Glucose sensor based on organic thin film transistor using glucose oxidase and conducting polymer

被引:148
作者
Liu, J. [1 ]
Agarwal, M. [1 ]
Varahramyan, K. [1 ]
机构
[1] Louisiana Tech Univ, Inst Micromfg, Ruston, LA 71272 USA
关键词
Glucose sensor; Glucose oxidase; Poly(3,4-ethylenedioxythiophene)-poly(styrene-sulfonate) (PEDOT-PSS); Cellulose acetate membrane; Organic thin film transistor (OTFT);
D O I
10.1016/j.snb.2008.08.009
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
An organic thin film transistor (OTFT)-based glucose sensor, with a channel consisting of glucose oxidase enzyme (GOx) immobilized with poly(3,4-ethylenedioxythiopliene-poly(styrene-Sulfonate) (PEDOT-PSS) conducting polymer film, has been investigated. GOx was immobilized on PEDOT-PSS conducting polymer film using a simple cost effective spin-coating technique, and was entrapped in the polymer matrix during electrochemical polymerization. Subsequently, the sensor was encapsulated within a cellulose acetate membrane to prevent dissolution of GOx and PEDOT-PSS matrix in glucose Solution (glucose in DI water). The resulting device characteristics were determined using a Keithley electrical measurement system. A linear increase in the OTFT drain Current was observed resulting from the increase in glucose concentration. The sensitivity of the developed glucose sensor was determined to be 1.65 mu A per 1 mM of glucose concentration. A linear range of response was found from 1.1 to 16.5 mM of glucose, with a response time of 10-20 s. The results obtained are indicative that the reported OTFT-based glucose sensor retains the enzyme bioactivity and can be applied as a glucose biosensor. (c) 2008 Elsevier B.V. All rights reserved.
引用
收藏
页码:195 / 199
页数:5
相关论文
共 29 条
[1]   Electroimmobilisation of sulphite oxidase into a polypyrrole film and its utilisation for flow amperometric detection of sulphite [J].
Adeloju, SB ;
Barisci, JN ;
Wallace, GG .
ANALYTICA CHIMICA ACTA, 1996, 332 (2-3) :145-153
[2]   Application of octadecanethiol self-assembled monolayer to cholesterol biosensor based on surface plasmon resonance technique [J].
Arya, Sunil K. ;
Solanki, Pratima R. ;
Singh, Ravindra P. ;
Pandey, Manoj K. ;
Datta, Monika ;
Malhotra, Bansi D. .
TALANTA, 2006, 69 (04) :918-926
[3]   Glucose oxidase electrodes of a terpolymer poly(aniline-co-o-anisidine-co-o-toluidine) as biosensors [J].
Borole, DD ;
Kapadi, UR ;
Mahulikar, PP ;
Hundiwale, DG .
EUROPEAN POLYMER JOURNAL, 2005, 41 (09) :2183-2188
[4]   Glucose oxidase electrodes of poly(o-anisidine), poly(o-toluidine), and their copolymer as biosensors:: A comparative study [J].
Borole, DD ;
Kapadi, UR ;
Mahulikar, PP ;
Hundiwale, DG .
JOURNAL OF APPLIED POLYMER SCIENCE, 2004, 94 (05) :1877-1884
[5]   Printed electrochemical devices using conducting polymers as active materials on flexible substrates [J].
Chen, MX .
PROCEEDINGS OF THE IEEE, 2005, 93 (07) :1339-1347
[6]   Electric-field induced ion-leveraged metal-insulator transition in conducting polymer-based field effect devices [J].
Epstein, AJ ;
Hsu, FC ;
Chiou, NR ;
Prigodin, VN .
CURRENT APPLIED PHYSICS, 2002, 2 (04) :339-343
[7]   Hybrid bioelectrocatalyst for hydrogen peroxide reduction: Immobilization of enzyme within organic-inorganic film of structured Prussian Blue and PEDOT [J].
Ernst, Andrzej ;
Makowski, Oktawian ;
Kowalewska, Barbara ;
Miecznikowski, Krzysztof ;
Kulesza, Pawel J. .
BIOELECTROCHEMISTRY, 2007, 71 (01) :23-28
[8]   Immobilization of GOD on electrochemically synthesized Ppy-PVS composite film by cross-linking via glutaraldehyde for determination of glucose [J].
Gade, V. K. ;
Shirale, D. J. ;
Gaikwad, P. D. ;
Savale, P. A. ;
Kakde, K. P. ;
Kharat, H. J. ;
Shirsat, M. D. .
REACTIVE & FUNCTIONAL POLYMERS, 2006, 66 (12) :1420-1426
[9]   Application of a multi-sulfonamide biosensor immunoassay for the detection of sulfadiazine and sulfamethoxazole residues in broiler serum and its use as a predictor of the levels in edible tissue [J].
Haasnoot, W ;
Bienenmann-Ploum, M ;
Lamminmäki, U ;
Swanenburg, M ;
van Rhijn, H .
ANALYTICA CHIMICA ACTA, 2005, 552 (1-2) :87-95
[10]   INTEGRATED MULTI-BIOSENSORS BASED ON AN ION-SENSITIVE FIELD-EFFECT TRANSISTOR USING PHOTOLITHOGRAPHIC TECHNIQUES [J].
HANAZATO, Y ;
NAKAKO, M ;
SHIONO, S ;
MAEDA, M .
IEEE TRANSACTIONS ON ELECTRON DEVICES, 1989, 36 (07) :1303-1310