Algorithms for scanned probe microscope image simulation, surface reconstruction, and tip estimation

被引:520
作者
Villarrubia, JS [1 ]
机构
[1] Natl Inst Stand & Technol, Gaithersburg, MD 20899 USA
关键词
algorithms; atomic force microscopy; blind reconstruction; dimensional metrology; image simulation; mathematical morphology; scanned probe microscopy; scanning tunneling microscopy; surface reconstruction; tip artifacts; tip estimation;
D O I
10.6028/jres.102.030
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
To the extent that tips are not perfectly sharp, images produced by scanned probe microscopies (SPM) such as atomic force microscopy and scanning tunneling microscopy are only approximations of the specimen surface, np-induced distortions are significant whenever the specimen contains features with aspect ratios comparable to the tip's. Treatment of the tip-surface interaction as a simple geometrical exclusion allows calculation of many quantities important for SPM dimensional metrology. Algorithms for many of these are provided here, including the following: (1) calculating an image given a specimen and a tip (dilation), (2) reconstructing the specimen surface given its image and the tip (erosion), (3) reconstructing the tip shape from the image of a known "tip characterizer" (erosion again), and (4) estimating the tip shape from an image of an unknown tip characterizer (blind reconstruction). Blind reconstruction, previously demonstrated only for simulated noiseless images, is here extended to images with noise or other experimental artifacts. The main body of the paper serves as a programmer's and user's guide. It includes theoretical background for all of the algorithms, detailed discussion of some algorithmic problems of interest to programmers, and practical recommendations for users.
引用
收藏
页码:425 / 454
页数:30
相关论文
共 41 条
[11]   DIMENSIONAL METROLOGY WITH SCANNING PROBE MICROSCOPES [J].
GRIFFITH, JE ;
MARCHMAN, HM ;
MILLER, GL ;
HOPKINS, LC .
JOURNAL OF VACUUM SCIENCE & TECHNOLOGY B, 1995, 13 (03) :1100-1105
[12]   PROBE CHARACTERIZATION FOR SCANNING PROBE METROLOGY [J].
GRIGG, DA ;
RUSSELL, PE ;
GRIFFITH, JE ;
VASILE, MJ ;
FITZGERALD, EA .
ULTRAMICROSCOPY, 1992, 42 :1616-1620
[13]   IMAGE-ANALYSIS USING MATHEMATICAL MORPHOLOGY [J].
HARALICK, RM ;
STERNBERG, SR ;
ZHUANG, XH .
IEEE TRANSACTIONS ON PATTERN ANALYSIS AND MACHINE INTELLIGENCE, 1987, 9 (04) :532-550
[14]   RECONSTRUCTION OF STM AND AFM IMAGES DISTORTED BY FINITE-SIZE TIPS [J].
KELLER, D .
SURFACE SCIENCE, 1991, 253 (1-3) :353-364
[15]   ENVELOPE RECONSTRUCTION OF PROBE MICROSCOPE IMAGES [J].
KELLER, DJ ;
FRANKE, FS .
SURFACE SCIENCE, 1993, 294 (03) :409-419
[16]  
LOWNEY JR, 1994, P SOC PHOTO-OPT INS, V2194, P85, DOI 10.1117/12.174167
[17]   ATOMIC-FORCE MICROSCOPY PROBE TIP VISUALIZATION AND IMPROVEMENT OF IMAGES USING A SIMPLE DECONVOLUTION PROCEDURE [J].
MARKIEWICZ, P ;
GOH, MC .
LANGMUIR, 1994, 10 (01) :5-7
[18]   A ROCKING BEAM ELECTROSTATIC BALANCE FOR THE MEASUREMENT OF SMALL FORCES [J].
MILLER, GL ;
GRIFFITH, JE ;
WAGNER, ER ;
GRIGG, DA .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1991, 62 (03) :705-709
[19]   SUBMICROMETER LINEWIDTH METROLOGY IN THE OPTICAL MICROSCOPE [J].
NYYSSONEN, D ;
LARRABEE, RD .
JOURNAL OF RESEARCH OF THE NATIONAL BUREAU OF STANDARDS, 1987, 92 (03) :187-204
[20]   TIP FINITE-SIZE EFFECTS ON ATOMIC-FORCE MICROSCOPY IN THE CONTACT MODE - SIMPLE GEOMETRICAL CONSIDERATIONS FOR RAPID ESTIMATION OF APEX RADIUS AND TIP ANGLE BASED ON THE STUDY OF POLYSTYRENE LATEX BALLS [J].
ODIN, C ;
AIME, JP ;
ELKAAKOUR, Z ;
BOUHACINA, T .
SURFACE SCIENCE, 1994, 317 (03) :321-340