A simple method for preparing calibration standards for the three working axes of scanning probe microscope piezo scanners

被引:12
作者
Alliata, D [1 ]
Cecconi, C [1 ]
Nicolini, C [1 ]
机构
[1] UNIV GENOA,INST BIOPHYS,I-16153 GENOA,ITALY
关键词
D O I
10.1063/1.1146803
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
A method for preparing samples suitable for calibrating scanning probe microscopes (SPM) and for eliminating any distortions in images is described. Samples consist of polystyrene particles organized in monolayers and bilayers with hexagonal-ordered domains. The monolayer is not uniform, but is characterized by areas without particles. These discontinuities allow the measurement of the thickness of the monolayer in order to calibrate the z axes, while the lattice constant of the domains can be used as a calibration standard for the x and y axes. The nondeformability of the particles after the deposition on the substrate has been studied by an optical microscope, equipped for interferometric measurements, scanning force microscopy, and scanning tunneling microscopy. The use of these standards directly as substrates for samples is proposed to correct the distortions in the SPM images. (C) 1996 American Institute of Physics.
引用
收藏
页码:748 / 751
页数:4
相关论文
共 14 条
[1]   A COMPACT SCANNING TUNNELLING MICROSCOPE WITH THERMAL COMPENSATION [J].
ALBREKTSEN, O ;
MADSEN, LL ;
MYGIND, J ;
MORCH, KA .
JOURNAL OF PHYSICS E-SCIENTIFIC INSTRUMENTS, 1989, 22 (01) :39-42
[2]   COMPUTER CORRECTION FOR DISTORTED STM IMAGES [J].
CAI, CZ ;
CHEN, XY ;
SHU, QQ ;
ZHENG, XL .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1992, 63 (12) :5649-5652
[3]   FROM ATOMS TO INTEGRATED-CIRCUIT CHIPS, BLOOD-CELLS, AND BACTERIA WITH THE ATOMIC FORCE MICROSCOPE [J].
GOULD, SAC ;
DRAKE, B ;
PRATER, CB ;
WEISENHORN, AL ;
MANNE, S ;
HANSMA, HG ;
HANSMA, PK ;
MASSIE, J ;
LONGMIRE, M ;
ELINGS, V ;
NORTHERN, BD ;
MUKERGEE, B ;
PETERSON, CM ;
STOECKENIUS, W ;
ALBRECHT, TR ;
QUATE, CF .
JOURNAL OF VACUUM SCIENCE & TECHNOLOGY A-VACUUM SURFACES AND FILMS, 1990, 8 (01) :369-373
[4]   ACTIN FILAMENT DYNAMICS IN LIVING GLIAL-CELLS IMAGED BY ATOMIC FORCE MICROSCOPY [J].
HENDERSON, E ;
HAYDON, PG ;
SAKAGUCHI, DS .
SCIENCE, 1992, 257 (5078) :1944-1946
[5]   Z CALIBRATION OF THE ATOMIC-FORCE MICROSCOPE BY MEANS OF A PYRAMIDAL TIP [J].
JENSEN, F .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1993, 64 (09) :2595-2597
[6]   POLYSTYRENE LATEX-PARTICLES AS A SIZE CALIBRATION FOR THE ATOMIC FORCE MICROSCOPE [J].
LI, Y ;
LINDSAY, SM .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1991, 62 (11) :2630-2633
[7]  
LOCATELLI M, 1991, REV SCI INSTRUM, V59, P661
[8]   TRUE ATOMIC-RESOLUTION BY ATOMIC FORCE MICROSCOPY THROUGH REPULSIVE AND ATTRACTIVE FORCES [J].
OHNESORGE, F ;
BINNIG, G .
SCIENCE, 1993, 260 (5113) :1451-1456
[9]   CALIBRATION OF THE ELECTRICAL RESPONSE OF PIEZOELECTRIC ELEMENTS AT LOW-VOLTAGE USING LASER INTERFEROMETRY [J].
RIIS, E ;
SIMONSEN, H ;
WORM, T ;
NIELSEN, U ;
BESENBACHER, F .
APPLIED PHYSICS LETTERS, 1989, 54 (25) :2530-2531
[10]   SURFACE ORDER AND STABILITY OF LANGMUIR-BLODGETT-FILMS [J].
SCHWARTZ, DK ;
GARNAES, J ;
VISWANATHAN, R ;
ZASADZINSKI, JAN .
SCIENCE, 1992, 257 (5069) :508-511