A study of a bio-mimetic recognition material for the BAW sensor by molecular imprinting and its application for the determination of paracetamol in the human serum and urine

被引:52
作者
Tan, YG
Zhou, ZL
Wang, P
Nie, LH
Yao, SZ [1 ]
机构
[1] Hunan Univ, Coll Chem & Chem Engn, Changsha 410082, Peoples R China
[2] Cent S Univ, Xiangya Med Coll, Changsha 410078, Peoples R China
基金
中国国家自然科学基金;
关键词
bulk acoustic wave; piezoelectric quartz crystal; bio-mimetic sensor; molecularly imprinted polymers; molecular recognition; paracetamol;
D O I
10.1016/S0039-9140(01)00442-8
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A new bio-mimetic bulk acoustic wave (BAW) sensor for paracetamol, with a high selectivity and sensitivity, was fabricated by using the molecularly imprinted polymer (MIP) as the sensing material. Non-covalent molecular imprinting polymers were synthesized simultaneously using two different functional monomers, namely the weakly basic 4-vinylpyridine (4-VP) and the acidic and hydrogen binding methacrylic acid (MAA). Such imprinted polymers improved recognition capability as compared with the polymers that were synthesized using only one of the functional monomers. Dynamic quartz crystal impedance analysis was employed to study the effect of the viscoelasticity of the polymer coating on the measurement. There was no variation in the viscoelasticity during the detection. The sensor was stable and exhibited good reproducibility. Satisfactory results were obtained in the determination of paracetamol in real samples. (C) 2001 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:337 / 347
页数:11
相关论文
共 42 条
[11]   Piezoelectric quartz crystal biosensors [J].
Bunde, RL ;
Jarvi, EJ ;
Rosentreter, JJ .
TALANTA, 1998, 46 (06) :1223-1236
[12]   Rational catalyst design via imprinted nanostructured materials [J].
Davis, ME ;
Katz, A ;
Ahmad, WR .
CHEMISTRY OF MATERIALS, 1996, 8 (08) :1820-1839
[13]  
DECORDORA MLA, 2000, J PHARM BIOMED ANAL, V22, P59
[14]   A GOAT-ANTI-HUMAN IGG MODIFIED PIEZOIMMUNOSENSOR FOR STAPHYLOCOCCUS-AUREUS DETECTION [J].
LE, D ;
HE, FJ ;
JIANG, TJ ;
NIE, LH ;
YAO, SZ .
JOURNAL OF MICROBIOLOGICAL METHODS, 1995, 23 (02) :229-234
[15]  
Dickert FL, 1998, ADV MATER, V10, P149, DOI 10.1002/(SICI)1521-4095(199801)10:2<149::AID-ADMA149>3.0.CO
[16]  
2-2
[17]  
EBERSOLE RC, 1990, J AM CHEM SOC, V112, P173
[18]   Determination of acetaminophen, caffeine, and butalbital in a commercial tablet dosage form by micellar electrokinetic chromatography [J].
Haque, A ;
Stewart, JT .
JOURNAL OF LIQUID CHROMATOGRAPHY & RELATED TECHNOLOGIES, 1999, 22 (14) :2159-2166
[19]   SURFACE ACOUSTIC-WAVE METHOD FOR INSITU DETERMINATION OF THE AMOUNTS OF ENZYME SUBSTRATE COMPLEX FORMED ON IMMOBILIZED GLUCOSE-OXIDASE DURING CATALYTIC REACTION [J].
INOUE, Y ;
KATO, Y ;
SATO, K .
JOURNAL OF THE CHEMICAL SOCIETY-FARADAY TRANSACTIONS, 1992, 88 (03) :449-454
[20]   A NOVEL IMMUNOSENSOR FOR HERPES VIRUSES [J].
KONIG, B ;
GRATZEL, M .
ANALYTICAL CHEMISTRY, 1994, 66 (03) :341-344