Accuracy of automated volumetry of pulmonary nodules across different multislice CT scanners
被引:77
作者:
Das, Marco
Ley-Zaporozhan, Julia
Gietema, H. A.
Czech, Andre
Muehlenbruch, Georg
Mahnken, Andreas H.
Katoh, Markus
Bakai, Annemarie
Salganicoff, Marcos
Diederich, Stefan
Prokop, Mathias
Kauczor, Hans-Ulrich
Guenther, Rolf W.
Wildberger, Joachim E.
机构:
[1] Rhein Westfal TH Aachen Univ, Dept Diagnost Radiol, D-52072 Aachen, Germany
[2] German Canc Res Ctr, Dept Radiol, D-6900 Heidelberg, Germany
[3] Univ Utrecht Hosp, Dept Diagnost Radiol, Utrecht, Netherlands
[4] Marien Hosp, Dept Diagnost Radiol, Dusseldorf, Germany
关键词:
multislice CT;
pulmonary nodule;
nodule volumetry;
automated volumetry;
chest CT;
D O I:
10.1007/s00330-006-0562-1
中图分类号:
R8 [特种医学];
R445 [影像诊断学];
学科分类号:
1002 ;
100207 ;
1009 ;
摘要:
The purpose of this study was to compare the accuracy of an automated volumetry software for phantom pulmonary nodules across various 16-slice multislice spiral CT (MSCT) scanners from different vendors. A lung phantom containing five different nodule categories (intraparenchymal, around a vessel, vessel attached, pleural, and attached to the pleura), with each category comprised of 7-9 nodules (total, n=40) of varying sizes (diameter 3-10 mm; volume 6.62 mm(3)-525 mm(3)), was scanned with four different 16-slice MSCT scanners (Siemens, GE, Philips, Toshiba). Routine and low-dose chest protocols with thin and thick collimations were applied. The data from all scanners were used for further analysis using a dedicated prototype volumetry software. Absolute percentage volume errors (APE) were calculated and compared. The mean APE for all nodules was 8.4% ( +/- 7.7%) for data acquired with the 16-slice Siemens scanner, 14.3% ( +/- 11.1%) for the GE scanner, 9.7% ( +/- 9.6%) for the Philips scanner and 7.5% ( +/- 7.2%) for the Toshiba scanner, respectively. The lowest APEs were found within the diameter size range of 5-10 mm and volumes > 66 mm(3). Nodule volumetry is accurate with a reasonable volume error in data from different scanner vendors. This may have an important impact for intraindividual follow-up studies.
引用
收藏
页码:1979 / 1984
页数:6