Attenuation-Based Automatic Kilovolt Selection in Abdominal Computed Tomography Effects on Radiation Exposure and Image Quality

被引:58
作者
Eller, Achim [1 ]
May, Matthias S. [1 ]
Scharf, Michael [1 ]
Schmid, Axel [1 ]
Kuefner, Michael [1 ]
Uder, Michael [1 ,2 ]
Lell, Michael M. [1 ,2 ]
机构
[1] Univ Erlangen Nurnberg, Dept Radiol, D-91054 Erlangen, Germany
[2] Imaging Sci Inst, Erlangen, Germany
关键词
computed tomography; dose reduction; tube potential; image quality; radiation dose; DOSE REDUCTION STRATEGY; DOUBLE-STRAND BREAKS; LOW-TUBE-VOLTAGE; ITERATIVE RECONSTRUCTION; MULTIDETECTOR CT; PATIENT SIZE; SPIRAL CT; ANGIOGRAPHY; OPTIMIZATION; PROTOCOLS;
D O I
10.1097/RLI.0b013e318260c5d6
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
100231 [临床病理学]; 100902 [航空航天医学];
摘要
Objectives: Dose reduction has become a major issue in computed tomography (CT). The benefit of kilovolt (kV) reduction has been demonstrated in CT angiography. We sought to evaluate an attenuation-based fully automated kV-selection and milliampere second-adaption algorithm for CT and to assess radiation dose and image quality in comparison with a standard 120 kV protocol in contrast-enhanced (CE) portal-venous thoracoabdominal imaging. Materials and Methods: One hundred patients (mean age, 58.4 +/- 5.7 years; mean body mass index [BMI], 26.1 +/- 5.1 kg/m(2)) underwent CE CT using automated selection of the tube potential (80-140 kV) with milliampere second adaption based on the attenuation profile of the scout scan. The estimated CT dose index was recorded for the proposed scan setting and standard 120-kV protocol. Regions of interest measurements were performed at different locations for objective assessment of image quality. Signal-to-noise ratio and contrast-to-noise ratio (CNR) were calculated. The subjective image quality was assessed by 2 observers with a 4-point scale using previous CT examinations with the 120-kV standard protocol as the reference for comparison. Results: The kV-selection algorithm could be applied in all examinations without problems. Image quality was high, and there were no significant differences compared with previous examinations of the patients performed at 120 kV. Eighty kilovolts was used in 9% of examinations (mean BMI, 22.8 +/- 2.8 kg/m(2)); 100 kV, in 75% (mean BMI, 23.7 +/- 4.7 kg/m(2)); 120 kV, in 16% (mean BMI, 30 +/- 3.3 kg/m(2)); and 140 kV, in a single case (BMI, 49.4 kg/m(2)). The average estimated CT dose index reduction was 25.3% in the 80-kV group, 14.5% in the 100-kV group, and 11.4% overall. The CNR did not differ significantly, whereas the signal-to-noise ratio was significantly higher in the 80- and 100-kV examinations. Conclusion: The attenuation-based kV-selection algorithm was demonstrated to be applicable in clinical routine of CE thoracoabdominal CT, to keep CNR constant, and to result in a significant dose reduction while preserving image quality.
引用
收藏
页码:559 / 565
页数:7
相关论文
共 42 条
[1]
Coronary computed tomography angiography with a consistent dose below 1 mSv using prospectively electrocardiogram-triggered high-pitch spiral acquisition [J].
Achenbach, Stephan ;
Marwan, Mohamed ;
Ropers, Dieter ;
Schepis, Tiziano ;
Pflederer, Tobias ;
Anders, Katharina ;
Kuettner, Axel ;
Daniel, Werner G. ;
Uder, Michael ;
Lell, Michael M. .
EUROPEAN HEART JOURNAL, 2010, 31 (03) :340-346
[2]
Radiation dose of cardiac computed tomography - what has been achieved and what needs to be done [J].
Alkadhi, Hatem ;
Leschka, Sebastian .
EUROPEAN RADIOLOGY, 2011, 21 (03) :505-509
[3]
Improvement of image quality of multislice spiral CT scans of the head and neck region using a raw data-based multidimensional adaptive filtering (MAF) technique [J].
Baum, U ;
Anders, K ;
Steinbichler, G ;
Lell, M ;
Greess, H ;
Riedel, T ;
Kachelriess, M ;
Kalender, WA ;
Bautz, WA .
EUROPEAN RADIOLOGY, 2004, 14 (10) :1873-1881
[4]
Iterative reconstruction in image space (IRIS) in cardiac computed tomography: initial experience [J].
Bittencourt, Marcio Sommer ;
Schmidt, Bernhard ;
Seltmann, Martin ;
Muschiol, Gerd ;
Ropers, Dieter ;
Daniel, Werner Guenther ;
Achenbach, Stephan .
INTERNATIONAL JOURNAL OF CARDIOVASCULAR IMAGING, 2011, 27 (07) :1081-1087
[5]
Using 100-instead of 120-kVp computed tomography to diagnose pulmonary embolism almost halves the radiation dose with preserved diagnostic quality [J].
Bjorkdahl, Peter ;
Nyman, Ulf .
ACTA RADIOLOGICA, 2010, 51 (03) :260-270
[6]
Current concepts - Computed tomography - An increasing source of radiation exposure [J].
Brenner, David J. ;
Hall, Eric J. .
NEW ENGLAND JOURNAL OF MEDICINE, 2007, 357 (22) :2277-2284
[7]
CT Radiation Dose: What Can You Do Right Now in Your Practice? [J].
Coakley, Fergus V. ;
Gould, Robert ;
Yeh, Benjamin M. ;
Arenson, Ronald L. .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2011, 196 (03) :619-625
[8]
Projected Cancer Risks From Computed Tomographic Scans Performed in the United States in 2007 [J].
de Gonzalez, Amy Berrington ;
Mahesh, Mahadevappa ;
Kim, Kwang-Pyo ;
Bhargavan, Mythreyi ;
Lewis, Rebecca ;
Mettler, Fred ;
Land, Charles .
ARCHIVES OF INTERNAL MEDICINE, 2009, 169 (22) :2071-2077
[9]
Effects of Adaptive Section Collimation on Patient Radiation Dose in Multisection Spiral CT [J].
Deak, Paul D. ;
Langner, Oliver ;
Lell, Michael ;
Kalender, Willi A. .
RADIOLOGY, 2009, 252 (01) :140-147
[10]
CT iterative reconstruction in image space: A phantom study [J].
Ghetti, C. ;
Ortenzia, O. ;
Serreli, G. .
PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2012, 28 (02) :161-165