摘要
A conducting polymer [polypyrrole (PPy)]-based amperometric biosensor fabricated on a platinum-coated nanoporous alumina electrode has been described. This fabricating process introduced artificial porosity into the PPy film, and the template pore sizes were carefully chosen to match the size of the glucose oxidase (GOx) molecule. The PF 6 -doped PPy film was synthesized with 0.05 M pyrrole and 0.1 M NaPF6 at a current density of 0.3 mA/CM2 for 90 s. Immobilization was done by physically adsorbing 5 pL of GOx on the nanoporous PPy film. Glutaraldehyde (0.1 wt.%, 5 pL) was used for cross-finking. The synthesized films were characterized by using an electrochemical technique and scanning electron microscopy (SEM). Amperometric responses were measured as a function of different concentrations of glucose at 0.4 V. Nanoporous electrodes lead to high enzyme loading, whereas the use of a cross-linking agent increased stability, sensitivity, reproducibility, repeatability, and shelf life.
引用
收藏
页码:1621 / 1626
页数:6
相关论文
共 19 条
[1]   Biomolecular immobilization on conducting polymers for biosensing applications [J].
Ahuja, Tarushee ;
Mir, Irfan Ahmad ;
Kumar, Devendra ;
Rajesh .
BIOMATERIALS, 2007, 28 (05) :791-805
[2]   Electrochemical adsorption of glucose oxidase onto polypyrrole film for the construction of a glucose biosensor [J].
Cho, JH ;
Shin, MC ;
Kim, HS .
SENSORS AND ACTUATORS B-CHEMICAL, 1996, 30 (02) :137-141
[3]   A glucose biosensor based on enzyme entrapment within polypyrrole films electrodeposited on mesoporous titanium dioxide [J].
Cosnier, S ;
Senillou, A ;
Grätzel, M ;
Comte, P ;
Vlachopoulos, N ;
Renault, NJ ;
Martelet, C .
JOURNAL OF ELECTROANALYTICAL CHEMISTRY, 1999, 469 (02) :176-181
[4]  
DAS HK, 2005, TXB BIOTECHNOLOGY, P695
[5]   Polypyrrole nanotube array sensor for enhanced adsorption of glucose oxidase in glucose biosensors [J].
Ekanayake, E. M. I. Mala ;
Preethichandra, D. M. G. ;
Kaneto, Keiichi .
BIOSENSORS & BIOELECTRONICS, 2007, 23 (01) :107-113
[6]  
EKANAYAKE EMI, 2007, P IMTC WARS POL MAY, P1
[7]   Immobilization of GOD on electrochemically synthesized Ppy-PVS composite film by cross-linking via glutaraldehyde for determination of glucose [J].
Gade, V. K. ;
Shirale, D. J. ;
Gaikwad, P. D. ;
Savale, P. A. ;
Kakde, K. P. ;
Kharat, H. J. ;
Shirsat, M. D. .
REACTIVE & FUNCTIONAL POLYMERS, 2006, 66 (12) :1420-1426
[8]   Bifunctional hydrogen peroxide electrode as an amperometric transducer for biosensors [J].
Gamburzev, S ;
Atanasov, P ;
Ghindilis, AL ;
Wilkins, E ;
Kaisheva, A ;
Iliev, I .
SENSORS AND ACTUATORS B-CHEMICAL, 1997, 43 (1-3) :70-77
[9]   Structural development of a minimally invasive sensor chip for blood glucose monitoring [J].
Kaimori, Shingo ;
Kitamura, Takahiko ;
Ichino, Moriyasu ;
Hosoya, Toshifumi ;
Kurusu, Fumiyo ;
Ishikawa, Tomoko ;
Nakamura, Hideaki ;
Gotoh, Masao ;
Karube, Isao .
ANALYTICA CHIMICA ACTA, 2006, 573 :104-109
[10]   Glucose oxidase electrodes based on microstructured polypyrrole films [J].
Ma, MM ;
Qu, LT ;
Shi, GQ .
JOURNAL OF APPLIED POLYMER SCIENCE, 2005, 98 (06) :2550-2554