Incidence of electromagnetic interference in implantable cardioverter defibrillators

被引:42
作者
Kolb, C
Zrenner, B
Schmitt, C
机构
[1] Deutsch Herzzentrum Munchen, D-80636 Munich, Germany
[2] Tech Univ Munich, Klinikum Rechts Isar, Med Klin 1, D-8000 Munich, Germany
来源
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY | 2001年 / 24卷 / 04期
关键词
electromagnetic interference; implantable cardioverter defibrillators;
D O I
10.1046/j.1460-9592.2001.00465.x
中图分类号
R5 [内科学];
学科分类号
1002 [临床医学]; 100201 [内科学];
摘要
Electromagnetic interference (EMI) with ICDs can lead to temporary inhibition of the device or to inappropriate delivery of antitachycardia pacing and shocks. The incidence of interactions between electronic devices and the current generation of ICDs is not known. In a retrospective study of 341 patients (665 patient-years) who underwent a regular follow-up every 3 months, five episodes of EMI were detected in four different patients. The risk for receiving inappropriate shocks due to EMI is < 1 % per year and patient. In conclusion, although inappropriate delivery of shocks by ICDs due to EMI rarely occurs, patient information should emphasize the avoidance of situations of possible interference. Further efforts concerning lead technology and detection algorithms are necessary to minimize the risk of EMI.
引用
收藏
页码:465 / 468
页数:4
相关论文
共 14 条
[1]
Electromagnetic interference of digital and analog cellular telephones with implantable cardioverter defibrillators: In vitro and in vivo studies [J].
Barbaro, V ;
Bartolini, P ;
Bellocci, F ;
Caruso, F ;
Donato, A ;
Gabrielli, D ;
Militello, C ;
Montenero, AS ;
Zecchi, P .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1999, 22 (04) :626-634
[2]
Cellular phone interference testing of implantable cardiac defibrillators in vitro [J].
Bassen, HI ;
Moore, HJ ;
Ruggera, PS .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1998, 21 (09) :1709-1715
[3]
Electromagnetic interference from a muscle stimulation device causing discharge of an implantable cardioverter defibrillator: Epicardial bipolar and endocardial bipolar sensing circuits are compared [J].
Glotzer, TV ;
Gordon, M ;
Sparta, M ;
Radoslovich, G ;
Zimmerman, J .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1998, 21 (10) :1996-1998
[4]
Interactions between electronic article surveillance systems and implantable cardioverter-defibrillators [J].
Groh, WJ ;
Boschee, SA ;
Engelstein, ED ;
Miles, WM ;
Burton, ME ;
Foster, PR ;
Crevey, BJ ;
Zipes, DP .
CIRCULATION, 1999, 100 (04) :387-392
[5]
VENTRICULAR-FIBRILLATION INDUCED BY LOW-ENERGY SHOCKS FROM PROGRAMMABLE IMPLANTABLE CARDIOVERTER-DEFIBRILLATORS IN PATIENTS WITH CORONARY-ARTERY DISEASE [J].
LAUER, MR ;
YOUNG, C ;
LIEM, LB ;
OTTOBONI, L ;
PETERSON, J ;
GOOLD, P ;
SUNG, RJ .
AMERICAN JOURNAL OF CARDIOLOGY, 1994, 73 (08) :559-563
[6]
Dysfunction of implantable defibrillators caused by slot machines [J].
Madrid, A ;
Sanchez, A ;
Bosch, E ;
Fernandez, E ;
Serrano, CM .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1997, 20 (01) :212-214
[7]
Implantable cardioverter defibrillator - An unusual case of inappropriate discharge during showering [J].
Manolis, AG ;
Katsivas, AG ;
Vassilopoulos, CV ;
Louvros, NE .
JOURNAL OF INTERVENTIONAL CARDIAC ELECTROPHYSIOLOGY, 2000, 4 (01) :265-268
[8]
Interaction between electronic article surveillance systems and implantable defibrillators: Insights from a fourth generation ICD [J].
Mathew, P ;
Lewis, C ;
Neglia, J ;
Krol, RB ;
Saksena, S .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1997, 20 (11) :2857-2859
[9]
Environmental electromagnetic interference from electronic article surveillance devices: Interactions with an ICD [J].
McIvor, ME .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1995, 18 (12) :2229-2230
[10]
PIUSKY SL, 1995, CIRCULATION, V92, P1651