Background and purpose: To describe the clinical implementation of dynamic multileaf collimation (DMLC). Custom compensated four-field treatments of carcinoma of the bladder have been used as a simple test site for the introduction of intensity modulated radiotherapy. Materials and Methods: Compensating intensity modulations are calculated from computed tomography (CT) data, accounting for scattered, as well as primary radiation. Modulations are converted to multileaf collimator (MLC) leaf and jaw settings for dynamic delivery on a linear accelerator. A full dose calculation is carried out, accounting for dynamic leaf and jaw motion and transmission through these components. Before treatment, a test run of the delivery is performed and an absolute dose measurement made in a water or solid water phantom. Treatments are verified by in vivo diode measurements and real-rime electronic portal imaging. Results: Seven patients have been treated using DMLC. The technique improves dose homogeneity within the target volume, reducing high dose areas and compensating for loss of scatter at the beam edge. A typical total treatment time is 20 min. Conclusions: Compensated bladder treatments have proven an effective test site for DMLC in an extremely busy clinic. (C) 2001 Elsevier Science Ireland Ltd. All rights reserved.
引用
收藏
页码:31 / 38
页数:8
相关论文
共 20 条
[1]   Temporal resolution requirements for intensity modulated radiation therapy delivered by multileaf collimators [J].
Budgell, GJ .
PHYSICS IN MEDICINE AND BIOLOGY, 1999, 44 (07) :1581-1596
[2]   Rectangular edge synchronization for intensity modulated radiation therapy with dynamic multileaf collimation [J].
Budgell, GJ ;
Sykes, JR ;
Wilkinson, JM .
PHYSICS IN MEDICINE AND BIOLOGY, 1998, 43 (10) :2769-2784
[3]   Requirements for leaf position accuracy for dynamic multileaf collimation [J].
Budgell, GJ ;
Mott, JHL ;
Williams, PC ;
Brown, KJ .
PHYSICS IN MEDICINE AND BIOLOGY, 2000, 45 (05) :1211-1227
[4]   Planning, delivery, and quality assurance of intensity-modulated radiotherapy using dynamic multileaf collimator: A strategy for large-scale implementation for the treatment of carcinoma of the prostate [J].
Burman, C ;
Chui, CS ;
Kutcher, G ;
Leibel, S ;
Zelefsky, M ;
LoSasso, T ;
Spirou, S ;
Wu, QW ;
Yang, J ;
Stein, J ;
Mohan, R ;
Fuks, Z ;
Ling, CC .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1997, 39 (04) :863-873
[5]   DOSE CALCULATION FOR PHOTON BEAMS WITH INTENSITY MODULATION GENERATED BY DYNAMIC JAW OR MULTILEAF COLLIMATIONS [J].
CHUI, CS ;
LOSASSO, T ;
SPIROU, S .
MEDICAL PHYSICS, 1994, 21 (08) :1237-1244
[6]   Testing of dynamic multileaf collimation [J].
Chui, CS ;
Spirou, S ;
LoSasso, T .
MEDICAL PHYSICS, 1996, 23 (05) :635-641
[7]   In vivo dosimetry during external photon beam radiotherapy [J].
Essers, M ;
Mijnheer, BJ .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1999, 43 (02) :245-259
[8]   Improving dose homogeneity in routine head and neck radiotherapy with custom 3-D compensation [J].
Harari, PM ;
Sharda, NN ;
Brock, LK ;
Paliwal, BR .
RADIOTHERAPY AND ONCOLOGY, 1998, 49 (01) :67-71
[9]   COMPUTER-ASSISTED GENERATION OF MULTI-LEAF COLLIMATOR SETTINGS FOR CONFORMATION THERAPY [J].
HOUNSELL, AR ;
SHARROCK, PJ ;
MOORE, CJ ;
SHAW, AJ ;
WILKINSON, JM ;
WILLIAMS, PC .
BRITISH JOURNAL OF RADIOLOGY, 1992, 65 (772) :321-326
[10]   Head scatter modelling for irregular field shaping and beam intensity modulation [J].
Hounsell, AR ;
Wilkinson, JM .
PHYSICS IN MEDICINE AND BIOLOGY, 1997, 42 (09) :1737-1749