A miniaturized liquid core waveguide-capillary electrophoresis system with flow injection sample introduction and fluorometric detection using light-emitting diodes

被引:121
作者
Wang, SL
Huang, XJ
Fang, ZL
Dasgupta, PK
机构
[1] Northeastern Univ, Res Ctr Analyt Sci, Shenyang 110006, Peoples R China
[2] Texas Tech Univ, Dept Chem & Biochem, Lubbock, TX 79409 USA
关键词
D O I
10.1021/ac010341a
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A novel miniaturized capillary electrophoresis (CE) system is described where a Teflon AF-coated silica capillary serves both as the separation channel and as a transversely illuminated liquid core waveguide. This device uniquely uses flow injection (FI)-based split-flow sample introduction through a falling-drop interface. An H-channel structure fixed on a microscope glass slide utilizes a horizontal separation capillary with tubular sidearms on each end that serve as inlet and outlet flow-through electrode reservoirs. The inlet reservoir also functions as a falling-drop interface for coupling to the Fl system. A blue LED is used as excitation source. A large-core optical fiber takes the emitted fluorescence to an inexpensive PMT with two layers of green plastic used for optical filtering. No focusing arrangement is needed. Continuous FI introduction of a series of 30-muL samples containing a mixture of of fluorescein isothiocyanate (FITC)-labeled amino acids allowed a throughput rate up to 144 samples/ h, with similar to2% carryover and good precision (3.2% RSD). Baseline separation was achieved for FITC-labeled arginine, phenylalanine, glycine, and FITC in sodium tetraborate buffer (pH 9.5) with plate heights of 5.4-5.5 mum and plate numbers of 2.34 x 10(4)-2.37 x 10(4) under electrical field strengths of 214 V/cm for injection and 500 V/cm for separation (14-cm capillary, 48-mum i.d.). Detection limits (S/N = 3) were 1.3 muM for arginine and 1.9 muM for phenylalanine and glycine.
引用
收藏
页码:4545 / 4549
页数:5
相关论文
共 13 条
[1]   Luminescence detection with a liquid care waveguide [J].
Dasgupta, PK ;
Zhang, GF ;
Li, JZ ;
Boring, CB ;
Jambunathan, S ;
Al-Horr, R .
ANALYTICAL CHEMISTRY, 1999, 71 (07) :1400-1407
[2]   GLASS CHIPS FOR HIGH-SPEED CAPILLARY ELECTROPHORESIS SEPARATIONS WITH SUBMICROMETER PLATE HEIGHTS [J].
EFFENHAUSER, CS ;
MANZ, A ;
WIDMER, HM .
ANALYTICAL CHEMISTRY, 1993, 65 (19) :2637-2642
[3]   Sequential injection sample introduction microfluidic-chip based capillary electrophoresis system [J].
Fang, Q ;
Wang, FR ;
Wang, SL ;
Liu, SS ;
Xu, SK ;
Fang, ZL .
ANALYTICA CHIMICA ACTA, 1999, 390 (1-3) :27-37
[4]   Combination of flow injection with capillary electrophoresis .1. The basic system [J].
Fang, ZL ;
Liu, ZS ;
Shen, Q .
ANALYTICA CHIMICA ACTA, 1997, 346 (02) :135-143
[5]   Developments in flow injection-capillary electrophoresis systems [J].
Fang, ZL ;
Chen, HW ;
Fang, Q ;
Pu, QS .
ANALYTICAL SCIENCES, 2000, 16 (02) :197-203
[6]  
FANG ZL, IN PRESS J ANAL CHEM
[7]   Combination of flow injection with capillary electrophoresis - Part 7. Microchip capillary electrophoresis system with flow injection sample introduction and amperometric detection [J].
Fu, CG ;
Fang, ZL .
ANALYTICA CHIMICA ACTA, 2000, 422 (01) :71-79
[8]   Capillary electrophoresis system with flow injection sample introduction and chemiluminescence detection on a chip platform [J].
Huang, XJ ;
Pu, QS ;
Fang, ZL .
ANALYST, 2001, 126 (03) :281-284
[9]   New interface for coupling flow-injection and capillary electrophoresis [J].
Kuban, P ;
Engstrom, A ;
Olsson, JC ;
Thorsen, G ;
Tryzell, R ;
Karlberg, B .
ANALYTICA CHIMICA ACTA, 1997, 337 (02) :117-124
[10]   Measurement of atmospheric hydrogen peroxide and hydroxymethyl hydroperoxide with a diffusion scrubber and light emitting diode-liquid core waveguide-based fluorometry [J].
Li, JZ ;
Dasgupta, PK .
ANALYTICAL CHEMISTRY, 2000, 72 (21) :5338-5347