Clinical wear measurement on low contact stress rotating platform knee bearings

被引:38
作者
Atwood, Sara A. [1 ]
Currier, John H. [1 ]
Mayor, Michael B. [1 ]
Collier, John P. [1 ]
Van Citters, Douglas W. [1 ]
Kennedy, Francis E. [1 ]
机构
[1] Dartmouth Coll, Thayer Sch Engn, Hanover, NH 03755 USA
关键词
TKA; mobile bearings; wear; polyethylene;
D O I
10.1016/j.arth.2007.06.005
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Whereas fixed-bearing total knee arthroplasty (TKA) designs secure the polyethylene bearing to the tibial tray, mobile-bearing TKAs allow the bearing to move relative to the tray. This study evaluated wear performance of the rotational articulation of the Low Contact Stress Rotating Platform mobile-bearing TKA (DePuy, Warsaw, Ind) by analyzing 100 retrievals. All retrieved bearings showed rotation surface damage, but severity of the damage did not correlate with duration. Rotation surface damage appeared to be caused by contaminant particles, which produced curvilinear scratches that were longer than the normal rotational excursion of the knee. Wear measurement indicated that wear was relatively uniform, long-term wear rates were low (mean, 54 mm(3)/y for durations >2 years) and decreased with longer duration, and damaged appearance did not correspond to high wear.
引用
收藏
页码:431 / 440
页数:10
相关论文
共 24 条
[1]  
ATWOOD SA, 2005, J TRIBOL, V128, P275
[2]   In vivo kinematics of cruciate-retaining and -substituting knee arthroplasties [J].
Banks, SA ;
Markovich, GD ;
Hodge, WA .
JOURNAL OF ARTHROPLASTY, 1997, 12 (03) :297-304
[3]   Factors affecting the severity of backside wear of modular tibial inserts [J].
Conditt, MA ;
Stein, JA ;
Noble, PC .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2004, 86A (02) :305-311
[4]   Polyethylene damage on the nonarticular surface of modular total knee prostheses [J].
Cuckler, JM ;
Lemons, J ;
Tamarapalli, JR ;
Beck, P .
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2003, (410) :248-253
[5]  
Currier BH, 2000, J BIOMED MATER RES, V53, P143, DOI 10.1002/(SICI)1097-4636(2000)53:2<143::AID-JBM3>3.0.CO
[6]  
2-5
[7]   Analysis of wear asymmetry in a series of 94 retrieved polyethylene tibial bearings [J].
Currier, JH ;
Bill, MA ;
Mayor, MB .
JOURNAL OF BIOMECHANICS, 2005, 38 (02) :367-375
[8]   Serial measurement of polyethylene wear of well-fixed cementless metal-backed acetabular component in total hip arthroplasty: An over 10 year follow-up study [J].
Dai, XS ;
Omori, H ;
Okumura, Y ;
Ando, M ;
Oki, H ;
Hashimoto, N ;
Baba, H .
ARTIFICIAL ORGANS, 2000, 24 (09) :746-751
[9]   Mobile-beading total knee arthroplasty - Do the polyethylene bearings rotate? [J].
Dennis, DA ;
Komistek, RD ;
Mahfouz, MR ;
Outten, JT ;
Sharma, A .
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2005, (440) :88-95
[10]  
Goodfellow JW, 2002, LCS MOBILE BEARING K, P3