Full range complex spectral domain optical coherence tomography without additional phase shifters

被引:127
作者
Baumann, Bernhard [1 ]
Pircher, Michael [1 ]
Goetzinger, Erich [1 ]
Hitzenberger, Christoph K. [1 ]
机构
[1] Med Univ Vienna, Ctr Biomed Engn & Phys, A-1090 Vienna, Austria
来源
OPTICS EXPRESS | 2007年 / 15卷 / 20期
基金
奥地利科学基金会;
关键词
D O I
10.1364/OE.15.013375
中图分类号
O43 [光学];
学科分类号
070207 ; 0803 ;
摘要
We demonstrate a new full range complex spectral domain optical coherence tomography ( FRC SD-OCT) method. Other than FRC SD-OCT systems reported in literature, which employed devices such as electro-/ acousto optic modulators or piezo-driven mirrors providing the phase modulations necessary for retrieval of the complex-valued signal, the system presented works without any additional phase shifting device. The required phase shift is introduced by the galvanometer scanner used for transversally scanning the sample beam. By means of a slight displacement of the probe beam with respect to the scanning mirror's pivot axis, the sample arm length and thus the phase is continuously modulated as the beam is scanned in lateral direction. From such modulated spectral data, the complex-valued data yielding a twofold increase of accessible depth range can be calculated using an algorithm based on the Hilbert transform. To demonstrate the performance of our method quantitative measurements of the suppression of mirror images as a function of induced phase shift were performed. In order to validate the FRC SD-OCT technique for high-speed imaging of biological tissue, we present full-range images of the human anterior chamber in vivo. (C) 2007 Optical Society of America.
引用
收藏
页码:13375 / 13387
页数:13
相关论文
共 22 条
[11]  
Häusler G, 1998, J BIOMED OPT, V3, P21, DOI 10.1117/1.429899
[12]   OPTICAL COHERENCE TOMOGRAPHY [J].
HUANG, D ;
SWANSON, EA ;
LIN, CP ;
SCHUMAN, JS ;
STINSON, WG ;
CHANG, W ;
HEE, MR ;
FLOTTE, T ;
GREGORY, K ;
PULIAFITO, CA ;
FUJIMOTO, JG .
SCIENCE, 1991, 254 (5035) :1178-1181
[13]  
Laser Institute of America, 2000, AM NAT STAND SAF US
[14]   Performance of fourier domain vs. time domain optical coherence tomography [J].
Leitgeb, R ;
Hitzenberger, CK ;
Fercher, AF .
OPTICS EXPRESS, 2003, 11 (08) :889-894
[15]   Phase-shifting algorithm to achieve high-speed long-depth-range probing by frequency-domain optical coherence tomography [J].
Leitgeb, RA ;
Hitzenberger, CK ;
Fercher, AF ;
Bajraszewski, T .
OPTICS LETTERS, 2003, 28 (22) :2201-2203
[16]   En-face coherence imaging using galvanometer scanner modulation [J].
Podoleanu, AG ;
Dobre, GM ;
Jackson, DA .
OPTICS LETTERS, 1998, 23 (03) :147-149
[17]   Coherence imaging by use of a Newton rings sampling function [J].
Podoleanu, AG ;
Dobre, GM ;
Webb, DJ ;
Jackson, DA .
OPTICS LETTERS, 1996, 21 (21) :1789-1791
[18]   Complex spectral OCT in human eye imaging in vivo [J].
Targowski, P ;
Wojtkowski, M ;
Kowalczyk, A ;
Bajraszewski, T ;
Szkulmowski, M ;
Gorczynska, W .
OPTICS COMMUNICATIONS, 2004, 229 (1-6) :79-84
[19]   In vivo full range complex Fourier domain optical coherence tomography [J].
Wang, Ruikang K. .
APPLIED PHYSICS LETTERS, 2007, 90 (05)
[20]   Full range complex spectral optical coherence tomography technique in eye imaging [J].
Wojtkowski, M ;
Kowalczyk, A ;
Leitgeb, R ;
Fercher, AF .
OPTICS LETTERS, 2002, 27 (16) :1415-1417