, W
    [J]. RADIOTHERAPY AND ONCOLOGY, 1994, 31 (02) : 161 - 168
  • [10] REPRODUCIBILITY OF PATIENT POSITIONING DURING ROUTINE RADIOTHERAPY, AS ASSESSED BY AN INTEGRATED MEGAVOLTAGE IMAGING-SYSTEM
    GILDERSLEVE, J
    DEARNALEY, DP
    EVANS, PM
    SWINDELL, W
    [J]. RADIOTHERAPY AND ONCOLOGY, 1995, 35 (02) : 151 - 160